Literature DB >> 25563247

Efficient scatter distribution estimation and correction in CBCT using concurrent Monte Carlo fitting.

G J Bootsma1, F Verhaegen2, D A Jaffray3.   

Abstract

PURPOSE: X-ray scatter is a significant impediment to image quality improvements in cone-beam CT (CBCT). The authors present and demonstrate a novel scatter correction algorithm using a scatter estimation method that simultaneously combines multiple Monte Carlo (MC) CBCT simulations through the use of a concurrently evaluated fitting function, referred to as concurrent MC fitting (CMCF).
METHODS: The CMCF method uses concurrently run MC CBCT scatter projection simulations that are a subset of the projection angles used in the projection set, P, to be corrected. The scattered photons reaching the detector in each MC simulation are simultaneously aggregated by an algorithm which computes the scatter detector response, SMC. SMC is fit to a function, SF, and if the fit of SF is within a specified goodness of fit (GOF), the simulations are terminated. The fit, SF, is then used to interpolate the scatter distribution over all pixel locations for every projection angle in the set P. The CMCF algorithm was tested using a frequency limited sum of sines and cosines as the fitting function on both simulated and measured data. The simulated data consisted of an anthropomorphic head and a pelvis phantom created from CT data, simulated with and without the use of a compensator. The measured data were a pelvis scan of a phantom and patient taken on an Elekta Synergy platform. The simulated data were used to evaluate various GOF metrics as well as determine a suitable fitness value. The simulated data were also used to quantitatively evaluate the image quality improvements provided by the CMCF method. A qualitative analysis was performed on the measured data by comparing the CMCF scatter corrected reconstruction to the original uncorrected and corrected by a constant scatter correction reconstruction, as well as a reconstruction created using a set of projections taken with a small cone angle.
RESULTS: Pearson's correlation, r, proved to be a suitable GOF metric with strong correlation with the actual error of the scatter fit, SF. Fitting the scatter distribution to a limited sum of sine and cosine functions using a low-pass filtered fast Fourier transform provided a computationally efficient and accurate fit. The CMCF algorithm reduces the number of photon histories required by over four orders of magnitude. The simulated experiments showed that using a compensator reduced the computational time by a factor between 1.5 and 1.75. The scatter estimates for the simulated and measured data were computed between 35-93 s and 114-122 s, respectively, using 16 Intel Xeon cores (3.0 GHz). The CMCF scatter correction improved the contrast-to-noise ratio by 10%-50% and reduced the reconstruction error to under 3% for the simulated phantoms.
CONCLUSIONS: The novel CMCF algorithm significantly reduces the computation time required to estimate the scatter distribution by reducing the statistical noise in the MC scatter estimate and limiting the number of projection angles that must be simulated. Using the scatter estimate provided by the CMCF algorithm to correct both simulated and real projection data showed improved reconstruction image quality.

Entities:  

Mesh:

Year:  2015        PMID: 25563247     DOI: 10.1118/1.4903260

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Monte Carlo study of the effects of system geometry and antiscatter grids on cone-beam CT scatter distributions.

Authors:  A Sisniega; W Zbijewski; A Badal; I S Kyprianou; J W Stayman; J J Vaquero; J H Siewerdsen
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

2.  A model-based scatter artifacts correction for cone beam CT.

Authors:  Wei Zhao; Don Vernekohl; Jun Zhu; Luyao Wang; Lei Xing
Journal:  Med Phys       Date:  2016-04       Impact factor: 4.071

3.  Paired cycle-GAN-based image correction for quantitative cone-beam computed tomography.

Authors:  Joseph Harms; Yang Lei; Tonghe Wang; Rongxiao Zhang; Jun Zhou; Xiangyang Tang; Walter J Curran; Tian Liu; Xiaofeng Yang
Journal:  Med Phys       Date:  2019-07-17       Impact factor: 4.071

4.  Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction.

Authors:  Xi Chen; Luo Ouyang; Hao Yan; Xun Jia; Bin Li; Qingwen Lyu; You Zhang; Jing Wang
Journal:  Med Phys       Date:  2017-09       Impact factor: 4.071

5.  Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter - Part I: Core algorithms and validation.

Authors:  Alexander Maslowski; Adam Wang; Mingshan Sun; Todd Wareing; Ian Davis; Josh Star-Lack
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

6.  X-ray scatter correction for dedicated cone beam breast CT using a forward-projection model.

Authors:  Linxi Shi; Srinivasan Vedantham; Andrew Karellas; Lei Zhu
Journal:  Med Phys       Date:  2017-04-25       Impact factor: 4.071

7.  Artificial Intelligence in Radiation Therapy.

Authors:  Yabo Fu; Hao Zhang; Eric D Morris; Carri K Glide-Hurst; Suraj Pai; Alberto Traverso; Leonard Wee; Ibrahim Hadzic; Per-Ivar Lønne; Chenyang Shen; Tian Liu; Xiaofeng Yang
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-08-24

8.  The role of off-focus radiation in scatter correction for dedicated cone beam breast CT.

Authors:  Linxi Shi; Srinivasan Vedantham; Andrew Karellas; Lei Zhu
Journal:  Med Phys       Date:  2017-12-16       Impact factor: 4.071

Review 9.  Modelling the physics in the iterative reconstruction for transmission computed tomography.

Authors:  Johan Nuyts; Bruno De Man; Jeffrey A Fessler; Wojciech Zbijewski; Freek J Beekman
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

10.  CBCT-based synthetic CT generation using deep-attention cycleGAN for pancreatic adaptive radiotherapy.

Authors:  Yingzi Liu; Yang Lei; Tonghe Wang; Yabo Fu; Xiangyang Tang; Walter J Curran; Tian Liu; Pretesh Patel; Xiaofeng Yang
Journal:  Med Phys       Date:  2020-03-28       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.