Measuring small forces is a major challenge in cell biology. Here we improve the spatial resolution and accuracy of force reconstruction of the well-established technique of traction force microscopy (TFM) using STED microscopy. The increased spatial resolution of STED-TFM (STFM) allows a greater than 5-fold higher sampling of the forces generated by the cell than conventional TFM, accessing the nano instead of the micron scale. This improvement is highlighted by computer simulations and an activating RBL cell model system.
Measuring small forces is a major challenge in cell biology. Here we improve the spatial resolution and accuracy of force reconstruction of the well-established technique of traction force microscopy (TFM) using STED microscopy. The increased spatial resolution of STED-TFM (STFM) allows a greater than 5-fold higher sampling of the forces generated by the cell than conventional TFM, accessing the nano instead of the micron scale. This improvement is highlighted by computer simulations and an activating RBL cell model system.
Entities:
Keywords:
Super-resolution microscopy; actin cytoskeleton; mechanobiology; traction force microscopy
It is becoming
increasingly
clear that mechanical force plays a crucial role in many biological
processes, including adhesion, migration, and cell signaling.[1−3] Forces act across many length scales, from tissue to the single
cell and ultimately down to the molecular level, as is true for cells
of the immune system where individual cell–cell and receptor–ligand
interactions are crucial.[4−6] In order to understand the role
of forces within a given biological system, it is important that we
have the appropriate tools and techniques that allow quantification
of mechanical forces at the relevant length scales. A commonly used
technique to measure forces on the micron-scale in cell biological
systems is traction force microscopy (TFM) (or simply traction microscopy).Beginning with the pioneering work of Harris et al., flexible substrates,
such as polyacrylamide (PAA) gels, have been used to investigate cellular
tractions and forces for over 30 years.[7] In a typical TFM experiment, a thin (20–30 μm) elastic
gel is formed on a glass coverslip onto which proteins facilitating
cell adherence can be attached (Figure a).[8] Within the gel, fluorescent
beads serve as fiducial markers and imaging of the bead positions
over time during the application of cellular tractions allows the
displacement of the gel to be quantified. By combining the measured
displacement field with knowledge of the mechanical properties of
the gel, the tractions responsible for the displacements can be calculated.
The optimal theoretical treatment of the traction solution has been
the subject of much research.[9−11] Dembo et al. provided a rigorous
mathematical framework for the use of elastic materials to measure
traction forces,[9] which was further developed
with the introduction of Fourier transform traction cytometry (FTTC)
whose treatment of the force reconstruction in Fourier space allowed
for greatly decreased computation time.[10] Owing to the inverse nature of the problem, work has shown the need
for regularization to control the influence of experimental noise
in the measured displacements on the traction solution.[11] In addition to FTTC, other methods of traction
reconstruction have been developed whereby experimental knowledge
of the traction locations is used to aid the traction recovery, for
example, traction reconstruction with point forces (TRPF)[11] and more recently, model based traction force
microscopy (MBTFM).[12] These methods are
only applicable in cases where the traction location can be inferred
from fluorescent data, as in the case of fluorescently labeled focal
adhesions. In the more general case, where no knowledge of the traction
location is assumed, FTTC is more suitable and is the methodology
used in this work.
Figure 1
Theoretical characterization of STFM. (a) Schematic representation
of a typical TFM setup. An elastic polyacrylamide gel filled with
fluorescent marker beads is covalently attached to a glass coverslip
and functionalized with proteins that facilitate cell adherence. Traction
forces applied by the cell to the top surface of the gel results in
lateral displacements of the gel which can be quantified by imaging
the displacement of the beads within the gel. (b) Theoretical relationship
between the sampling density and the Nyquist limit (dashed line), with three different
bead densities highlighted (red, blue, green as labeled), exemplifying
that a bead density of 15 μm–2 would allow
the recovery of tractions 500 nm in size. Crosses show the smallest
recoverable tractions from simulations performed at the three bead
densities shown in Figure . Open circles show the smallest recoverable traction from
simulations where noise is added and regularization used.
Theoretical characterization of STFM. (a) Schematic representation
of a typical TFM setup. An elastic polyacrylamide gel filled with
fluorescent marker beads is covalently attached to a glass coverslip
and functionalized with proteins that facilitate cell adherence. Traction
forces applied by the cell to the top surface of the gel results in
lateral displacements of the gel which can be quantified by imaging
the displacement of the beads within the gel. (b) Theoretical relationship
between the sampling density and the Nyquist limit (dashed line), with three different
bead densities highlighted (red, blue, green as labeled), exemplifying
that a bead density of 15 μm–2 would allow
the recovery of tractions 500 nm in size. Crosses show the smallest
recoverable tractions from simulations performed at the three bead
densities shown in Figure . Open circles show the smallest recoverable traction from
simulations where noise is added and regularization used.
Figure 2
Outline of the simulation process. (a) A uniform
circular traction
field Tsimulated(x) is
simulated and the corresponding displacement field u(x) calculated (heat map; high traction magnitude
warm colors, low traction magnitudes cold colors, white arrows: traction
direction). The displacement field is then subsampled at a confocal
and STED density (red dots: bead positions, black arrows: bead displacements),
the traction field recovered Trecovered(x) and the simulation and recovery compared by
the deviation of traction magnitude (DTM). Scale bar 1 μm. (b)
DTM for varying traction diameters at three sampling densities, confocal
(red), medium STED (blue), and maximum STED (green). A DTM of 0 represents
a perfect traction recovery, whereas a DTM of −1 represents
a complete underestimation. Dotted line: DTM for no subsampling. Line
deviates from zero at large tractions due to artifacts introduced
by the finite size of the simulated gel area. (c) Same as b with the
addition of artificial noise and using the regularized solution, showing
very similar dependency as b except for the no subsampling case (dotted
line), where regularization masks the recovered tractions at length
scales matching that of the artificial noise. (d) Simulation and traction
recovery for a 1 μm diameter circular traction zone (0.3 kPa).
Scale bar 2 μm. (e) Simulation and traction recovery for a 1
μm wavelength periodic traction pattern (0–0.3 kPa).
Scale bar 2 μm.
The greatest shortcoming of classical TFM is its
limited sensitivity
due to the finite density at which the displacement field can be sampled
within the gel.[13] The density of fiducial
markers must be high enough to reflect the complexity of the traction
field that is applied by the cell. If the bead density is too low,
areas of the gel will move without being reported by any bead movement
and the traction information is lost. This can be thought of as a
sampling problem, where to meet the Nyquist criteria the spatial sampling
frequency of the displacement field must be twice that of any details
that may be resolved in the displacement field (Figure b).[13] Experimentally,
this limit is imposed by the finite size point spread function (PSF)
that results from each marker bead. At high densities, the PSF of
each individual bead begin to overlap, meaning nearby beads can no
longer be resolved individually, obscuring details of their relative
displacement. A first attempt to overcome this limitation involved
the use of two different colors of marker beads which proved that
the recovery of micron sized tractions are feasible.[13] However, due to its reliance on the spectral separation
of the beads, this technique is ultimately limited by the spectral
range of the microscope. Because the beads must be imaged at the top
surface of the gel, this requires a microscope technique that can
operate away from the coverslip, meaning TIRF or near-field microscopy
are not suitable. Cellular traction fields are typically on the nanoscale
range rather than on the micron-scale. Consequently, there remains
a need to improve the spatial resolution of TFM. While improved analysis
tools might introduce some advancements in resolving force fields
(e.g., using TRPF involving knowledge from additional fluorescence
data of the sample, as outlined above), these approaches are experimentally
still limited by the finite size of the PSF as given by diffraction
for conventional optical microscopes (i.e., they only push the TFM
read-out to its ultimate limit as given by diffraction). To overcome
these challenges, here we improved the spatial resolution and accuracy
of force reconstruction of TFM by using super-resolution optical STED
microscopy.[14]To examine the effects
of the sampling density on traction force
recovery, we first conducted computer simulations in which a gel of
defined stiffness (3 kPa) was exposed to a uniform circular traction
field (T) (0.3 kPa) of varying spatial sizes (0.1–4.0
μm). The resulting displacement field was then calculated using
the mathematical framework provided by FTTC (Supporting Information, eq S3) (Figure a). To simulate the discrete nature of bead sampling,
the displacement field was subsampled at random points, with sampling
densities corresponding to those attainable by confocal or STED microscopy:
15 beads per μm2 for high, theoretically achievable
STED resolution (40 nm), 3 beads per μm2 for STED
resolution achievable in the current experiments (80 nm), and 0.5
beads per μm2 for confocal. The subsampled displacement
field was then transformed back into a traction field (Supporting Information, eq S5). The recovered
and simulated traction field were then compared, and the difference
quantified via a metric known as the deviation of traction magnitude
(DTM), where a DTM of −1 represents a complete underestimation
and 0 represents a perfect recovery of the traction.[13]Outline of the simulation process. (a) A uniform
circular traction
field Tsimulated(x) is
simulated and the corresponding displacement field u(x) calculated (heat map; high traction magnitude
warm colors, low traction magnitudes cold colors, white arrows: traction
direction). The displacement field is then subsampled at a confocal
and STED density (red dots: bead positions, black arrows: bead displacements),
the traction field recovered Trecovered(x) and the simulation and recovery compared by
the deviation of traction magnitude (DTM). Scale bar 1 μm. (b)
DTM for varying traction diameters at three sampling densities, confocal
(red), medium STED (blue), and maximum STED (green). A DTM of 0 represents
a perfect traction recovery, whereas a DTM of −1 represents
a complete underestimation. Dotted line: DTM for no subsampling. Line
deviates from zero at large tractions due to artifacts introduced
by the finite size of the simulated gel area. (c) Same as b with the
addition of artificial noise and using the regularized solution, showing
very similar dependency as b except for the no subsampling case (dotted
line), where regularization masks the recovered tractions at length
scales matching that of the artificial noise. (d) Simulation and traction
recovery for a 1 μm diameter circular traction zone (0.3 kPa).
Scale bar 2 μm. (e) Simulation and traction recovery for a 1
μm wavelength periodic traction pattern (0–0.3 kPa).
Scale bar 2 μm.By calculating the DTM for circular traction zones of varying
diameters
at the three different sampling densities, it is evident that increasing
the sampling density allows for the successful recovery of spatially
more confined tractions (Figure b). By adding artificial Gaussian distributed noise
to the displacement field at a level consistent with the experiment
(10% of the maximum) and using the regularized solution, it is clear
that the relationship between sampling density and traction recovery
is maintained (Supporting Information,
eq S7) (Figure c).
For each sampling density, defining a DTM of −0.2 as the minimum
required to recover a traction, it is possible to plot the corresponding
traction size on Figure b, which displayed good agreement between simulation and the Nyquist
limit. To further demonstrate this improvement, the simulation and
recovery is shown for a circular traction (0.3 kPa) of 1 μm
diameter at the three different sampling densities (Figure d). We also show the simulation
and corresponding recovery of a more complex periodic traction field
(0–0.3 kPa) with a wavelength of 1 μm (Figure e). In all cases, we find that
the higher the sampling density, the more detail is recovered in the
traction field.In addition to the sampling density, another
important factor in
determining the accuracy of TFM is the method used to recover the
displacement of the beads from the fluorescent images. The most common
methods of extracting bead displacements are those based on single
particle tracking (SPT), where each individual bead must be localized,
and those methods based on statistical comparisons of fluorescent
images, such as particle image velocimetry (PIV). To this end, the
image is divided into a grid, and each grid element is spatially correlated
between frames to assess the degree of movement. PIV does not require
localization of each bead but is limited spatially by the size of
the grid elements required to give accurate correlations. It can be shown
that the increased resolution of STED allows for more accurate recovery
of the displacement field and hence a more accurate force field in
both SPT and PIV (Supporting Information, Figure S3 and Table S1). Note, the magnitude of PIV recovered high
bead density based force fields can be close to the simulated values
but increased spatial accuracy necessitates STED.To observe
the effect of increased sampling experimentally in the
single cell environment, RBL cells, which express high levels of the
Fcε receptor-1 (FcεRI), were allowed to interact with
a 3 kPa gel loaded with 40 nm red-fluorescent beads and coated with
the antibody, IgE (Figure a). The Fc portion of IgE binds with high affinity (equilibrium
constant Ka = 1010 M–1, off-rate koff = 10–5 s–1) to the FcεRI present on the RBL cell
surface, which results in cell spreading and activation.[15,16] By fluorescently labeling actin filaments via Lifeact-citrine, we
were able to visualize the dynamically spreading cell edge as the cell-gel
contact area increased (Figure b). On visual inspection (Movie S1), the beads within this area are seen to move elastically in a directed
manner toward the direction of cell motion, indicating forces being
applied by the cell to the compliant gel beneath via the FcεRI-IgE
interaction. By using a water immersion objective, optical aberrations
from imaging through the gel layer were low, allowing STED imaging
with a spatial resolution of around 80 nm (Figure S1), and we have observed no visible sign of cell degradation
(such as cell contraction) during the around 120 s long recordings.
Figure 3
Experimental
demonstration of STFM. (a) Gel functionalization.
(Left) Scheme: The roughly 30 μm thick PAA gel layer (light
blue) was loaded with 40 nm-large red fluorescent beads (red dots)
and surface-coated with poly-l-lysine (light green) followed
by attachment of IgE (green). (Middle, right) Confocal z–x profile images of the gel cross-section
showing concentration of Alexa488 labeled IgE (green, middle) and
red fluorescent beads (red, right) at the top surface of the gel.
Scale bar 30 μm. (b) Representative confocal image of fluorescent
F-actin (Lifeact-citrine) expressing RBL cell (green) interacting
with IgE coated 3 kPa PAA gel loaded with the red fluorescent beads
(red). Scale bar 10 μm. (c) Time-lapse imaging of the spreading
cell edge results in the displacement of the beads within the gel,
monitored for different conditions as labeled. (Left panels) Confocal
images of fluorescent F-actin (green) and confocal or STED images
of red fluorescent beads (red) at a certain time point together with
the temporal displacement tracks of the beads (time color-coded as
labeled), for low (0.4 μm–2) and high (2.2
μm–2) bead density. Scale bar 2 μm.
For confocal at high bead density (lower left) no bead tracks could
be resolved; instead a bar chart is shown, quantifying the ability
to successfully locate and track beads in the high density confocal
case compared to the high density STED case (total number of beads:
140 STED, 60 confocal). (d) Recovered traction field for the high
density STED tracking of c (left) and extrapolated low density effective
confocal tracking (right) with force color-coded in kPa. (e) Quantification
of the F-actin flow from the high density STED recording of c by optical
flow (left) and correlation (color coded with 1.0 showing maximum
correlation) with the bead displacement (right).
Experimental
demonstration of STFM. (a) Gel functionalization.
(Left) Scheme: The roughly 30 μm thick PAA gel layer (light
blue) was loaded with 40 nm-large red fluorescent beads (red dots)
and surface-coated with poly-l-lysine (light green) followed
by attachment of IgE (green). (Middle, right) Confocal z–x profile images of the gel cross-section
showing concentration of Alexa488 labeled IgE (green, middle) and
red fluorescent beads (red, right) at the top surface of the gel.
Scale bar 30 μm. (b) Representative confocal image of fluorescent
F-actin (Lifeact-citrine) expressing RBL cell (green) interacting
with IgE coated 3 kPa PAA gel loaded with the red fluorescent beads
(red). Scale bar 10 μm. (c) Time-lapse imaging of the spreading
cell edge results in the displacement of the beads within the gel,
monitored for different conditions as labeled. (Left panels) Confocal
images of fluorescent F-actin (green) and confocal or STED images
of red fluorescent beads (red) at a certain time point together with
the temporal displacement tracks of the beads (time color-coded as
labeled), for low (0.4 μm–2) and high (2.2
μm–2) bead density. Scale bar 2 μm.
For confocal at high bead density (lower left) no bead tracks could
be resolved; instead a bar chart is shown, quantifying the ability
to successfully locate and track beads in the high density confocal
case compared to the high density STED case (total number of beads:
140 STED, 60 confocal). (d) Recovered traction field for the high
density STED tracking of c (left) and extrapolated low density effective
confocal tracking (right) with force color-coded in kPa. (e) Quantification
of the F-actin flow from the high density STED recording of c by optical
flow (left) and correlation (color coded with 1.0 showing maximum
correlation) with the bead displacement (right).Next, to demonstrate the influence of the bead sampling density,
we introduce four different scenarios; low density confocal and STED
(0.4 beads μm–2) (Movies S2 and S3), and high density confocal
and STED (2.2 beads μm–2) (Figure c) (Movies S4 and S5). Here we define low density
as the maximum trackable density by confocal, and high density as
the maximum trackable density of our current STED experiments. Note,
bead sampling values in the scenario of high density STED were a moderate
and robust choice considering the experimental optical conditions
and needs of the biological specimen. However, they were below the
computationally predicted possible advances of STFM. Specifically,
the bead densities are a function of the microscopes PSF size and
could be improved in future work by optimizing the imaging conditions,
e.g., minimizing optical aberrations (see discussion).To assess
the displacement of the beads we chose to use SPT as
it allows the movements of each individual bead to be captured. PIV
is generally best suited to tractions where collective bead movements
are expected, for example focal adhesions. For RBL cells, forces may
arise from localized receptor–ligand interactions and may be
spatially complex. In the low density case, applying a custom written
MATLAB SPT algorithm allowed the displacements of all beads within
the field of view to be measured in both the confocal and STED image
sequences (Supporting Information and Figure S2). In the high density case, confocal imaging resulted in a significant
number of overlapping PSFs, preventing reliable bead tracking. However,
on applying the STED beam, beads were resolved individually and the
tracking was successful (Figure c). In all cases, bead tracks were interpolated onto
a regular mesh and the corresponding traction field calculated using
the appropriate degree of regularization (Figure S4). Obviously, in the high density case only the STED imaging
yielded a traction field (Figure d). To directly compare the effect of sampling density
on the ability to accurately recover the traction field of the same
cell, beads in the high density STED case were randomly deleted until
the bead density was equal to that attainable by confocal tracking
(Figure d). It is
clear that this reduces the information content present in the displacement
field, and hence reduces the fine detail in the traction field.Finally, to identify the origin of mechanical force generation
in RBL cells, we combined fluorescent imaging and STFM. The technique
of optical flow enables the spatial change in pixel intensity to be
quantified over time and is commonly applied in computer vision to
assess the spatial shift between image frames. Here, we apply this
technique to extract the retrograde flow vector field of fluorescently
labeled filamentous actin (Lifeact-citrine). This vector field was
then correlated via a dot product with the displacement field of the
beads in the gel, yielding a spatial correlation of actin and bead
displacement (Figure e).[17] The correlation highlights that
areas of the cell showing the most dynamic actin coincide with the
areas of greatest bead displacement, highlighting that it was indeed
the flow of actin within the cell that was responsible for the observed
tractions. As in the case of traction forces, the higher bead density
leads to a higher information content in the displacement field and
hence a more reliable correlation between the two vector fields.In summary, the increase in accuracy of STFM is important when
considering cellular forces on small length scales, as is the case
for receptor-antagonist interactions in immune cells.[18] Using STFM, we are now better able to make links between
the forces generated by the cell, and those molecules which are responsible
for force generation. This is particularly valuable when force measurements
are coupled to fluorescent data, as is shown in Figure e.We have focused on the experimental
aspects of TFM. Moreover, recent
work has suggested that theoretical aspects may be equally important
in optimizing the accuracy of force reconstruction.[19] For example, in the case of sparse focal adhesions, it
has been demonstrated that the L1-norm is favorable to the L2-norm
used to assess to degree of regularization,[20,21] owing to a greater retention
of the high resolution detail in the force maps. We had no prior knowledge
of the traction field induced by the RBL cells, therefore we choose
L2-norm regularization as the more general case. However, further
work should focus on optimizing both experimental and computational
aspects of the technique to further increase the accuracy of STFM.Notably, increasing the location accuracy of STFM is theoretically
not limited since it scales with the applied STED power.[14] This needs to be balanced with other optical
factors such as maximal bead density within the gel, the ability to
track the beads, and fluorescence light sensitivity of the biological
specimen. Moreover, the nature of the (S)TFM setup requires all imaging
to be done at the top surface of the gel, meaning imaging is subject
to aberrations induced by the mismatch in refractive index of the
gel and the immersion media. Improvements in aberration correction,
for example using adaptive optics would reduce the effect of these
aberrations and would result in an improved STED resolution, possibly
along all three spatial dimensions,[22] allowing
even higher bead densities to be used and the accuracy of experiments
to approach those shown possible by simulations. This also presents
the opportunity of performing 3D-TFM in high resolution. In the same
way that 2D-STED can increase the accuracy of the tangential force
reconstruction, using 3D-STED would allow for a greater sampling of
the forces perpendicular to the gel surface.
Authors: James P Butler; Iva Marija Tolić-Nørrelykke; Ben Fabry; Jeffrey J Fredberg Journal: Am J Physiol Cell Physiol Date: 2002-03 Impact factor: 4.249
Authors: Jiro Kitaura; Jinming Song; Mindy Tsai; Koichi Asai; Mari Maeda-Yamamoto; Attila Mocsai; Yuko Kawakami; Fu-Tong Liu; Clifford A Lowell; B George Barisas; Stephen J Galli; Toshiaki Kawakami Journal: Proc Natl Acad Sci U S A Date: 2003-10-20 Impact factor: 11.205
Authors: Celeste M Nelson; Ronald P Jean; John L Tan; Wendy F Liu; Nathan J Sniadecki; Alexander A Spector; Christopher S Chen Journal: Proc Natl Acad Sci U S A Date: 2005-07-27 Impact factor: 11.205
Authors: Erdem Tabdanov; Sasha Gondarenko; Sudha Kumari; Anastasia Liapis; Michael L Dustin; Michael P Sheetz; Lance C Kam; Thomas Iskratsch Journal: Integr Biol (Camb) Date: 2015-07-09 Impact factor: 2.192
Authors: Ivan Pushkarsky; Peter Tseng; Dylan Black; Bryan France; Lyndon Warfe; Cynthia J Koziol-White; William F Jester; Ryan K Trinh; Jonathan Lin; Philip O Scumpia; Sherie L Morrison; Reynold A Panettieri; Robert Damoiseaux; Dino Di Carlo Journal: Nat Biomed Eng Date: 2018-02-06 Impact factor: 25.671
Authors: Lothar Schermelleh; Alexia Ferrand; Thomas Huser; Christian Eggeling; Markus Sauer; Oliver Biehlmaier; Gregor P C Drummen Journal: Nat Cell Biol Date: 2019-01-02 Impact factor: 28.824
Authors: Toshio Ando; Satya Prathyusha Bhamidimarri; Niklas Brending; H Colin-York; Lucy Collinson; Niels De Jonge; P J de Pablo; Elke Debroye; Christian Eggeling; Christian Franck; Marco Fritzsche; Hans Gerritsen; Ben N G Giepmans; Kay Grunewald; Johan Hofkens; Jacob P Hoogenboom; Kris P F Janssen; Rainer Kaufman; Judith Klumpermann; Nyoman Kurniawan; Jana Kusch; Nalan Liv; Viha Parekh; Diana B Peckys; Florian Rehfeldt; David C Reutens; Maarten B J Roeffaers; Tim Salditt; Iwan A T Schaap; Ulrich S Schwarz; Paul Verkade; Michael W Vogel; Richard Wagner; Mathias Winterhalter; Haifeng Yuan; Giovanni Zifarelli Journal: J Phys D Appl Phys Date: 2018-08-31 Impact factor: 3.207
Authors: Nils M Kronenberg; Philipp Liehm; Anja Steude; Johanna A Knipper; Jessica G Borger; Giuliano Scarcelli; Kristian Franze; Simon J Powis; Malte C Gather Journal: Nat Cell Biol Date: 2017-06-19 Impact factor: 28.824