Literature DB >> 26987898

Achieving fast timing performance with multiplexed SiPMs.

M F Bieniosek1, J W Cates, C S Levin.   

Abstract

Using time of flight (ToF) measurements for positron emission tomography (PET) is an attractive avenue for increasing the signal to noise (SNR) ratio of PET images. However, achieving excellent time resolution required for high SNR gain using silicon photomultipliers (SiPM) requires many resource heavy high bandwidth readout channels. A method of multiplexing many SiPM signals into a single electronic channel would greatly simplify ToF PET systems. However, multiplexing SiPMs degrades time resolution because of added dark counts and signal shaping. In this work the relative contribution of dark counts and signal shaping to timing degradation is simulated and a baseline correction technique to mitigate the effect of multiplexing on the time resolution of analog SiPMs is simulated and experimentally verified. A charge sharing network for multiplexing is proposed and tested. Results show a full width at half maximum (FWHM) coincidence time resolution of [Formula: see text] ps for a single 3 mm  ×  3 mm  ×  20 mm LYSO scintillation crystals coupled to an array of sixteen 3 mm  ×  3 mm SiPMs that are multiplexed to a single timing channel (in addition to 4 position channels). A [Formula: see text] array of 3 mm  ×  3 mm  ×  20 mm LFS crystals showed an average FWHM coincidence time resolution of [Formula: see text] ps using the same timing scheme. All experiments were performed at room temperature with no thermal regulation. These results show that excellent time resolution for ToF can be achieved with a highly multiplexed analog SiPM readout.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26987898     DOI: 10.1088/0031-9155/61/7/2879

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Highly multiplexed signal readout for a time-of-flight positron emission tomography detector based on silicon photomultipliers.

Authors:  Joshua W Cates; Matthew F Bieniosek; Craig S Levin
Journal:  J Med Imaging (Bellingham)       Date:  2017-03-23

2.  Analog filtering methods improve leading edge timing performance of multiplexed SiPMs.

Authors:  M F Bieniosek; J W Cates; A M Grant; C S Levin
Journal:  Phys Med Biol       Date:  2016-08-03       Impact factor: 3.609

3.  A multiplexed TOF and DOI capable PET detector using a binary position sensitive network.

Authors:  M F Bieniosek; J W Cates; C S Levin
Journal:  Phys Med Biol       Date:  2016-10-14       Impact factor: 3.609

4.  Evaluation of a clinical TOF-PET detector design that achieves ⩽100 ps coincidence time resolution.

Authors:  Joshua W Cates; Craig S Levin
Journal:  Phys Med Biol       Date:  2018-06-07       Impact factor: 3.609

5.  Time-over-threshold for pulse shape discrimination in a time-of-flight phoswich PET detector.

Authors:  Chen-Ming Chang; Joshua W Cates; Craig S Levin
Journal:  Phys Med Biol       Date:  2016-12-17       Impact factor: 3.609

6.  Multiplexing Readout for Time-of-Flight (TOF) PET Detectors Using Striplines.

Authors:  Heejong Kim; Chien-Min Kao; Yuexuan Hua; Qingguo Xie; Chin-Tu Chen
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-01-13
  6 in total