Comparison of pulse oximeters: effects of vasoconstriction and venous engorgement.

C J Wilkins1, M Moores, C D Hanning.   

Abstract

The effects of cold-induced vasoconstriction and venous occlusion on the detection of induced hypoxaemia by four pulse oximeters were examined in 10 volunteers. In three further subjects vasoconstriction was maintained until at least one instrument failed to detect the induced hypoxaemia. Time taken to detect hypoxaemia was increased for all instruments to between two and three times the instrument's own control value for both vasoconstriction and venous engorgement (P less than 0.01). There was highly significant variation in detected minimum saturation between the instruments (P less than 0.001). One instrument failed to detect the full extent of desaturation under the experimental conditions and was more likely to fail completely to detect desaturation than the other test instruments when influenced by vasoconstriction (P less than 0.05). Significant impairment in the performance of all the instruments tested occurred in the presence of normal pulse signals. The duration of detected reductions in oxygen saturation was not significantly affected.

Entities:  

Mesh:

Year:  1989        PMID: 2706182     DOI: 10.1093/bja/62.4.439

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  8 in total

1.  A comparison of pulse oximetry and near infrared spectroscopy (NIRS) in the detection of hypoxaemia occurring with pauses in nasal airflow in neonates.

Authors:  S L Watkin; S A Spencer; P W Dimmock; Y A Wickramasinghe; P Rolfe
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

2.  Reflection and transmission pulse oximetry during compromised peripheral perfusion.

Authors:  H Pälve
Journal:  J Clin Monit       Date:  1992-01

3.  Decreased circulation time in the upper limb reduces the lag time of the finger pulse oximeter response.

Authors:  Z N Ding; K Shibata; K Yamamoto; T Kobayashi; S Murakami
Journal:  Can J Anaesth       Date:  1992-01       Impact factor: 5.063

4.  Misleading information from pulse oximetry and the usefulness of continuous blood gas monitoring in a post cardiac surgery patient.

Authors:  J P Benson; B Venkatesh; V Patla
Journal:  Intensive Care Med       Date:  1995-05       Impact factor: 17.440

5.  The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy.

Authors:  J Ibáñez; J Velasco; J M Raurich
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

6.  Evaluation of finger and forehead pulse oximeters during mild hypothermic cardiopulmonary bypass.

Authors:  Ken Yamaura; Kazuo Irita; Tadashi Kandabashi; Kohei Tohyama; Shosuke Takahashi
Journal:  J Clin Monit Comput       Date:  2007-06-20       Impact factor: 1.977

7.  Photoplethysmography for blood volumes and oxygenation changes during intermittent vascular occlusions.

Authors:  T Y Abay; P A Kyriacou
Journal:  J Clin Monit Comput       Date:  2017-05-25       Impact factor: 2.502

8.  Pulse oximeter oxygen saturation in prediction of arterial oxygen saturation in liver transplant candidates.

Authors:  Seiyed Mohammad Ali Ghayumi; Abolfazl Khalafi-Nezhad; Zahra Jowkar
Journal:  Hepat Mon       Date:  2014-04-07       Impact factor: 0.660

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.