| Literature DB >> 17018390 |
Mark A Tanner1, Taigang He, Mark A Westwood, David N Firmin, Dudley J Pennell.
Abstract
The transferability of the T2* technique for measurement of tissue iron between magnetic resonance (MR) scanners is unknown. Heart and liver multi-breath-hold T2* sequences were installed on MR scanners at six different sites. T2* was assessed locally in five or more patients with thalassemia major (n=39), and subjects were re-scanned at the standardization center in London. Inter-center reproducibility of T2* in heart and liver was 5.0% and 7.1%, with mean absolute differences in T2* of 1.3 ms and 0.45 ms, respectively. The MR multi-breath-hold T2* technique for tissue iron quantification is transferable between scanners with good reproducibility.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17018390
Source DB: PubMed Journal: Haematologica ISSN: 0390-6078 Impact factor: 9.941