Comprehensive Comparison of Virtual Monoenergetic and Linearly Blended Reconstruction Techniques in Third-Generation Dual-Source Dual-Energy Computed Tomography Angiography of the Thorax and Abdomen.

Moritz H Albrecht1, Jesko Trommer, Julian L Wichmann, Jan-Erik Scholtz, Simon S Martin, Thomas Lehnert, Thomas J Vogl, Boris Bodelle.   

Abstract

OBJECTIVES: The aim of this study was to perform an objective and subjective image analysis of traditional and advanced noise-optimized virtual monoenergetic imaging (VMI) algorithms and standard linearly blended images in third-generation dual-source dual-energy computed tomography angiography (DE-CTA) of the thorax and abdomen.
MATERIALS AND METHODS: Thoracoabdominal DE-CTA examinations of 55 patients (36 male; mean age, 64.2 ± 12.7 years) were included in this retrospective institutional review board-approved study. Dual-energy computed tomography angiography data were reconstructed using standard linearly blended M_0.6 (merging 60% low kiloelectron volt [90 kV] with 40% high kiloelectron volt [150 kV] spectrum), traditional (VMI), and advanced VMI (VMI+) algorithms. Monoenergetic series were calculated ranging from 40 to 120 keV with 10 keV increments. Attenuation and standard deviation of 8 arteries and various anatomical landmarks of the thorax and abdomen were measured to calculate contrast-to-noise ratio values. Two radiologists subjectively assessed image quality, contrast conditions, noise, and visualization of small arterial branches using 5-point Likert scales.
RESULTS: Vascular attenuation of VMI and VMI+ series showed a gradual increase from high to low kiloelectron volt levels without significant differences between both algorithms (P < 0.894). VMI+ 40-keV series showed the highest contrast-to-noise ratio for both thoracic and abdominal DE-CTA (P < 0.001), albeit revealing higher noise than M_0.6 images (objectively and subjectively, P < 0.001) and were rated best for visualization of small arterial branches in the subjective analysis (P < 0.109). Substantially increased noise was found for VMI 40 and 50 keV series compared with all other reconstructions (objectively and subjectively, P < 0.001). VMI+ images at 100 keV+ were rated best regarding image noise (P < 0.843), whereas VMI+ reconstructions at 70 keV were found to have superior subjective image quality (P < 0.031) compared with other series except for 60 and 80 keV VMI+ series (P < 0.587). Contrast conditions at 50 keV VMI+ were rated superior compared with 60 to 100 keV VMI and VMI+ reconstructions (P < 0.012).
CONCLUSIONS: General image quality of DE-CTA examinations can be substantially improved using the VMI+ algorithm with observer preference of 70 keV, while 40 to 50 keV series provide superior contrast and improved visualization of small arterial branches compared with traditional VMI and standard linearly blended series.

Entities:  

Mesh:

Year:  2016        PMID: 26953565     DOI: 10.1097/RLI.0000000000000272

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  32 in total

1.  Assessment of an advanced virtual monoenergetic reconstruction technique in cerebral and cervical angiography with third-generation dual-source CT: Feasibility of using low-concentration contrast medium.

Authors:  Lu Zhao; Fengtan Li; Zewei Zhang; Zhang Zhang; Yingjian Jiang; Xinyu Wang; Jun Gu; Dong Li
Journal:  Eur Radiol       Date:  2018-04-13       Impact factor: 5.315

2.  Diagnostic accuracy of low and high tube voltage coronary CT angiography using an X-ray tube potential-tailored contrast medium injection protocol.

Authors:  Moritz H Albrecht; John W Nance; U Joseph Schoepf; Brian E Jacobs; Richard R Bayer; Sheldon E Litwin; Michael A Reynolds; Katharina Otani; Stefanie Mangold; Akos Varga-Szemes; Domenico De Santis; Marwen Eid; Georg Apfaltrer; Christian Tesche; Markus Goeller; Thomas J Vogl; Carlo N De Cecco
Journal:  Eur Radiol       Date:  2017-11-27       Impact factor: 5.315

3.  Modified calcium subtraction in dual-energy CT angiography of the lower extremity runoff: impact on diagnostic accuracy for stenosis detection.

Authors:  Domenico De Santis; Carlo N De Cecco; U Joseph Schoepf; John W Nance; Ricardo T Yamada; Brooke A Thomas; Katharina Otani; Brian E Jacobs; D Alan Turner; Julian L Wichmann; Marwen Eid; Akos Varga-Szemes; Damiano Caruso; Katharine L Grant; Bernhard Schmidt; Thomas J Vogl; Andrea Laghi; Moritz H Albrecht
Journal:  Eur Radiol       Date:  2019-02-25       Impact factor: 5.315

Review 4.  Dual energy computed tomography virtual monoenergetic imaging: technique and clinical applications.

Authors:  Tommaso D'Angelo; Giuseppe Cicero; Silvio Mazziotti; Giorgio Ascenti; Moritz H Albrecht; Simon S Martin; Ahmed E Othman; Thomas J Vogl; Julian L Wichmann
Journal:  Br J Radiol       Date:  2019-04-09       Impact factor: 3.039

5.  Noise-optimised virtual monoenergetic imaging of dual-energy CT: effect on metal artefact reduction in patients with lumbar internal fixation.

Authors:  Yanwei Zeng; Kai Deng; Haitao Yang; Yi Tan; Jun Liu; Daoying Geng; Jun Zhang
Journal:  Eur Spine J       Date:  2019-04-30       Impact factor: 3.134

6.  90-kVp low-tube-voltage CT pulmonary angiography in combination with advanced modeled iterative reconstruction algorithm: effects on radiation dose, image quality and diagnostic accuracy for the detection of pulmonary embolism.

Authors:  Doris Leithner; Tatjana Gruber-Rouh; Martin Beeres; Julian L Wichmann; Scherwin Mahmoudi; Simon S Martin; Lukas Lenga; Moritz H Albrecht; Christian Booz; Thomas J Vogl; Jan-Erik Scholtz
Journal:  Br J Radiol       Date:  2018-06-05       Impact factor: 3.039

7.  Improving iodine contrast to noise ratio using virtual monoenergetic imaging and prior-knowledge-aware iterative denoising (mono-PKAID).

Authors:  Shengzhen Tao; Kishore Rajendran; Wei Zhou; Joel G Fletcher; Cynthia H McCollough; Shuai Leng
Journal:  Phys Med Biol       Date:  2019-05-16       Impact factor: 3.609

8.  Comparison of virtual monoenergetic and polyenergetic images reconstructed from dual-layer detector CT angiography of the head and neck.

Authors:  Victor Neuhaus; Nils Große Hokamp; Nuran Abdullayev; Volker Maus; Christoph Kabbasch; Anastasios Mpotsaris; David Maintz; Jan Borggrefe
Journal:  Eur Radiol       Date:  2017-10-10       Impact factor: 5.315

9.  Metal artifacts in patients with large dental implants and bridges: combination of metal artifact reduction algorithms and virtual monoenergetic images provides an approach to handle even strongest artifacts.

Authors:  Kai Roman Laukamp; David Zopfs; Simon Lennartz; Lenhard Pennig; David Maintz; Jan Borggrefe; Nils Große Hokamp
Journal:  Eur Radiol       Date:  2019-01-16       Impact factor: 5.315

10.  Dual-energy CT in early acute pancreatitis: improved detection using iodine quantification.

Authors:  Simon S Martin; Franziska Trapp; Julian L Wichmann; Moritz H Albrecht; Lukas Lenga; James Durden; Christian Booz; Thomas J Vogl; Tommaso D'Angelo
Journal:  Eur Radiol       Date:  2018-11-28       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.