Evaluation of intraocular lens position during phacoemulsification using intraoperative spectral-domain optical coherence tomography.

Lyubomyr M Lytvynchuk1, Carl G Glittenberg2, Christiane I Falkner-Radler2, Beatrix Neumaier-Ammerer2, Eva Smretschnig2, Stefan Hagen2, Siamak Ansari-Shahrezaei2, Susanne Binder2.   

Abstract

PURPOSE: To assess the position of intraocular lenses (IOLs) at the end of standard phacoemulsification with intraoperative spectral-domain optical coherence tomography (SD-OCT). SETTINGS: Department of Ophthalmology, Rudolf Foundation Hospital, Vienna, Austria.
DESIGN: Prospective case series.
METHODS: Standard phacoemulsification with IOL implantation was performed. The Rescan 700 SD-OCT system was used for intraoperative imaging. The anterior segment of the eye was scanned using SD-OCT at the end of the surgery. The distance from the IOL optic center and the IOL optic edge to the posterior capsule was measured postoperatively using graphic software.
RESULTS: The study comprised 74 patients (101 eyes). The mean axial length was 23.97 mm (range 21.43 to 28.61 mm). The mean IOL power was 20.39 diopters (D) (range 6.5 to 27.5 D). Contact between the IOL and posterior capsule was absent in 88 cases (87.13%), and partial or full contact was present in 13 cases (12.87%). The mean distance between the IOL central optic and posterior capsule was 0.71 pixel (range 0.06 to 1.38 pixels) in 99 cases (98.02%). In 42 cases (57.53%), partial contact between the IOL edges and the posterior capsule was noticed. The mean distance between the IOL edge and posterior capsule was 0.21 pixel (range 0.04 to 0.92 pixel).
CONCLUSIONS: Intraoperative SD-OCT facilitated the imaging of IOL position during standard phacoemulsification. Contact between the IOL central optic and posterior capsule at the end of the surgery occurred rarely. Improved IOL design should be considered. FINANCIAL DISCLOSURE: Drs. Binder and Glittenberg are consultants to Carl Zeiss Meditech AG. None of the other authors has a financial or proprietary interest in any material or method mentioned.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27255245     DOI: 10.1016/j.jcrs.2016.01.044

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  Influence of lens position as detected by an anterior segment analysis system on postoperative refraction in cataract surgery.

Authors:  Jia-Ju Zhang; Jian-Qing Li; Chen Li; Yi-Hong Cao; Pei-Rong Lu
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

Review 2.  Intraoperative Anterior Segment Optical Coherence Tomography in the Management of Cataract Surgery: State of the Art.

Authors:  Mario Damiano Toro; Serena Milan; Daniele Tognetto; Robert Rejdak; Ciro Costagliola; Sandrine Anne Zweifel; Chiara Posarelli; Michele Figus; Magdalena Rejdak; Teresio Avitabile; Adriano Carnevali; Rosa Giglio
Journal:  J Clin Med       Date:  2022-07-04       Impact factor: 4.964

3.  Combined anterior segment OCT and wavefront-based autorefractor using a shared beam.

Authors:  Marco Ruggeri; Giulia Belloni; Yu-Cherng Chang; Heather Durkee; Ettore Masetti; Florence Cabot; Sonia H Yoo; Arthur Ho; Jean-Marie Parel; Fabrice Manns
Journal:  Biomed Opt Express       Date:  2021-10-06       Impact factor: 3.562

4.  Optimizing optical outcomes of intraocular lens implantation by achieving centration on visual axis.

Authors:  Prakhyat Roop
Journal:  Indian J Ophthalmol       Date:  2017-12       Impact factor: 1.848

5.  Optical Coherence Tomography Observation of Gonio Structures during Microhook Ab Interno Trabeculotomy.

Authors:  Masaki Tanito
Journal:  J Ophthalmol       Date:  2017-08-06       Impact factor: 1.909

Review 6.  Intraoperative optical coherence tomography in anterior segment surgeries.

Authors:  Jeewan S Titiyal; Manpreet Kaur; Ruchita Falera
Journal:  Indian J Ophthalmol       Date:  2017-02       Impact factor: 1.848

7.  Observation of Gonio Structures during Microhook Ab Interno Trabeculotomy Using a Novel Digital Microscope with Integrated Intraoperative Optical Coherence Tomography.

Authors:  Akiko Ishida; Kazunobu Sugihara; Tomoki Shirakami; Aika Tsutsui; Kaoru Manabe; Masaki Tanito
Journal:  J Ophthalmol       Date:  2020-07-17       Impact factor: 1.909

8.  Quantitative Analysis of Gap Between the Intraocular Lens and Posterior Capsule Using Microscope-Integrated Optical Coherence Tomography in Eyes Undergoing Phacoemulsification.

Authors:  Ritu Nagpal; Chetan Shakkarwal; Rinky Agarwal; Rahul Kumar Bafna; Prafulla Kumar Maharana; Namrata Sharma
Journal:  Clin Ophthalmol       Date:  2021-05-10

Review 9.  Clinical applications for intraoperative optical coherence tomography: a systematic review.

Authors:  Marc B Muijzer; Peter A W J Schellekens; Henny J M Beckers; Joke H de Boer; Saskia M Imhof; Robert P L Wisse
Journal:  Eye (Lond)       Date:  2021-07-16       Impact factor: 3.775

Review 10.  Diagnostic imaging versus surgical procedure: intra- and postoperative OCT evaluation of sutureless scleral-fixated intraocular lens implantation and possible related complications.

Authors:  Raffaele Nuzzi; Alessandro Rossi
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-03-12       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.