Cécile Chaix1, Jared W Moore1, Roel Van Holen2, Harrison H Barrett1, Lars R Furenlid1.
Abstract
In this paper, we present the imaging aperture of an adaptive SPECT imaging system being developed at the Center for Gamma Ray Imaging (AdaptiSPECT). AdaptiSPECT is designed to automatically change its configuration in response to preliminary data, in order to improve image quality for a particular task. In a traditional pinhole SPECT imaging system, the characteristics (magnification, resolution, field of view) are set by the geometry of the system, and any modification can be accomplished only by manually changing the collimator and the distance of the detector to the center of the field of view. Optimization of the imaging system for a specific task on a specific individual is therefore difficult. In an adaptive SPECT imaging system, on the other hand, the configuration can be conveniently changed under computer control. A key component of an adaptive SPECT system is its aperture. In this paper, we present the design, specifications, and fabrication of the adaptive pinhole aperture that will be used for AdaptiSPECT, as well as the controls that enable autonomous adaptation.Entities:
Keywords: Imaging; SPECT; collimator design; instrumentation; small-animal
Year: 2012 PMID: 27019577 PMCID: PMC4807117 DOI: 10.1109/NSSMIC.2012.6551816