Performance of cone-beam computed tomography and multidetector computed tomography in diagnostic imaging of the midface: A comparative study on Phantom and cadaver head scans.

Simon Veldhoen1,2, Maximilian Schöllchen3, H Hanken3, C Precht3, F O Henes4, G Schön5, H D Nagel6, U Schumacher7, M Heiland3, G Adam4, M Regier4.   

Abstract

PURPOSE: To compare multidetector computed tomography (MDCT) and cone-beam computed tomography (CBCT) regarding radiation, resolution, image noise, and image quality.
METHODS: CBCT and 256-MDCT were compared based on three scan protocols: Standard-dose (≈24 mGy), reduced-dose (≈9 mGy), and low-dose (≈4 mGy). MDCT images were acquired in standard- and high-resolution mode (HR-MDCT) and reconstructed using filtered back projection (FBP) and iterative reconstruction (IR). Spatial resolution in linepairs (lp) and objective image noise (OIN) were assessed using dedicated phantoms. Image quality was assessed in scans of 25 cadaver heads using a Likert scale.
RESULTS: OIN was markedly higher in FBP-MDCT when compared to CBCT. IR lowered the OIN to comparable values in standard-mode MDCT only. CBCT provided a resolution of 13 lp/cm at standard-dose and 11 lp/cm at reduced-dose vs. 11 lp/cm and 10 lp/cm in HR-MDCT. Resolution of 10 lp/cm was observed for both devices using low-dose settings. Quality scores of MDCT and CBCT did not differ at standard-dose (CBCT, 3.4; MDCT, 3.3-3.5; p > 0.05). Using reduced- and low-dose protocols, CBCT was superior (reduced-dose, 3.2 vs. 2.8; low dose, 3.0 vs. 2.3; p < 0.001).
CONCLUSION: Using the low-dose protocol, the assessed CBCT provided better objective and subjective image quality and equality in resolution. Similar image quality, but better resolution using CBCT was observed at higher exposure settings. KEY POINTS: • The assessed CBCT device provided better image quality at lower doses. • Objective and subjective image quality were comparable using higher exposure settings. • CBCT showed superior spatial resolution in standard-dose and reduced-dose settings. • Modern noise-reducing tools are used in CBCT devices currently. • MDCT should be preferred for assessment of soft-tissue injuries and oncologic imaging.

Keywords:  Cone-beam computed tomography; Multidetector computed tomography; Phantom imaging; Prospective study; Radiometry

Mesh:

Year:  2016        PMID: 27169574     DOI: 10.1007/s00330-016-4387-2

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  29 in total

1.  Transthoracic fine-needle aspiration biopsy of the lungs using a C-arm cone-beam CT system: diagnostic accuracy and post-procedural complications.

Authors:  W J Lee; S Chong; J S Seo; H J Shim
Journal:  Br J Radiol       Date:  2011-10-18       Impact factor: 3.039

Review 2.  CT & CBCT imaging: assessment of the orbits.

Authors:  David C Hatcher
Journal:  Oral Maxillofac Surg Clin North Am       Date:  2012-09-13       Impact factor: 2.802

3.  Cortical bone thickness of the alveolar process measured with cone-beam computed tomography in patients with different facial types.

Authors:  Fulya Ozdemir; Murat Tozlu; Derya Germec-Cakan
Journal:  Am J Orthod Dentofacial Orthop       Date:  2013-02       Impact factor: 2.650

4.  Accuracy of volumetric measurement of simulated root resorption lacunas based on cone beam computed tomography.

Authors:  Y Wang; S He; Y Guo; S Wang; S Chen
Journal:  Orthod Craniofac Res       Date:  2013-02-19       Impact factor: 1.826

5.  CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT).

Authors:  Seppo K Koskinen; Ville V Haapamäki; Jari Salo; Nina C Lindfors; Mika Kortesniemi; Lauri Seppälä; Kimmo T Mattila
Journal:  Skeletal Radiol       Date:  2012-09-19       Impact factor: 2.199

6.  [Image quality of thin- and thick-slice MSCT reconstructions in low-contrast objects (liver lesions) with equal doses].

Authors:  U Wedegärtner; M Lorenzen; H D Nagel; A Koops; C Weber; C Nolte-Ernsting; V Schoder; G Adam
Journal:  Rofo       Date:  2004-11

7.  Variability of dental cone beam CT grey values for density estimations.

Authors:  R Pauwels; O Nackaerts; N Bellaiche; H Stamatakis; K Tsiklakis; A Walker; H Bosmans; R Bogaerts; R Jacobs; K Horner
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

8.  [Cross-sectional imaging in dentomaxillofacial diagnostics: dose comparison of dental MSCT and NewTom 9000 DVT].

Authors:  E Coppenrath; F Draenert; U Lechel; R Veit; T Meindl; M Reiser; U Mueller-Lisse
Journal:  Rofo       Date:  2008-05

9.  Multidetector CT urography: experimental analysis of radiation dose reduction in an animal model.

Authors:  Joern Kemper; Marc Regier; Paul M Bansmann; Philipp G C Begemann; Alexander Stork; Hans Dieter Nagel; Gerhard Adam; Claus Nolte-Ernsting
Journal:  Eur Radiol       Date:  2007-02-14       Impact factor: 5.315

10.  Apical root resorption due to orthodontic treatment detected by cone beam computed tomography.

Authors:  Iury O Castro; Ana H G Alencar; José Valladares-Neto; Carlos Estrela
Journal:  Angle Orthod       Date:  2012-07-19       Impact factor: 2.079

View more
  12 in total

1.  Initial evaluation of image performance of a 3-D x-ray system: phantom-based comparison of 3-D tomography with conventional computed tomography.

Authors:  Robyn Melanie Benz; Meritxell Alzamora Garcia; Felix Amsler; Johannes Voigt; Andreas Fieselmann; Anna Lucja Falkowski; Bram Stieltjes; Anna Hirschmann
Journal:  J Med Imaging (Bellingham)       Date:  2018-03-06

Review 2.  Precision of cone beam CT to assess periodontal bone defects: a systematic review and meta-analysis.

Authors:  Letícia Fernanda Haas; Glaucia Santos Zimmermann; G De Luca Canto; Carlos Flores-Mir; Márcio Corrêa
Journal:  Dentomaxillofac Radiol       Date:  2017-10-27       Impact factor: 2.419

3.  Artefacts in multimodal imaging of titanium, zirconium and binary titanium-zirconium alloy dental implants: an in vitro study.

Authors:  Ralf Smeets; Maximilian Schöllchen; Tobias Gauer; Ghazal Aarabi; Alexandre T Assaf; Carsten Rendenbach; Benedicta Beck-Broichsitter; Jan Semmusch; Jan Sedlacik; Max Heiland; Jens Fiehler; Susanne Siemonsen
Journal:  Dentomaxillofac Radiol       Date:  2016-12-02       Impact factor: 2.419

4.  Predicting image properties in penalized-likelihood reconstructions of flat-panel CBCT.

Authors:  Wenying Wang; Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Med Phys       Date:  2018-11-20       Impact factor: 4.071

5.  Improved scan method for dental imaging using multidetector computed tomography: a phantom study.

Authors:  Yuki Sakai; Kazutoshi Okamura; Erina Kitamoto; Yukiko N Kami; Takashi Shirasaka; Ryoji Mikayama; Masato Tatsumi; Masatoshi Kondo; Toyoyuki Kato; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2020-04-29       Impact factor: 2.419

Review 6.  4D-DSA: Development and Current Neurovascular Applications.

Authors:  K L Ruedinger; S Schafer; M A Speidel; C M Strother
Journal:  AJNR Am J Neuroradiol       Date:  2020-11-26       Impact factor: 3.825

7.  Is MRI a viable alternative to CT/CBCT to identify the course of the inferior alveolar nerve in relation to the roots of the third molars?

Authors:  Florian Beck; Stephanie Austermann; Kristina Bertl; Christian Ulm; Stefan Lettner; Andrea Toelly; André Gahleitner
Journal:  Clin Oral Investig       Date:  2020-12-07       Impact factor: 3.573

8.  Accuracy of Bone Measurements in the Vicinity of Titanium Implants in CBCT Data Sets: A Comparison of Radiological and Histological Findings in Minipigs.

Authors:  Alexander Gröbe; Jan Semmusch; Maximilian Schöllchen; Henning Hanken; Michael Hahn; Wolfgang Eichhorn; Gerhard Schön; Ole Jung; Jamal M Stein; Aline Reitmeier; Max Heiland; Ralf Smeets; Clarissa Precht
Journal:  Biomed Res Int       Date:  2017-07-17       Impact factor: 3.411

9.  3D cone-beam CT with a twin robotic x-ray system in elbow imaging: comparison of image quality to high-resolution multidetector CT.

Authors:  Jan-Peter Grunz; Andreas Max Weng; Andreas Steven Kunz; Maike Veyhl-Wichmann; Rainer Schmitt; Carsten Herbert Gietzen; Lenhard Pennig; Stefan Herz; Süleyman Ergün; Thorsten Alexander Bley; Tobias Gassenmaier
Journal:  Eur Radiol Exp       Date:  2020-09-08

Review 10.  Dose reduction in CT imaging for facial bone trauma in adults: A narrative literature review.

Authors:  Tayla Hooper; Grace Eccles; Talia Milliken; Josephine R Mathieu-Burry; Warren Reed
Journal:  J Med Radiat Sci       Date:  2019-02-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.