Literature DB >> 27228018

Participant-generated Cochlear Implant Programs: Speech Recognition, Sound Quality, and Satisfaction.

Robert T Dwyer1, Tony Spahr, Smita Agrawal, Chris Hetlinger, Jourdan T Holder, René H Gifford.   

Abstract

OBJECTIVE: To determine whether patient-derived programming of one's cochlear implant (CI) stimulation levels may affect performance outcomes.
BACKGROUND: Increases in patient population, device complexity, outcome expectations, and clinician responsibility have demonstrated the necessity for improved clinical efficiency.
METHODS: Eighteen postlingually deafened adult CI recipients (mean = 53 years; range, 24-83 years) participated in a repeated-measures, within-participant study designed to compare their baseline listening program to an experimental program they created.
RESULTS: No significant group differences in aided sound-field thresholds, monosyllabic word recognition, speech understanding in quiet, speech understanding in noise, nor spectral modulation detection (SMD) were observed (p > 0.05). Four ears (17%) improved with the experimental program for speech presented at 45 dB SPL and two ears (9%) performed worse. Six ears (27.3%) improved significantly with the self-fit program at +10 dB signal-to-noise ratio (SNR) and four ears (26.6%) improved in speech understanding at +5 dB SNR. No individual scored significantly worse when speech was presented in quiet at 60 dB SPL or in any of the noise conditions tested. All but one participant opted to keep at least one of the self-fitting programs at the completion of this study. Participants viewed the process of creating their program more favorably (t = 2.11, p = 0.012) and thought creating the program was easier than the traditional fitting methodology (t = 2.12, p = 0.003). Average time to create the self-fit program was 10 minutes, 10 seconds (mean = 9:22; range, 4:46-24:40).
CONCLUSIONS: Allowing experienced adult CI recipients to set their own stimulation levels without clinical guidance is not detrimental to success.

Entities:  

Mesh:

Year:  2016        PMID: 27228018      PMCID: PMC5302051          DOI: 10.1097/MAO.0000000000001076

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  25 in total

1.  "Mini-mental state". A practical method for grading the cognitive state of patients for the clinician.

Authors:  M F Folstein; S E Folstein; P R McHugh
Journal:  J Psychiatr Res       Date:  1975-11       Impact factor: 4.791

2.  User preference and reliability of bilateral hearing aid gain adjustments.

Authors:  Benjamin W Y Hornsby; H Gustav Mueller
Journal:  J Am Acad Audiol       Date:  2008-02       Impact factor: 1.664

3.  Spectral modulation detection and vowel and consonant identifications in cochlear implant listeners.

Authors:  Aniket A Saoji; Leonid Litvak; Anthony J Spahr; David A Eddins
Journal:  J Acoust Soc Am       Date:  2009-09       Impact factor: 1.840

4.  Clinical assessment of spectral modulation detection for adult cochlear implant recipients: a non-language based measure of performance outcomes.

Authors:  René H Gifford; Andrea Hedley-Williams; Anthony J Spahr
Journal:  Int J Audiol       Date:  2014-01-23       Impact factor: 2.117

5.  Expanded pediatric cochlear implant candidacy.

Authors:  J Thomas Roland; Susan B Waltzman
Journal:  Otolaryngol Head Neck Surg       Date:  2015-04       Impact factor: 3.497

6.  The effects of different frequency responses on sound quality judgments and speech intelligibility.

Authors:  A Gabrielsson; B N Schenkman; B Hagerman
Journal:  J Speech Hear Res       Date:  1988-06

7.  A Within-Subject Comparison of Bimodal Hearing, Bilateral Cochlear Implantation, and Bilateral Cochlear Implantation With Bilateral Hearing Preservation: High-Performing Patients.

Authors:  René H Gifford; Colin L W Driscoll; Timothy J Davis; Pam Fiebig; Alan Micco; Michael F Dorman
Journal:  Otol Neurotol       Date:  2015-09       Impact factor: 2.311

8.  Cochlear implantation in nontraditional candidates: preliminary results in adolescents with asymmetric hearing loss.

Authors:  Jamie H Cadieux; Jill B Firszt; Ruth M Reeder
Journal:  Otol Neurotol       Date:  2013-04       Impact factor: 2.311

9.  Unilateral deafness in children: audiologic and subjective assessment of hearing ability after cochlear implantation.

Authors:  Frederike Hassepass; Antje Aschendorff; Thomas Wesarg; Stefanie Kröger; Roland Laszig; Rainer L Beck; Christian Schild; Susan Arndt
Journal:  Otol Neurotol       Date:  2013-01       Impact factor: 2.311

10.  A short form of the Speech, Spatial and Qualities of Hearing scale suitable for clinical use: the SSQ12.

Authors:  William Noble; Niels Søgaard Jensen; Graham Naylor; Navjot Bhullar; Michael A Akeroyd
Journal:  Int J Audiol       Date:  2013-06       Impact factor: 2.117

View more
  1 in total

1.  Validating a Quick Spectral Modulation Detection Task.

Authors:  David M Landsberger; Robert T Dwyer; Natalia Stupak; René H Gifford
Journal:  Ear Hear       Date:  2019 Nov/Dec       Impact factor: 3.570

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.