Abstract

OBJECTIVE: The Tensoval Duo Control II is an automated upper arm device that uses a combination of oscillometric and auscultatory technology to determine blood pressure noninvasively. The accuracy of this device was assessed according to the British Hypertension Society (BHS) protocol and the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) in an adult population.
METHODS: Ethical approval was obtained. Eighty-five and 33 adult individuals, respectively, were recruited to fulfil the requirements of each protocol. Trained observers took nine sequential same-arm measurements alternating between a mercury sphygmomanometer and the device. The device had to achieve at least a B grade for both systolic and diastolic pressures to pass the BHS protocol and had to fulfil the criteria of all three phases of the ESH-IP2 protocol to receive recommendation.
RESULTS: The device achieved an A/A grading for the BHS protocol and passed all three phases of the ESH-IP2 protocol. The mean difference±SD for the BHS/ESH protocols, respectively, was -1.8±6.5/-0.7±5.7 mmHg for systolic pressure and 1.9±5.1/2.4±4.5 mmHg for diastolic pressure. The device maintained its A/A grading throughout the low-pressure, medium-pressure and high-pressure ranges.
CONCLUSION: The Tensoval Duo Control II device is recommended for clinical and home use according to both the BHS and the ESH-IP2 standard.

Entities:  

Mesh:

Year:  2013        PMID: 23640067     DOI: 10.1097/MBP.0b013e328360fb52

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Validation of the RisingSun RS-651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060-2:2013 Standard.

Authors:  Jin She; Xizhou Guan; Yanyong Liu; Haiyan Xiang
Journal:  J Clin Hypertens (Greenwich)       Date:  2016-06-07       Impact factor: 3.738

2.  Quantitative Assessment of Blood Pressure Measurement Accuracy and Variability from Visual Auscultation Method by Observers without Receiving Medical Training.

Authors:  Wenai Chen; Fei Chen; Yong Feng; Aiqing Chen; Dingchang Zheng
Journal:  Biomed Res Int       Date:  2017-12-20       Impact factor: 3.411

3.  Influence of Malondialdehyde and Matrix Metalloproteinase-9 on Progression of Carotid Atherosclerosis in Chronic Renal Disease with Cardiometabolic Syndrome.

Authors:  Senija Rašić; Damir Rebić; Sabaheta Hasić; Ismar Rašić; Marina Delić Šarac
Journal:  Mediators Inflamm       Date:  2015-10-11       Impact factor: 4.711

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.