Jiayuan Peng1,2, Zhen Zhang1,2, Jiazhou Wang1,2, Jiang Xie1,2, Weigang Hu1,2.
Abstract
PURPOSE: 4DCT delineated internal target volume (ITV) was applied to determine the tumor motion and used as planning target in treatment planning in lung cancer stereotactic body radiotherapy (SBRT). This work is to study the accuracy of using ITV to predict the real target dose in lung cancer SBRT.Entities:
Keywords: 4DCT; ITV; SBRT; dose evaluation; lung cancer
Mesh:
Year: 2016 PMID: 26968812 PMCID: PMC5008378 DOI: 10.18632/oncotarget.8000
Source DB: PubMed Journal: Oncotarget ISSN: 1949-2553
Figure 1AVG CT of a respiratory phantom case with ITV contoured in MIP (red contour), chestwall (brown contour), and lung (between tumor and chestwall)
Figure 2An example of field distribution in SBRT planning in patient case
Dose and target coverage variations between ITV and GTV_4D (real target) in different target size and target motion
| Target Size (cm) | Target Motion (cm) | |||
|---|---|---|---|---|
| 0.5 | 1 | 2 | 3 | |
| 1.00 | −1.56 | −2.87 | −8.03 | −8.03 |
| 2.00 | −0.17 | −0.08 | −1.31 | −1.31 |
| 3.00 | −0.03 | −0.08 | 0.22 | 0.22 |
| 4.00 | 0.01 | 0.10 | 0.10 | 0.10 |
| 5.00 | 0.00 | 0.23 | 0.58 | 0.58 |
| 1.00 | −16.31 | −11.71 | −18.05 | −17.86 |
| 2.00 | −11.77 | −14.24 | −19.80 | −19.76 |
| 3.00 | −12.36 | −12.08 | −16.14 | −15.97 |
| 4.00 | −10.16 | −11.69 | −15.72 | −14.88 |
| 5.00 | −9.47 | −9.81 | −12.65 | −12.36 |
| 1.00 | −11.26 | −10.52 | −15.2 | −15.99 |
| 2.00 | −9.17 | −11.42 | −14.86 | −14.56 |
| 3.00 | −7.66 | −8.87 | −11.52 | −11.45 |
| 4.00 | −5.8 | −7.12 | −9.44 | −9.11 |
| 5.00 | −4.43 | −5.48 | −7.05 | −7.07 |
| 1.00 | −5.00 | −5.00 | −5.00 | −5.00 |
| 2.00 | −5.00 | −5.00 | −5.00 | −5.00 |
| 3.00 | −5.00 | −5.00 | −5.00 | −5.00 |
| 4.00 | −5.00 | −5.00 | −5.00 | −5.00 |
| 5.00 | −5.00 | −5.00 | −5.00 | −5.00 |
The dose variation: (DX_ITV-DX_GTV_4D)/DX_GTV_4D *100%, DX represents D1, D99, or D95. The V100 variation: V100_ITV-V100_GTV_4D.
Figure 3a. The ITV target (dash purple ellipse) located on the color washed ITV dose plane. b. Dose profile comparison related to ITV and real target (GTV_4D) dose planes in white arrow line position. Dose profiles out of the target were plotted as dash lines. c. The real target (dash red circle) located on color washed GTV_4D dose plane.
Tumor characteristic, dose, and dose coverage analysis of five patients
| Patient No | location | size (cm) | Amplitude (cm) | regimine | D99(cGy) | D95(cGy) | D1(cGy) | V100(%) | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ITV | GTV_4D | ITV | GTV_4D | ITV | GTV_4D | ITV | GTV_4D | |||||
| 1 | RL | 2.5 | 1.2 | 50Gy/5f | 4694(89.3%) | 5258 | 5048(92.8%) | 5440 | 6665(100.1%) | 6657 | 95.88%(−3.96%) | 99.84% |
| 2 | RL | 1.8 | 0.7 | 50Gy/4f | 4795(93.7%) | 5115 | 4980(95.3%) | 5223 | 5870(100.5%) | 5842 | 93.75%(−6.12%) | 99.87% |
| 3 | LU | 2.9 | 0.1 | 50Gy/4f | 4738(98.8%) | 4794 | 4936(99.3%) | 4972 | 5950(100.0%) | 5948 | 93.36%(−0.87%) | 94.23% |
| 4 | RU | 2.3 | 0.1 | 50Gy/5f | 4787(99.4%) | 4814 | 4949(99.2%) | 4991 | 5604(100.0%) | 5601 | 93.00%(−1.00) | 94.00% |
| 5 | LL | 1.8 | 0.9 | 50Gy/4f | 4846(96.2%) | 5037 | 4920(94.6%) | 5200 | 6343(99.9%) | 6348 | 92.99%(−6.59%) | 99.58% |
For location column, RL is right left lobe, LU is left upper lobe, RU is right upper lobe, and LL is left lower lobe.
For D99, D95, and D1 in ITV column, it both listed the absolute dose and the relative dose (in the brackets). The relative dose was the absolute dose of ITV relative to the absolute dose of GTV_4D (real target). For V100 in ITV column, the brackets listed the V100 difference (V100_ITV-V100_GTV_4D).
Figure 4a. The ITV target (purple contour) located on color washed dose distribution on AVG CT. b. Real target (red contour) located on color washed dose distribution of cumulated plan on CT_30. c. Dose volume histogram comparison between ITV and real target. d. Dose profile comparison related to the white line position in (a) and (b). The target area and location diagrams of ITV and real target were plotted under the dose profiles. Dose profiles out of the target were plotted as dash lines.