A comparative study on visual and optical performance of Akreos AO and Kontur AB IOLs after phacoemulsification cataract surgery.

Hamid Gharaee1, Masoud Zabihifard1, Alireza Eslampour2, Samira Hassanzadeh1, Masoud Shafiee1.   

Abstract

PURPOSE: Akreos AO and Kontur AB are two commonly used intraocular lenses (IOLs) in Iran. This study was designed to evaluate the visual performance of these lenses.
METHODS: In a comparative interventional study, 35 patients (70 eyes) were recruited, and each IOL was implanted in one eye of the patients, randomly. Best corrected visual acuity (BCVA), contrast sensitivity, aberrometric analysis, and depth of focus were evaluated 1 month and 3 months postoperatively. A visual quality questionnaire was also filled for each eye, and the results were compared.
RESULTS: Mean age of the patients was 60.97 ± 7.00 years. BCVA was not significantly different between the two eyes, before, 1 month, and 3 months postoperatively (p > 0.05 for all). Photopic and mesopic contrast sensitivity was not different between the two lenses instead of photopic 18 cycles per degree, 3 months postoperatively and in mesopic 6 cycles per degree 1 month postoperatively (p = 0.034 and p = 0.002, respectively). Aberrometric factors including HoRMS, Total RMS, and Higher order without Z(4,0) were not significantly different between the two lenses (p > 0.05 for all), but they were slightly lower for Akreos AO. Post-operative distance-corrected visual acuity for intermediate and near vision were not different between the two groups (p > 0.05, respectively).
CONCLUSION: Visual performance of Akreos AO and Kontur AB is similar. However, contrast sensitivity and aberrometric parameters are slightly better for Akreos AO IOL.

Entities:  

Keywords:  Aberrometry; Akreos AO; Contrast sensitivity; Depth of focus; Kontur AB

Year:  2016        PMID: 27239596      PMCID: PMC4881222          DOI: 10.1016/j.joco.2016.02.001

Source DB:  PubMed          Journal:  J Curr Ophthalmol        ISSN: 2452-2325


Introduction

The human cornea has positive spherical aberration because it focuses peripheral light anterior to the point of convergence for rays traced through the center. With age, positive spherical aberration increases, and, accordingly, mesopic contrast sensitivity decreases. In such an eye, implantation of a traditional spherical intraocular lens (IOL) will add the positive spherical aberration, and despite the superior clarity of the IOL, it may not improve mesopic functional acuity contrast in phakic eyes. To balance positive spherical aberration of the cornea, commercially available aspherical IOLs present different amounts of spherical aberration. New aspherical IOL designs such as Akreos Advanced Optic (Akreos AO; Bausch & Lomb, Inc., Rochester, New York, USA) are considered aberration-free because their anterior and posterior prolate surfaces generate, theoretically, no negative spherical aberration. In this study, we tried to compare two commonly used aspherical IOLs (Akreos AO vs Kontur AB Medicontur Medical Engineering LTD, Zsámbék, Hungary) for postoperative objective and subjective visual function (aberrometric and contrast sensitivity results). In order to enhance post-operative visual function of our patients, we tried to find out the IOL which results in better postsurgical visual and optical outcomes. It is notable that we found no previous report about visual performance of Kontur AB IOL in scientific databases.

Methods

Seventy eyes of 35 patients with diagnosis of bilateral senile cataract grade III or more (according to LOCS criteria) were included in this comparative interventional study after receiving ethical consent. Exclusion criteria were axial length >24.9 mm or <22.5 mm, corneal astigmatism >1.50 diopter, any history of ocular pathology including opaque or irregular cornea, dry eye, amblyopia, glaucoma, and retinal disorders. There were also postsurgical exclusion criteria including any postoperative complication, postoperative best corrected visual acuity (BCVA) less than 20/30 (0.18Log MAR), IOL tilt or decentration, and not taking part in follow up sessions. A complete preoperative examination was performed including dilated funduscopy and slit lamp evaluation, measuring of axial length by Lenstar (Haag Streit, USA), and intraocular pressure (IOP) evaluation by Goldmann applanation tonometer. Patients were randomly allocated Akreos AO (Akreos AO, Bausch & Lomb, Inc) or Kontur AB IOLs to each eye. An expert surgeon performed a standard phaco chop phacoemulsification technique. Each IOL was implanted in an interval of about one month. Table 1 shows the specification of each study lens.
Table 1

Specifications of study lenses.

Lens typeLens materialPower range: Total diameterOptical diameterHaptic designAngulationA-constantOpticPCO protection
Akreos AOHydrophil acrylic (26.0%water content)0.0 to +15.0: 11.0 mm+15.5 to +22.0: 10.7 mm+22.5 to +30.0: 10.5 mm6.0 mm4 closed loops118.0Aberration-free, biconvex, aspheric360° anti PCO berrier, square edge design
Kontur ABHydrophil acrylic (26.0%water content))0.0 to +15.0: 11.0 mm+15.5 to +22.0: 10.7 mm+22.5 to +35.0: 10.5 mm6.0 mm4 closed loops118.0Aberration-free, biconvex, aspheric360° Special square edge

PCO: Posterior capsule opacity.

Preoperatively and 1 and 3 months postoperatively, all patients were evaluated and BCVA in 4 m and 40 cm was assessed by near and distance ETDRS acuity charts. One and 3 months postoperatively, photopic and mesopic (chart luminance: 85 cd/m2 and 3.0 cd/m2 respectively) contrast sensitivity was evaluated by CSV1000E chart (Vector Vision Inc., Dayton, Ohio, USA). Test distance was 2.5 m, and the results were recorded as log10contrast. Aberrometric findings were recorded 1 and 3 months postoperatively. Pupils were dilated with tropicamide 1.0%, and root mean square (RMS) and spherical aberration (z4) were measured by Technolas aberrometer (Zyoptix system; Bausch & Lamb, USA) for 5 and 6 mm pupil. Depth of focus was another factor which was compared between eyes. To measure this factor, we corrected distance refractive error and measured visual acuity (in Log MAR) for intermediate (1 m) and near (33.0 cm) distances without incorporating any addition. We also used a visual quality questionnaire which was previously used by Tester et al to compare postoperative quality of vision for two lenses.

Statistical analysis

SPSS statistical software (version 18.0, SPSS, Inc, Chicago, Illinois, USA) was used for statistical analysis. Non parametric Wilcoxon paired t-test was applied for comparisons between the two groups. For all results, p < 0.05 was considered significant.

Results

Seventy eyes of 35 patients were enrolled in this study. The mean age of patients was 60.97 years ± 7.00 (SD) (range: 49–72 years). All patients were attended in 1 and 3 months follow up after surgery. There was no significant difference in IOP, axial length, and keratometry between the two groups (Table 2).
Table 2

Demographic data.

Akreos AOKantur ABp-value
IOP (mmHg)Preoperative17.49 ± 2.1617.46 ± 2.150.956
Axial length (mm)Preoperative23.31 ± 0.4523.35 ± 0.040.654
Keratometry (D)Preoperative43.39 ± 0.9643.28 ± 1.010.630
1 month post-op43.63 ± 0.9543.58 ± 1.060.836
3months post-op43.38 ± 1.0643.48 ± 1.100.70
BCVA (LogMAR)Preoperative0.52 ± 0.180.51 ± 0.130.793
1 month post-op0.08 ± 0.070.08 ± 0.0540.98
3 months post-op0.053 ± 0.0450.051 ± 0.0410.77
Spherical equivalent (D)1 month post-op0.07 ± 1.370.11 ± 1.200.65
3 months post-op−0.06 ± 1.100.04 ± 1.070.793

D: Diopters, BCVA: Best corrected visual acuity, IOP: Intraocular pressure.

Before surgery and 1 and 3 months postsurgery, mean BCVA and mean spherical equivalent of refractive error were not significantly different between the 2 groups (Table 2). Three months postsurgically, photopic contrast sensitivity was significantly different between the 2 eyes in high spatial frequencies (18 cycles/deg) (p = 0.034). There was no statistically significant difference between the 2 groups in other spatial frequencies (p > 0.05 for all). One month after surgery, in mesopic condition, mean contrast sensitivity (in 6 cycles/deg) in eyes implanted with Akreos AO was significantly better than Kontur AB (p = 0.002), but there was no difference in other spatial frequencies and in 3 months postsurgery (p > 0.05 for all). Fig. 1(A–D) shows photopic and mesopic contrast sensitivity changes in 2 groups, 1 and 3 months postoperatively.
Fig. 1

Comparison of contrast sensitivity changes between 2 groups.

Aberrometric parameters (spherical aberration or Z(4,0), total RMS and HoRMS) was also compared between the two groups for 5 and 6 mm pupil. Results showed that there was no significant difference between groups for aberrometric parameters (p > 0.05 for all, Table 3). Also, mean undilated pupil size was not significantly different between the two groups (p = 0.658) (3.37 mm ± 0.64 in Akreos AO vs 3.38 mm ± 0.66 in Kontur AB groups.)
Table 3

Mean ± SD of HoRMS, total RMS and Z4 (μm) in two groups, 1 and 3 months postoperatively.

Pupil sizeAkreos AOKantur ABp-value
HoRMS1 month post op5 mm0.31 ± 0.050.31 ± 0.040.626
6 mm0.57 ± 0.080.55 ± 0.070.758
3 months post op5 mm0.30 ± 0.040.30 ± 0.040.937
6 mm0.58 ± 0.080.57 ± 0.090.525
Total RMS1 month post op5 mm1.67 ± 0.271.71 ± 0.260.113
6 mm2.79 ± 0.402.94 ± 0.370.104
3 months post op5 mm1.43 ± 0.261.54 ± 0.260.094
6 mm2.55 ± 0.392.72 ± 0.370.054
Z (4,0)1 month post op5 mm0.23 ± 0.050.24 ± 0.070.483
6 mm0.43 ± 0.090.43 ± 0.110.962
3 months post op5 mm0.22 ± 0.050.23 ± 0.060.26
6 mm0.40 ± 0.090.41 ± 0.110.787
Depth of focus was the other parameter which was compared between eyes implanted with Akreos AO and Kontur AB. Table 4 shows no significant difference for distance-corrected intermediate and near visual acuity, 1 and 3 months postoperatively.
Table 4

Mean distance-corrected visual acuity (LogMAR).

Akreos AOKantur ABp-value
1 month post-opIntermediate0.49 ± 0.120.52 ± 0.150.354
Near0.65 ± 0.200.69 ± 0.210.444
3 months post-opIntermediate0.46 ± 0.120.49 ± 0.130.418
Near0.60 ± 0.180.65 ± 0.180.276
Subjective reports of visual disturbances and light-induced complications of two lenses, 3 months postsurgery are given in Table 5.
Table 5

Subjective reports of visual complications 3 months post-surgery.

Visual disturbancesAkreos AO
Kantur AB
NoneMinimalAnnoyingDebilitatingNoneMinimalAnnoyingDebilitating
Light induced glare85.7%14.3%0.0%0.0%77.2%22.8%0.0%0.0%
Unwanted images88.6%11.4%0.0%0.0%82.9%17.1%0.0%0.0%
Increased eye sensitivity85.7%14.3%0.0%0.0%77.2%22.8%0.0%0.0%
Driving in to sunset-sunrise85.7%14.3%0.0%0.0%85.5%14.3%0.0%0.0%
Bright sunny day at noon85.7%14.3%0.0%0.0%80.0%20.0%0.0%0.0%
Brightly lit environment85.7%14.3%0.0%0.0%80.0%20.0%0.0%0.0%
Oncoming headlight at night85.7%14.3%0.0%0.0%77.2%22.8%0.0%0.0%

Discussion

Spherical and aspherical IOL designs have been compared in previous studies.3, 6, 7 The results show that mesopic contrast sensitivity and quality of vision can be improved by incorporating aspheric designs. On the other hand, comparing aspherical IOL designs, Johansson and associates showed that the higher perceived quality of vision (with the Akreos AO IOL in their study) could be because of differences in depth of field, IOL material, or IOL design. Our study was performed to compare quality of vision after implantation of two commonly used IOLs with aspheric designs and hydrophilic acrylic which were both claimed to be aberration free. Akreos AO is a well-known IOL, and previous studies have evaluated and compared it with other IOLs.3, 6, 8 We also assessed another aspherical IOL (Kontur AB), which to our knowledge, no study has been conducted on its performance. For minimization of the effects of factors other than the IOLs, such as ocular dominance or refractive deviations, which could affect the outcomes, we implanted each lens in each eye of one patient. Santiago and associates showed that aspherical Akreos AO IOL, induced lower values of spherical aberration, more patient satisfaction, better performance in the mesopic contrast sensitivity test and similar depth of focus compared with the spherical IOL Akreos Fit (study IOLs). According to our findings, 3 months postoperatively in relation to 1 month postoperatively, photopic and mesopic contrast sensitivity was increased in both eyes. Three months postoperatively, Akreos photopic contrast sensitivity in high spatial frequencies (18 cycles/deg) was significantly better than Kontur AB, which was in coordination with BCVA. Like previous studies we used aberrometric analysis as a determinant of postoperative visual function.1, 2, 3, 6, 7, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27 Santiago and associates in their study found that eyes implanted with the aspherical IOL Akreos AO showed lower values of spherical aberration and improved contrast sensitivity compared with eyes implanted with the spherical IOL Akreos Fit. By comparing HoRMS, total RMS and Z4 between two eyes (for 5 and 6 mm pupil), we found no significant difference between two lenses, but these factors were decreased 3 months postoperatively in comparison with 1 month postoperatively, which were in relation with postoperative BCVA and contrast sensitivity. According to previous studies, aspherical IOLs can decrease depth of focus (DoF) and worsen distance-corrected near vision by eliminating spherical aberration that a higher amount of spherical aberration results in a better DoF. In our study, we compared postoperative DoF for two aspherical IOLs. This factor is an important factor in postoperative patient satisfaction. Kontur AB and Akreos AO are considered aberration-free because their anterior and posterior prolate surfaces generate, theoretically, no negative spherical aberration compensate for the positive spherical aberration of the cornea,2, 16 so there is a higher amount of spherical aberration in the optical system. It seems that there will be no significant reduction of depth of focus with that value of spherical aberration. In our work, we incorporated Santiago and associates' technique in which distance-corrected near and intermediate visual acuity were studied as a measurement of depth of focus. Our results showed no significant difference between the two lenses at 1 and 3 months postoperatively. Also according to results of questionnaire of visual quality, 3 months after surgery more than 75% of patients in both groups had no visual disturbances and light-induced problems. Obviously, comprehensive studies with larger sample sizes, considering interfering factors in patient satisfaction, can result in a better comparison of these IOLs. In summary, we concluded that visual quality of Akreos AO and Kontur AB IOLs are similar, and both lenses have the same postoperative patient satisfaction. However, high spatial frequency mesopic and photopic contrast sensitivity and aberrometric findings of Akreos AO are more acceptable.
  27 in total

Review 1.  Analysis of the possible benefits of aspheric intraocular lenses: review of the literature.

Authors:  Robert Montés-Micó; Teresa Ferrer-Blasco; Alejandro Cerviño
Journal:  J Cataract Refract Surg       Date:  2009-01       Impact factor: 3.351

2.  Clinical outcomes and postoperative intraocular optical quality with a microincision aberration-free aspheric intraocular lens.

Authors:  Jorge L Alió; David P Piñero; Dolores Ortiz; Raúl Montalbán
Journal:  J Cataract Refract Surg       Date:  2009-09       Impact factor: 3.351

3.  Visual performance of aspherical and spherical intraocular lenses: intraindividual comparison of visual acuity, contrast sensitivity, and higher-order aberrations.

Authors:  Thomas Kasper; Jens Bühren; Thomas Kohnen
Journal:  J Cataract Refract Surg       Date:  2006-12       Impact factor: 3.351

4.  Comparison of pseudophakic visual quality in spherical and aspherical intraocular lenses.

Authors:  Hongguang Cui; Rongrong Hu; Yingying Zhang; Dinghua Lou
Journal:  Can J Ophthalmol       Date:  2009-06       Impact factor: 1.882

5.  Visual and optical performance of the Akreos Adapt Advanced Optics and Tecnis Z9000 intraocular lenses: Swedish multicenter study.

Authors:  Björn Johansson; Staffan Sundelin; Anna Wikberg-Matsson; Peter Unsbo; Anders Behndig
Journal:  J Cataract Refract Surg       Date:  2007-09       Impact factor: 3.351

6.  Binocular implantation of the Tecnis Z9000 or AcrySof MA60AC intraocular lens in routine cataract surgery: prospective randomized controlled trial comparing VF-14 scores.

Authors:  Bruce Allan
Journal:  J Cataract Refract Surg       Date:  2007-09       Impact factor: 3.351

7.  Intraindividual comparison of higher-order aberrations after implantation of aspherical and spherical intraocular lenses as a function of pupil diameter.

Authors:  Thomas Kasper; Jens Bühren; Thomas Kohnen
Journal:  J Cataract Refract Surg       Date:  2006-01       Impact factor: 3.351

8.  Improvements in vision-related quality of life with AcrySof IQ SN60WF aspherical intraocular lenses.

Authors:  I-Chan Lin; I-Jong Wang; Mei-Shu Lei; Luke L-K Lin; Fung-Rong Hu
Journal:  J Cataract Refract Surg       Date:  2008-08       Impact factor: 3.351

9.  Wavefront aberrations, depth of focus, and contrast sensitivity with aspheric and spherical intraocular lenses: fellow-eye study.

Authors:  Mayank A Nanavaty; David J Spalton; James Boyce; Shouvik Saha; John Marshall
Journal:  J Cataract Refract Surg       Date:  2009-04       Impact factor: 3.351

10.  Aspheric versus Spherical Posterior Chamber Intraocular Lenses.

Authors:  Mohammad-Reza Jafarinasab; Sepehr Feizi; Ahmad-Reza Baghi; Hossein Ziaie; Mehdi Yaseri
Journal:  J Ophthalmic Vis Res       Date:  2010-10
View more
  1 in total

Review 1.  Personalized aspheric intraocular lens implantation based on corneal spherical aberration: a review.

Authors:  Wei Du; Wei Lou; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2019-11-18       Impact factor: 1.779

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.