Mohammed H Alam1,2, Taigang He1,2,3, Dominique Auger1,2, Gillian C Smith1,2, Peter Drivas1, Rick Wage1, Cemil Izgi1, Karen Symmonds1, Andreas Greiser4, Bruce S Spottiswoode5, Lisa Anderson3, David Firmin1,2, Dudley J Pennell6,7.
Abstract
BACKGROUND: There is a need for improved worldwide access to tissue iron quantification using T2* cardiovascular magnetic resonance (CMR). One route to facilitate this would be simple in-line T2* analysis widely available on MR scanners. We therefore compared our clinically validated and established T2* method at Royal Brompton Hospital (RBH T2*) against a novel work-in-progress (WIP) sequence with in-line T2* measurement from Siemens (WIP T2*).Entities:
Keywords: CMR; Heart; In-line processing; Iron overload; Liver; T2*
Mesh:
Substances:
Year: 2016 PMID: 27121114 PMCID: PMC4847205 DOI: 10.1186/s12968-016-0243-4
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Fig. 1Typical short axis mid-ventricular slices obtained from a thalassemia patient (left- RBH T2* sequence; middle- WIP sequence; right- corresponding in-line T2* map)
Fig. 2Scatter plot with the line of identity (left) and Bland-Altman plot (right) showing the comparison for the white blood RBH and WIP T2* cardiac sequences. The 95 % confidence intervals are shown as dotted lines on the Bland-Altman plots
Fig. 3Scatter plot with the line of identity (left) and Bland-Altman plot (right) showing the comparison for the black blood RBH and WIP T2* cardiac sequences. The 95 % confidence intervals are shown as dotted lines on the Bland-Altman plots
Fig. 4Scatter plot with the line of identity (left) and Bland-Altman plot (right) showing the comparison for the RBH and WIP T2* liver sequences. The 95 % confidence intervals are shown as dotted lines on the Bland-Altman plots
Reproducibility data
| Sequence | Intraobserver | Interobserver | Interstudy | Artefact score |
|---|---|---|---|---|
| CoV (%) | CoV (%) | CoV (%) | (mean) | |
| RBH/WIP | RBH/WIP | RBH/WIP | RBH/WIP | |
| 1.5 T Heart White blood | 3.2/5.1 | 3.3/10.4 | 5.5/9.3 | 4.25/3.27 |
| 1.5 T Heart Black blood | 1.5/4.1 | 2.8/10.1 | 4.6/7.5 | 4.39/4.24 |
| 1.5 T Liver | 2.0/6.1 | 4.2/16.6 | 5.7/10.2 | 4.06/4.06 |
CoV coefficient of variance, RBH Royal Brompton Hospital, WIP work-in-progress
Fig. 5Example showing that the placing of an ROI in the septum was generally straightforward using the magnitude images (left), but sometimes difficult on the WIP T2* map (right)