Improved scan method for dental imaging using multidetector computed tomography: a phantom study.

Yuki Sakai1, Kazutoshi Okamura2, Erina Kitamoto2, Yukiko N Kami2, Takashi Shirasaka1, Ryoji Mikayama1, Masato Tatsumi1, Masatoshi Kondo1, Toyoyuki Kato1, Kazunori Yoshiura2.   

Abstract

OBJECTIVES: This study aimed to propose an improved scan method to shorten irradiation time and reduce radiation exposure.
METHODS: The maxilla of a human head CT phantom and a Catphan phantom were used for qualitative and quantitative assessment, respectively. The phantoms were scanned by a 160-row multidetector CT scanner using volumetric and helical scanning. In volumetric scanning, the tube current varied from 120 to 60 to 30 to 20 mA with a tube voltage of 120 kV. Images were reconstructed with a bone kernel using iterative reconstruction (IR) and filtered back projection. As a reference protocol, helical scanning was performed using our clinical setting with 120 kV. Two dental radiologists independently graded the quality of dental images using a 4-point scale (4, superior to reference; 1, unacceptable). For the quantitative assessment, we assessed the system performance from each scan.
RESULTS: There was no significant difference between the image quality of volumetric scanning using the 60 mA protocol reconstructed with IR and that of the reference (3.08 and 3.00, p = 0.3388). The system performance values at 1.0 cycles/mm of volumetric scanning and 60 mA protocol reconstructed with IR and reference were 0.0038 and 0.0041, respectively. The effective dose of volumetric scanning using the 60 mA protocol was 51.8 µSv, which is a 64.2% reduction to that of the reference.
CONCLUSIONS: We proposed an improved scan method resulting in a 64.2% reduction of radiation dose with one-fourth of irradiation time by combining volumetric scanning and IR technique in multidetector CT.

Entities:  

Keywords:  dental imaging; dentomaxillofacial CT; image quality; radiation dose

Mesh:

Year:  2020        PMID: 32302213      PMCID: PMC7461741          DOI: 10.1259/dmfr.20190462

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  24 in total

1.  Evaluation of the accuracy of cone beam computed tomography on measuring impacted supernumerary teeth.

Authors:  Levent Demiriz; Ebru Hazar Bodrumlu; Murat İçen; Mustafa Cenk Durmuşlar
Journal:  Scanning       Date:  2016-01-19       Impact factor: 1.932

2.  Radiation dose reduction with Sinogram Affirmed Iterative Reconstruction technique for abdominal computed tomography.

Authors:  Mannudeep K Kalra; Mischa Woisetschläger; Nils Dahlström; Sarabjeet Singh; Maria Lindblom; Garry Choy; Petter Quick; Bernhard Schmidt; Martin Sedlmair; Michael A Blake; Anders Persson
Journal:  J Comput Assist Tomogr       Date:  2012 May-Jun       Impact factor: 1.826

3.  CT iterative reconstruction in image space: a phantom study.

Authors:  C Ghetti; O Ortenzia; G Serreli
Journal:  Phys Med       Date:  2011-04-15       Impact factor: 2.685

4.  Lung nodule detection performance in five observers on computed tomography (CT) with adaptive iterative dose reduction using three-dimensional processing (AIDR 3D) in a Japanese multicenter study: Comparison between ultra-low-dose CT and low-dose CT by receiver-operating characteristic analysis.

Authors:  Yukihiro Nagatani; Masashi Takahashi; Kiyoshi Murata; Mitsuru Ikeda; Tsuneo Yamashiro; Tetsuhiro Miyara; Hisanobu Koyama; Mitsuhiro Koyama; Yukihisa Sato; Hiroshi Moriya; Satoshi Noma; Noriyuki Tomiyama; Yoshiharu Ohno; Sadayuki Murayama
Journal:  Eur J Radiol       Date:  2015-04-02       Impact factor: 3.528

5.  Investigation of an optimized scanning protocol for the dentomaxillofacial region using 320-slice multidetector computed tomography.

Authors:  Wei-Chan Lin; Hsueh-Han Wang; Wen-Lin Hsu; Bi-Hui Cao; De-Ji Chen; Chia-Jung Tsai
Journal:  Dentomaxillofac Radiol       Date:  2017-03-07       Impact factor: 2.419

6.  Performance of cone-beam computed tomography and multidetector computed tomography in diagnostic imaging of the midface: A comparative study on Phantom and cadaver head scans.

Authors:  Simon Veldhoen; Maximilian Schöllchen; H Hanken; C Precht; F O Henes; G Schön; H D Nagel; U Schumacher; M Heiland; G Adam; M Regier
Journal:  Eur Radiol       Date:  2016-05-11       Impact factor: 5.315

7.  Assessment of the variation in CT scanner performance (image quality and Hounsfield units) with scan parameters, for image optimisation in radiotherapy treatment planning.

Authors:  Anne T Davis; Antony L Palmer; Silvia Pani; Andrew Nisbet
Journal:  Phys Med       Date:  2017-12-19       Impact factor: 2.685

Review 8.  Multidetector CT in children: current concepts and dose reduction strategies.

Authors:  Rutger A J Nievelstein; Ingrid M van Dam; Aart J van der Molen
Journal:  Pediatr Radiol       Date:  2010-06-10

9.  Reducing CT radiation dose with iterative reconstruction algorithms: the influence of scan and reconstruction parameters on image quality and CTDIvol.

Authors:  Thorsten Klink; Verena Obmann; Johannes Heverhagen; Alexander Stork; Gerhard Adam; Philipp Begemann
Journal:  Eur J Radiol       Date:  2014-06-05       Impact factor: 3.528

10.  Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study.

Authors:  Mark S Pearce; Jane A Salotti; Mark P Little; Kieran McHugh; Choonsik Lee; Kwang Pyo Kim; Nicola L Howe; Cecile M Ronckers; Preetha Rajaraman; Alan W Sir Craft; Louise Parker; Amy Berrington de González
Journal:  Lancet       Date:  2012-06-07       Impact factor: 79.321

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.