| Literature DB >> 27087975 |
Catriona Hargrave1, Nicole Mason2, Robyn Guidi2, Julie-Anne Miller2, Jillian Becker2, Matthew Moores3, Kerrie Mengersen3, Michael Poulsen2, Fiona Harden4.
Abstract
INTRODUCTION: Time-consuming manual methods have been required to register cone-beam computed tomography (CBCT) images with plans in the Pinnacle(3) treatment planning system in order to replicate delivered treatments for adaptive radiotherapy. These methods rely on fiducial marker (FM) placement during CBCT acquisition or the image mid-point to localise the image isocentre. A quality assurance study was conducted to validate an automated CBCT-plan registration method utilising the Digital Imaging and Communications in Medicine (DICOM) Structure Set (RS) and Spatial Registration (RE) files created during online image-guided radiotherapy (IGRT).Entities:
Keywords: Adaptive radiotherapy; automated registration; cone‐beam computed tomography; treatment planning system
Mesh:
Year: 2015 PMID: 27087975 PMCID: PMC4775828 DOI: 10.1002/jmrs.141
Source DB: PubMed Journal: J Med Radiat Sci ISSN: 2051-3895
Figure 1Workflow for the phantom QA study.
Departmental planning CT and CBCT scan acquisition protocols
| Planning CT scan acquisition protocols | |
| Pelvis | 120 kVp, 35 mAs, 650 mm FOV, pitch = 0.75, 2 mm slice width |
| Head and neck | 120 kVp, 35 mAs, 650 mm FOV, pitch = 1.2, 3 mm slice width |
| CBCT scan acquisition protocols | |
| Pelvis | 125 kVp, 80 mA, scan mode: half‐fan, image matrix: 512 × 512, filter: sharp, slice thickness: 2 mm |
| Head and neck | 100 kVp, 20 mA, scan mode: Full‐fan, image matrix: 512 × 512, filter: sharp, slice thickness: 2 mm |
The different setup conditions used to acquire the images for the second series of CBCTs acquired on Linac 1
| Couch lateral position (Planiso = linac isocentre) | Setup error applied after Planiso setup | Image match saved | Couch shifts applied | DICOM‐RT RS file created | DICOM‐RT RE files created |
|---|---|---|---|---|---|
| 0 cm | No | No | No | Yes | 0 |
| Yes | No | Yes | 1 | ||
| No | Yes | Yes | 1 | ||
|
|
|
|
| ||
| 0 cm | Yes | No | No | Yes | 0 |
| Yes | No | Yes | 1 | ||
| No | Yes | Yes | 1 | ||
|
|
|
|
| ||
| 8 cm | Yes | No | No | Yes | 0 |
| Yes | No | Yes | 1 | ||
| No | Yes | Yes | 1 | ||
|
|
|
|
|
The rows in bold equate to the OBI workflow for online IGRT with a 0 mm action level.
Figure 2Example of the contour based FM mid‐point localisation, showing from left to right, the mid‐point position of the left FM in the transverse, sagittal and coronal planes.
Point of interest (POI) naming conventions and method of creation
| POI name | Definition | Method of creation |
|---|---|---|
| Plan iso | Treatment plan isocentre = localisation of isocentre on linac prior to online IGRT | The treatment plan isocentre. POI attached to planning CT scan during plan generation |
| CBCTmidpt | Half‐way between coordinates of first and last slice in | POI manually added to the CBCT image once imported to Pinnacle3 that equates to the mid‐point of the image |
| FMmidpt | The mid‐point of the imaged position of the anterior and right and left lateral fiducial markers on the CBCT image | POIs automatically placed in the middle of threshold‐based contours the FMs |
| AcqIsocenter | The linac isocentre during CBCT acquisition | POI automatically created and associated with the CBCT image when DICOM‐RT RS file imported |
| InitLaserIso | The position of the linac isocentre prior to CBCT scan initialisation | POI automatically created and associated with the CBCT image when DICOM‐RT RS file imported |
| InitMatchIso | The position of linac isocentre after CBCT acquisition, just prior to online image matching | POI automatically created and associated with the CBCT image when DICOM‐RT RS file imported |
Figure 3The difference in the RS file created CBCT points when anti‐collision couch re‐positioning has occurred during CBCT acquisition. On each image above the Planiso = the green POI and the CBCT midpt = the purple POI. (a) When no anti‐collision couch repositioning is applied, the DICOM‐RT RS generated POIs have the same coordinates (the orange POI). (b) When anti‐collision couch repositioning occurs, the AcqIsocentre (the light blue POI) = the isocentre during scanning and the InitLaserIso and InitMatchIso POIs = the isocentre position before and after couch repositioning and will have the same coordinates (the orange POI).
The difference between the Pinnacle3 coordinates of the image mid‐point POI and the AcqusitionIso POI for the CBCT images of the phantom
| CBCTmidpt POI – AcqIso POI coordinates (cm): pelvis scanning protocol | CBCTmidpt POI – AcqIso POI coordinates (cm): head and neck CBCT image mid‐point | |||||
|---|---|---|---|---|---|---|
| Linear accelerator | L/R ( | A/P ( | S/I ( | L/R ( | A/P ( | S/I ( |
| Linac 1: series 1 | 0.00 | 0.05 | −0.04 | 0.00 | 0.05 | −0.04 |
| Linac 1: series 2 | 0.00 | 0.05 | −0.04 | |||
| Linac 2 | −0.01 | 0.05 | −0.02 | −0.01 | 0.05 | −0.02 |
| Linac 3 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 |
| Linac 4 | 0.00 | 0.00 | −0.04 | 0.00 | 0.00 | −0.02 |
CBCTmidpt POI, CBCT mid‐point point of interest; AcqIso POI, AcquisitionIso point of interest; L/R, left/right; A/P, anterior/posterior; S/I, superior/inferior.
Figure 4The procedure for automated CBCT‐planning CT registration in Pinnacle3 for replicating delivered treatments after online IGRT with a 0 mm action level.
The mathematical and visual rule‐based system developed for verifying the CBCT isocentre alignment with the treatment plan isocentre
| If the recorded MOSAIQ shifts are | The sign of the value after performing: Coordinatestreatment plan isocentre − CoordinatesInitMatchIso should be: | The InitMatchIso POI should be visually on the transverse, coronal and sagittal 2D views in Pinnacle3 |
|---|---|---|
| Left | +ve (positive) | To the right of the treatment plan isocentre POI |
| Right | −ve (negative) | To the left of the treatment plan isocentre POI |
| Ant | +ve (positive) | To the post of the treatment plan isocentre POI |
| Post | −ve (negative) | To the ant of the treatment plan isocentre POI |
| Sup | −ve (negative) | To the inf of the treatment plan isocentre POI |
| Inf | +ve (positive) | To the sup of the treatment plan isocentre POI |
Figure 5Visual check of CBCT‐plan registration in Pinnacle3 against the MOSAIQ record of the online image match. (a) CBCT‐plan registration in Pinnacle3 produced by the automated method, (b) Recorded online soft‐tissue matching of the bladder in MOSAIQ corresponding to the CBCT registered in Pinnacle3.