Jin She1, Xizhou Guan2, Yanyong Liu3, Haiyan Xiang3.
Abstract
This study validated the RisingSun RS-651 blood pressure (BP) monitor based on auscultation in adults according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard. The RS-651 device was evaluated in a study of 97 participants. The same arm simultaneous method, as defined in the ANSI/AAMI/ISO standard, was used. The mean differences±standard deviation for criterion 1 were 0.8±2.3 mm Hg for systolic BP (SBP) and -0.1±2.9 mm Hg for diastolic BP (DBP). Analysis for criterion 2 resulted in values of 0.8±1.5 mm Hg for SBP and -0.1±2.1 mm Hg for DBP. All of the data fulfilled the ANSI/AAMI/ISO 81060-2:2013 standard requirements to pass the validation. The RisingSun RS-651 device can be recommended for both clinical and self/home use in adults according to the ANSI/AAMI/ISO 81060-2:2013 standard.Entities:
Mesh:
Year: 2016 PMID: 27271472 PMCID: PMC5216443 DOI: 10.1111/jch.12859
Source DB: PubMed Journal: J Clin Hypertens (Greenwich) ISSN: 1524-6175 Impact factor: 3.738
Figure 1Bland‐Altman plots for the differences between the RS‐651 readings and the observers’ measurements for systolic blood pressure. The x‐axis represents the average of systolic blood pressure measured by the RS‐651 device (test) and the mercury sphygmomanometer (reference). The y‐axis represents the difference between the systolic blood pressure obtained by the RS‐651 device and the mercury sphygmomanometer. A positive value indicates that the device measurement is greater than the observer's measurement.
Figure 2Bland‐Altman plots for the differences between the RS‐651 readings and the observers’ measurements for diastolic blood pressure. The x‐axis represents the average of diastolic blood pressure measured by the RS‐651 device (test) and the mercury sphygmomanometer (reference). The y‐axis represents the difference between the diastolic blood pressure obtained by the RS‐651 device and the mercury sphygmomanometer. A positive value indicates that the device measurement is greater than the observer's measurement.