| Literature DB >> 27074452 |
Ramamoorthy Ravichandran1, John Pichy Binukumar, Iqbal Al Amri, Cheriyathmanjiyil Antony Davis.
Abstract
Diamond detectors (DD) are preferred in small field dosimetry of radiation beams because of small dose profile penumbras, better spatial resolution, and tissue-equivalent properties. We investigated a commercially available 'microdiamond' detector in realizing absorbed dose from first principles. A microdiamond detector, type TM 60019 with tandem electrometer is used to measure absorbed doses in water, nylon, and PMMA phantoms. With sensitive volume 0.004 mm3, radius 1.1mm, thickness 1 x10(-3) mm, the nominal response is 1 nC/Gy. It is assumed that the diamond detector could collect total electric charge (nC) developed during irradiation at 0 V bias. We found that dose rate effect is less than 0.7% for changing dose rate by 500 MU/min. The reproducibility in obtaining readings with diamond detector is found to be ± 0.17% (1 SD) (n = 11). The measured absorbed doses for 6 MV and 15 MV photons arrived at using mass energy absorption coefficients and stop-ping power ratios compared well with Nd, water calibrated ion chamber measured absorbed doses within 3% in water, PMMA, and nylon media. The calibration factor obtained for diamond detector confirmed response variation is due to sensitivity due to difference in manufacturing process. For electron beams, we had to apply ratio of electron densities of water to carbon. Our results qualify diamond dosimeter as a transfer standard, based on long-term stability and reproducibility. Based on micro-dimensions, we recommend these detectors for pretreatment dose verifications in small field irradiations like stereotactic treatments with image guidance.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27074452 PMCID: PMC5875569 DOI: 10.1120/jacmp.v17i2.5690
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Listed and calculated ratios of mass energy absorption coefficients
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| 6 MV | 1.111 | 1.078 | 1.090 | 0.997 | 1.120 | 1.087 | 1.099 |
| 15 MV | 1.105 | 1.063 | 1.075 | 0.986 | 1.121 | 1.078 | 1.093 |
Listed and calculated average restricted mass stopping power ratios
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
| 6 MV | 1.127 | 1.093 | 1.129 | 1.002 | 1.125 | 1.091 | 1.127 |
| 15 MV | 1.106 | 1.074 | 1.097 | 0.982 | 1.126 | 1.094 | 1.117 |
Figure 1Estimation of absorbed dose in (a) nylon, (b) PMMA, and (c) water phantom.
Electron density in media and ratios of e” /gm against carbon
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|
|
|
|
| ||||
| 3.343 | 3.248 | 3.300 | 3.008 | 1.111 | 1.080 | 1.013 |
Figure 2Geometry of calibration set up for diamond detector.
Linearity of diamond detector (60019) in Perspex phantom at 5 cm depth, 300 MU/min
|
| |||
|---|---|---|---|
|
|
|
|
|
| 25 | 9.112 | 9.114 | 9.113 |
| 50 | 18.140 | 18.120 | 18.130 |
| 75 | 27.120 | 27.120 | 27.120 |
| 100 | 36.150 | 36.090 | 36.120 |
| 150 | 54.140 | 54.050 | 54.100 |
| 200 | 72.040 | 72.020 | 72.030 |
| 250 | 90.020 | 90.000 | 90.010 |
| 300 | 107.900 | 107.900 | 107.900 |
| 400 | 143.900 | 143.900 | 143.900 |
Dose rate effect of microdiamond detector for linac X‐rays
|
|
|
| |||||
|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| 100 | 0.04 | 1.6405 | 1.6300 | ||||
| 200 | 0.04 | 1.6340 | 1.6250 | ||||
| 300 | 0.04 | 1.6310 | 1.6240 |
|
| 1.6692 | 1.6534 |
| 400 | 0.04 | 1.6305 | 1.6230 | ||||
| 500 | 0.04 | 1.6295 | 1.6215 | ||||
| 600 | 0.04 | 1.6295 | 1.6215 | ||||
Field size response of diamond detector linac X‐rays
|
|
|
|
| ||||
|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
| ||
| 6 MV (2300CD) | 1.073 | 1.256 | 1.1705 | 361.3 | 416.2 | 1.1519 | 1.016 |
| 15 MV (2300CD) | 1.226 | 1.358 | 1.1076 | 417.7 | 457.5 | 1.0953 | 1.011 |
| 6 MV (600CD) | 1.047 | 1.196 | 1.1423 | 228.8 | 260.1 | 1.1368 | 1.005 |
Field size factors for mMLC (Brainlab) small fields in 6 MV X‐ray beam. (Perspex phantom, 5 cm depth, comparison of diamond detector CC 01 ion chamber (IC))
|
|
|
|
| ||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
| |||
|
| Diam | 0.5627 | 0.6725 | 0.6756 | 0.6780 | 0.6782 | 0.6791 | 0.6797 |
|
| IC | 0.4682 | 0.5782 | 0.5831 | 0.5845 | 0.5860 | 0.5870 | 0.5878 | 1.65 | |
|
| Diam | 0.7709 | 0.7894 | 0.7976 | 0.7996 | 0.7997 | 0.8016 |
| |
| IC | 0.7582 | 0.7849 | 0.7924 | 0.7961 | 0.7941 | 0.7972 | 0.46 | ||
|
| Diam | 0.8240 | 0.8473 | 0.8523 | 0.8532 | 0.8548 |
| ||
| IC | 0.8204 | 0.8442 | 0.8503 | 0.8506 | 0.8513 | 0.08 | |||
|
| Diam | 0.8703 | 0.8794 | 0.8833 | 0.8847 |
| |||
| IC | 0.8655 | 0.8749 | 0.8765 | 0.8810 | 0.14 | ||||
|
| Diam | 0.8842 | 0.8967 | 0.9028 | 0.9066 |
| |||
| IC | 0.8815 | 0.8958 | 0.9017 | 0.9032 | 0.14 | ||||
|
| Diam | 0.8968 | 0.9113 | 0.9193 | 0.9240 |
| |||
| IC | 0.8957 | 0.9117 | 0.9193 | 0.9233 | 0.07 | ||||
|
| Diam | 0.9057 | 0.9221 | 0.9321 | 0.9377 |
| |||
| IC | 0.9056 | 0.9199 | 0.9316 | 0.9344 | 0.16 | ||||
|
| Diam | 0.9442 | 0.9572 | 0.9660 |
| ||||
| IC | 0.9436 | 0.9567 | 0.9646 | 0.06 | |||||
|
| Diam | 0.9749 | 0.9860 |
| |||||
| IC | 0.9733 | 0.9871 | 0.20 | ||||||
|
| Diam | 1.0000 |
| ||||||
| IC | 1.0000 |
| |||||||
Field size factors for mMLC (Brainlab) effect of phantoms. (Comparison of measured factors in water at dmax and Perspex at 5 cm depth with CC01 IC)
|
|
|
|
| ||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| ||
|
| Water | 0.4042 | 0.5963 | 0.6005 | 0.6018 | 0.6027 | 0.6208 | 0.6220 |
|
| Ppx | 0.4682 | 0.5782 | 0.5831 | 0.5845 | 0.5860 | 0.5870 | 0.5878 | 1.50 | |
|
| Water | 0.7563 | 0.7946 | 0.8069 | 0.8057 | 0.8100 | 0.8110 |
| |
| Ppx | 0.7582 | 0.7849 | 0.7924 | 0.7961 | 0.7941 | 0.7972 | 0.81 | ||
|
| Water | 0.8469 | 0.8583 | 0.8621 | 0.8639 | 0.8654 |
| ||
| Ppx | 0.8204 | 0.8442 | 0.8503 | 0.8506 | 0.8513 | 0.74 | |||
|
| Water | 0.8784 | 0.8862 | 0.8902 | 0.8936 |
| |||
| Ppx | 0.8655 | 0.8749 | 0.8765 | 0.8810 | 0.11 | ||||
|
| Water | 0.8944 | 0.9059 | 0.9128 | 0.9164 |
| |||
| Ppx | 0.8815 | 0.8958 | 0.9017 | 0.9032 | 0.17 | ||||
|
| Water | 0.9039 | 0.9192 | 0.9274 | 0.9332 |
| |||
| Ppx | 0.8957 | 0.9117 | 0.9193 | 0.9233 | 0.16 | ||||
|
| Water | 0.9117 | 0.9286 | 0.9386 | 0.9453 |
| |||
| Ppx | 0.9056 | 0.9199 | 0.9316 | 0.9344 | 0.24 | ||||
|
| Water | 0.9472 | 0.9623 | 0.9723 |
| ||||
| Ppx | 0.9436 | 0.9567 | 0.9646 | 0.21 | |||||
|
| Water | 0.9760 | 0.9905 |
| |||||
| Ppx | 0.9733 | 0.9871 | 0.34 | ||||||
|
| Water | 1.0000 |
| ||||||
| Ppx | 1.0000 |
| |||||||
Measurements in water phantom with photons
|
| ||||
|---|---|---|---|---|
|
|
|
|
|
|
| 1 | 6 MV | 0.6932 | 0.6823 |
|
| 15 MV | 0.7872 | 0.7833 |
| |
| 2300 CD | ||||
| 2 | 6 MV | 0.6932 | 0.6722 |
|
| 15 MV | 0.7856 | 0.7737 |
| |
| 2300 CD | ||||
| 3 | 6 MV | 0.6694 | 0.6558 |
|
| 600 CD | ||||
Calibration factors for ‘diamond detector’ against ion chamber ND,water
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
| ||
|
|
|
|
|
|
| |
| 6 MV Clinac 2300 | 1)0.6932 | 1)0.6823 | 1.0160 |
|
| 1.0142 |
| 2)0.6932 | 2)0.6722 | 1.0312 |
|
| ||
| 15 MV Clinac 2300 | 1)0.7872 | 2)0.7833 | 1.0050 |
|
| 0.9963 |
| 2)0.7873 | 2)0.7737 | 1.0175 |
|
| ||
| 6 MV | 1)0.6694 | 1)0.6558 | 1.0207 | 1) |
| 1.0186 |
|
|
| |||||
| Clinac 600 | 2) |
| 1.0191 | |||
|
|
| |||||
Measurements in Perspex tool in linac X‐rays
| Diamond Detector Dose | ||||
|---|---|---|---|---|
|
|
|
|
|
|
| 1 | 600 CD | 0.8704 | 0.8557 |
|
| 6 MV | FC 65) | |||
| 2 | 2300 CD | |||
| 6 MV | 0.8948 | 0.8816 |
| |
| 15 MV | 0.9632 | 0.9835 |
| |
| (FC 65) | ||||
| 3 | 2300 CD | |||
| 6 MV | 0.8946 | 0.8806 |
| |
| 15 MV | 0.9624 | 0.9639 |
| |
| 4 | 2300 CD | (CC13) | ||
| 6 MV | 0.8956 | 0.8757 |
| |
| 15 MV | 0.9623 | 0.9597 |
| |
| (CC13) | ||||
| 5 | 600 CD | |||
| 6 MV | 0.8683 | 0.8510 |
| |
| (CC13) | ||||
Measurements in nylon miniphantom in linac X‐rays
|
| ||||
|---|---|---|---|---|
|
|
|
|
|
|
| 1 | 6 MV | 1.0003 | 0.9836 |
|
| 15 MV 2300 CD | 1.0482 | 1.0460 |
| |
| 2 | 6 MV | 1.0210 | 0.9870 |
|
| 15 MV 2300 CD | 1.0625 | 1.0413 |
| |
| 3 | 6 MV 600 CD | 0.9675 | 0.9493 |
|
Measurements in water phantom with high‐energy electrons
|
|
|
| |||||
|---|---|---|---|---|---|---|---|
|
|
|
| |||||
|
|
|
|
|
|
|
| |
| 1 | 6 MeV/ 14.0 | 1.0814 | 1.0875 | 1.0056 | 0.9799 |
|
|
| 2 | 9 MeV/ 21.2 | 1.0290 | 1.0343 | 0.9909 | 0.9656 |
|
|
| 3 | 12 eV/29.4 | 1.0438 | 1.0485 | 1.0155 | 0.9990 |
|
|
| 4 | 15 eV/37.3 | 1.1279 | 1.1351 | 1.0968 | 1.0955 |
|
|
| 5 | 18 eV/45.0 | 1.0701 | 1.0777 | 1.0498 | 1.0502 |
|
|
| 6 | 22 eV/52.2 | 0.9992 | 1.0128 | 0.9896 | 0.9965 |
|
|