The air matters - sleeve air cavity as a marker guiding image-guided helical tomotherapy to target cervical cancer.

Ya-Ting Jan1, Chih-Long Chang2, Hung-Chi Tai3, Yu-Chuen Huang4, Chia-Ling Liao3, Yu-Jen Chen3.   

Abstract

PURPOSE: Radiotherapy with concurrent chemotherapy has been recommended as standard treatment for locally advanced cervical cancer. To validate the main tumor location before each high-precision helical tomotherapy (HT) fraction, the development of a more reliable marker or indicator is of clinical importance to avoid inadequate coverage of the main tumor.
MATERIAL AND METHODS: A 61-year-old woman with cervical cancer, TMN stage cT2b2N1M1, FIGO stage IVB was presented. Extended field external beam radiotherapy (EBRT) with concurrent chemotherapy and the interdigitated delivery of intracavitary brachytherapy was performed. Helical tomotherapy equipped with megavoltage cone beam computed tomography (MV-CBCT) was used for image-guided radiotherapy. For the insertion of tandem of brachytherapy applicator, a silicone sleeve with a central hollow canal was placed into the endocervical canal with the caudal end stopping at the outer surface of the cervical os, and making contact with the distal boundary of the cervical tumor during the entire brachytherapy course.
RESULTS: In the remaining EBRT fractions, we found that the air cavity inside the central hollow canal of the sleeve could be clearly identified in daily CBCT images. The radiation oncologists matched the bony markers to adjust the daily setup errors because the megavoltage of the CBCT images could not provide a precise boundary between the soft tissue and the tumor, but the sleeve air cavity, with a clear boundary, could be used as a surrogate and reliable marker to guide the daily setup errors, and to demonstrate the primary tumor location before delivery of each HT fraction.
CONCLUSIONS: The application of the sleeve during the interdigitated course of HT and brachytherapy in this patient provided information for the feasibility of using the sleeve air cavity as a surrogate marker for the localization of the main primary tumor before the daily delivery of image-guided HT.

Entities:  

Keywords:  brachytherapy; cervical cancer; sleeve air; tomotherapy

Year:  2016        PMID: 26985201      PMCID: PMC4793067          DOI: 10.5114/jcb.2016.57803

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

For locally advanced cervical cancer, concurrent chemotherapy and radiation therapy (CCRT) has been recommended as a standard treatment [1, 2]. However, radiotherapy (RT) complications remain as clinical issues to be improved because of the vicinity of bone marrow and the small bowel, rectum, cervix uteri, and urinary bladder within RT fields [3, 4]. Helical tomotherapy (HT), an advanced RT technique providing rotational beam delivery, has been shown to improve target conformality and to spare more normal tissue in comparison with intensity-modulated radiotherapy (IMRT) [5, 6]. For the accurate delivery of RT to the target volumes, kilovoltage (kV) cone beam computed tomography (CBCT) scanners were integrated into IMRT machines to perform image-guided radiotherapy (IGRT). The energy of CBCT scanners equipped on HT machines is in megavolts (MV), which does not have the same performance characteristics as diagnostic computed tomography (CT) scanners and kV-CBCT because of the lower contrast resolution [7]. The inferior resolution of MV-CBCT images without contrast medium enhancement usually results in an insufficient ability to differentiate tumor and surrounding tissues during the daily verification of patient positioning. The combination of variable factors, including tumor shrinkage, anatomical alteration, body contour change, and daily setup errors, renders the registration of initial planning CT to CBCT images difficult. In clinical practice, radiation therapists commonly need to manually adjust the registered images according to bony markers, the most identifiable structure in CBCT, to fit the initial planning target volumes. To validate the main tumor location before each high-precision HT fraction, the development of a more reliable marker or indicator is of clinical importance to avoid inadequate coverage of the main tumor. Here, we present a case of locally advanced cervical cancer that received HT with the interdigitated combination of intracavitary brachytherapy. For the insertion of tandem of brachytherapy applicator, a silicone sleeve was placed into the endocervix before the beginning of brachytherapy. We found that the air cavity in the sleeve had a clear boundary and could serve as a reliable indicator for locating the main primary tumor during the daily verification of patient positioning in HT practice.

Case report

A 61-year-old woman presented with post-menopause vaginal bleeding was diagnosed as having cervical squamous cell carcinoma by biopsy with parametrial involvement and paraaortic lymphadenopathy noted in CT images (Figure 1A), FIGO stage cT2b2N1M1, IVB. External beam radiotherapy (EBRT) with concurrent chemotherapy, and the interdigitated delivery of intracavitary brachytherapy (high-dose-rate by using 192Ir) after 34 Gy of EBRT was performed from December 2015 to January 2016. External beam radiotherapy was planned for 60 Gy for the main tumor and parametrium as well as 50 Gy to the pelvic and paraaaortic lymph node areas in 30 fractions using the simultaneously integrated boost technique. Helical tomotherapy (Hi Art II with dynamic jaws) in conjunction with MV-CBCT was used to administer EBRT in an image-guided mode. Brachytherapy was designed for 30 Gy in 6 fractions once a week during the EBRT course and twice a week after the ending of EBRT. The dose of brachytherapy was set 100% at point A. For external beam treatment, the dose to 95% volume of PTV for 50 Gy and 60 Gy was achieved. Mean doses to surrounding normal organs were as follows: 53.0 Gy for rectum, 46.0 Gy for urinary bladder, 24.4 Gy for small bowel, 25.6 Gy for left femoral head, and 26.2 Gy for right femoral head. External beam treatment was planned by using CT-based simulation (Figure 2A) and the dose volume histogram of tomotherapy planning was demonstrated in Figure 2D. For insertion of tandem of brachytherapy applicator, a silicone sleeve (manufactured by Fortune Medical Instrument Corp., New Taipei City, Taiwan) with a central hollow canal (Figure 3A) was used. It was placed into the endocervical canal with the caudal end stopping at the outer surface of the cervical os, and making contact with the distal boundary of the cervical tumor during the entire brachytherapy course. Subsequently, a planned kilovoltage CT scan was performed for visualization and validation of the sleeve insertion (Figure 3B–3D).
Fig. 1

Diagnostic computed tomography images for primary cervical cancer and lymphadenopathy. A) Before concurrent chemotherapy and radiation therapy (CCRT) treatment, B) after CCRT treatment. Upper panel: renal vein level, arrow indicates enlarged paraaortic lymph node; middle panel: external iliac vein level, arrow indicates enlarged right pelvic lymph node; lower panel: acetabulum level, arrow indicates cervical tumor mass

Fig. 2

Image registration of initial planning and cone beam computed tomography (CT). A) Initial planning CT images registered with cone beam CT images. B) Image registration without sleeve insertion. Upper left panel: cross section. Upper right panel: sagittal section. Lower left panel: coronal view. Lower right panel: coronal section with isodose lines. Image registration of initial planning and cone beam computed tomography (CT). C) Image registration with sleeve insertion. Upper left panel: cross section. Upper right panel: sagittal section. Lower left panel: coronal view. Lower right panel: coronal section with isodose lines. Grey scale: initial planning CT image; Blue scale: cone beam CT image; Arrow: sleeve air cavity. D) Dose volume histogram of tomotherapy planning

Fig. 3

Demonstration of tandem insertion through sleeve and kilovoltage computed tomography scan images after sleeve insertion. A) White apparatus indicates sleeve, B) oblique coronal view, C) axial view, D) sagittal view

Diagnostic computed tomography images for primary cervical cancer and lymphadenopathy. A) Before concurrent chemotherapy and radiation therapy (CCRT) treatment, B) after CCRT treatment. Upper panel: renal vein level, arrow indicates enlarged paraaortic lymph node; middle panel: external iliac vein level, arrow indicates enlarged right pelvic lymph node; lower panel: acetabulum level, arrow indicates cervical tumor mass Image registration of initial planning and cone beam computed tomography (CT). A) Initial planning CT images registered with cone beam CT images. B) Image registration without sleeve insertion. Upper left panel: cross section. Upper right panel: sagittal section. Lower left panel: coronal view. Lower right panel: coronal section with isodose lines. Image registration of initial planning and cone beam computed tomography (CT). C) Image registration with sleeve insertion. Upper left panel: cross section. Upper right panel: sagittal section. Lower left panel: coronal view. Lower right panel: coronal section with isodose lines. Grey scale: initial planning CT image; Blue scale: cone beam CT image; Arrow: sleeve air cavity. D) Dose volume histogram of tomotherapy planning Demonstration of tandem insertion through sleeve and kilovoltage computed tomography scan images after sleeve insertion. A) White apparatus indicates sleeve, B) oblique coronal view, C) axial view, D) sagittal view

Results

In the remaining EBRT fractions overlapping the brachytherapy course, we found that the air cavity inside the central hollow canal of the sleeve could be clearly identified in the daily MV-CBCT images (Figure 2C). The radiation oncologists matched the bony markers to adjust the daily setup errors because the MV-CBCT images could not provide a precise boundary for the soft tissue or the tumor, only for the bone (Figure 2B). Moreover, the main structure of sleeve was not visible in MV-CBCT images due to lower contrast resolution. The visualization of sleeve by kilovoltage planning CT scan could verify the limitation of MV-CBCT in this scenario. However, the sleeve air cavity had a clear boundary and could be used as a surrogate and reliable marker to guide the daily setup errors, and to demonstrate the primary tumor location before the delivery of each HT fraction. After the completion of the HT and brachytherapy course, the tumor and pelvic and paraaortic lymphadenopathy responded to CCRT (Figure 1B) with acceptable toxicity. No specific complications were noted related to the insertion of the sleeve for 11 months.

Discussion

In summary, the application of the sleeve during the interdigitated course of HT and brachytherapy in this patient provided information for the feasibility of using the sleeve air cavity as a surrogate marker for the localization of the main primary tumor before the daily delivery of image-guided HT. This result may suggest the desirability of designing a clinical proposal for early integration of the sleeve during the HT course for a more precise target localization and validation. The possible limitation of using the sleeve air cavity as an indicator for the main tumor location remains to be carefully considered. For example, the insertion of the sleeve early in the EBRT course might be difficult due to the anatomical alteration caused by the bulky tumor at the cervical os [8, 9]. The relatively long term of insertion of the sleeve may increase the pain scale, vaginal edema, and the risk of infection [10]. The unexpected scattering of the radiation dose from the sleeve needs further dose measurement in the phantom, especially when the scatter correction of CBCT is concerned [11]. Taken together, the early use of the sleeve during the EBRT course, instead of the insertion of the sleeve in a standard interdigitated brachytherapy course, should be performed with caution and could be validated by further clinical trials. Concerning the possible application of magnetic resonance imaging (MRI) simulation in the future [12, 13], the scattering and artifacts surrounding the sleeve might be an issue that should be addressed. To enhance the applicability in both CT- and MRI-based simulations, the modification of the sleeve component could be expedited in collaboration with material scientists [14]. The application of HT in the treatment of cervical cancer is on the increase due to the concerns of normal tissue toxicity and quality of life. Therefore, the efforts to further augment precision in the daily delivery of EBRT by HT is a crucial issue to be addressed. Previous use of metal clips to indicate tumor at our center was not satisfactory, as the clips were not clearly visible on the CBCT images of the HT machine. The adoption of the sleeve air cavity as a marker guiding HT is the first in the literature, so the clinical validation for this issue is therefore warranted. Regardless of using tomotherapy or IMRT, using sleeve may help image guidance during CBCT.

Conclusions

The application of the sleeve during the interdigitated course of HT and brachytherapy in this patient provided information for the feasibility of using the sleeve air cavity as a surrogate marker for the localization of the main primary tumor before the daily delivery of image-guided HT. Further clinical validation for this issue is however warranted.
  14 in total

1.  Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit.

Authors:  Sanford L Meeks; Joseph F Harmon; Katja M Langen; Twyla R Willoughby; Thomas H Wagner; Patrick A Kupelian
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

2.  Efficient scatter distribution estimation and correction in CBCT using concurrent Monte Carlo fitting.

Authors:  G J Bootsma; F Verhaegen; D A Jaffray
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

3.  Concurrent cisplatin-based radiotherapy and chemotherapy for locally advanced cervical cancer.

Authors:  P G Rose; B N Bundy; E B Watkins; J T Thigpen; G Deppe; M A Maiman; D L Clarke-Pearson; S Insalaco
Journal:  N Engl J Med       Date:  1999-04-15       Impact factor: 91.245

4.  Intrauterine high dose rate afterloading brachytherapy: experience of fractionated therapy using a cervical sleeve technique.

Authors:  L K Tyrie; P J Hoskin
Journal:  Clin Oncol (R Coll Radiol)       Date:  1996       Impact factor: 4.126

5.  MRI-based treatment planning for radiotherapy: dosimetric verification for prostate IMRT.

Authors:  Lili Chen; Robert A Price; Lu Wang; Jinsheng Li; Lihong Qin; Shawn McNeeley; C-M Charlie Ma; Gary M Freedman; Alan Pollack
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-10-01       Impact factor: 7.038

6.  Integral dose in three-dimensional conformal radiotherapy, intensity-modulated radiotherapy and helical tomotherapy.

Authors:  R Yang; S Xu; W Jiang; C Xie; J Wang
Journal:  Clin Oncol (R Coll Radiol)       Date:  2009-08-27       Impact factor: 4.126

7.  Cisplatin, radiation, and adjuvant hysterectomy compared with radiation and adjuvant hysterectomy for bulky stage IB cervical carcinoma.

Authors:  H M Keys; B N Bundy; F B Stehman; L I Muderspach; W E Chafe; C L Suggs; J L Walker; D Gersell
Journal:  N Engl J Med       Date:  1999-04-15       Impact factor: 91.245

8.  Dosimetric comparisons of three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, and helical tomotherapy in whole abdominopelvic radiotherapy for gynecologic malignancy.

Authors:  Yong Bae Kim; Joo Ho Kim; Kyung Keun Jeong; Jinsil Seong; Chang Ok Suh; Gwi Eon Kim
Journal:  Technol Cancer Res Treat       Date:  2009-10

9.  Dosimetric comparison of bone marrow-sparing intensity-modulated radiotherapy versus conventional techniques for treatment of cervical cancer.

Authors:  Loren K Mell; Hanifi Tiryaki; Kang-Hyun Ahn; Arno J Mundt; John C Roeske; Bulent Aydogan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-08-01       Impact factor: 7.038

10.  Image guided adaptive brachytherapy for cervical cancer: dose contribution to involved pelvic nodes in two cancer centers.

Authors:  Willemien van den Bos; Sushil Beriwal; Laura Velema; Astrid A C de Leeuw; Christel N Nomden; Ina-M Jürgenliemk-Schulz
Journal:  J Contemp Brachytherapy       Date:  2014-04-03
View more
  4 in total

Review 1.  Problems and solutions in IGRT for cervical cancer.

Authors:  Iván Ríos; Ilse Vásquez; Elsa Cuervo; Óscar Garzón; Johnny Burbano
Journal:  Rep Pract Oncol Radiother       Date:  2018-05-26

2.  Research on Improving Radiotherapy Accuracy Based on Image-Guided Radiotherapy.

Authors:  Shunping Huang; Yang Xiao; Heng Li; Daochang Li
Journal:  Contrast Media Mol Imaging       Date:  2022-08-09       Impact factor: 3.009

3.  Proton Arc Therapy vs Interstitial HDR Brachytherapy in Gynecologic Cancer with Parametrial/pelvic Side Wall Extension.

Authors:  ByongYong Yi; Sina Mossahebi; Arezoo Modiri; Elizabeth M Nichols; Mariana Guerrero; Narottam Lamichhane; Pranshu Mohindra
Journal:  Int J Part Ther       Date:  2022-06-28

4.  Dosimetric effects of the Smit sleeve on high-dose-rate brachytherapy tandem and ovoids plans for patients with locally advanced cervical cancer.

Authors:  Shahil Mehta; Benjamin Farnia; Alberto de la Zerda; Robabeh Rahimi; Aaron Wolfson; Lorraine Portelance
Journal:  J Contemp Brachytherapy       Date:  2019-12-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.