| Literature DB >> 26914656 |
Fabian Niess1,2,3, Georg B Fiedler1,2, Albrecht I Schmid1,2, Sigrun Goluch1,2,4, Roberta Kriegl1,2, Michael Wolzt5, Ewald Moser1,2, Martin Meyerspeer1,2.
Abstract
PURPOSE: Separate measurements are required when investigating multiple exercising muscles with singlevoxel-localized dynamic 31 P-MRS. With multivoxel spectroscopy, 31 P-MRS time-series spectra are acquired from multiple independent regions during one exercise-recovery experiment with the same time resolution as for singlevoxel measurements.Entities:
Keywords: dynamic 31P MRS; exercising muscle; multivoxel; spectroscopic localization
Mesh:
Substances:
Year: 2016 PMID: 26914656 PMCID: PMC4996323 DOI: 10.1002/mrm.26172
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668
Figure 1Schematic representation of the cylindrical phantom (containing two small compartments) atop the multichannel transceiver coil (a), with a transversal localizer image showing the compartments and the shim volume [green rectangle (b)]. The effect of partially overlapping voxels was quantified in an unstructured test object, by moving voxel 2 along the x‐axis in steps of 5 mm, while holding the position of voxel 1 constant (c). Transversal gradient echo image of a human calf muscle with the positions of two VOIs for acquiring 31P spectra from gastrocnemius medialis and soleus. Shaded sections schematically show the slice affected by the excitation pulses (d).
Figure 2Comparison of signals acquired separately using singlevoxel spectroscopy (T R = 8 s each) from the left (a) and right (c) compartment inside the structured phantom, with data collected with the multivoxel acquisition approach [left (b) and right (d) within T R = 8 s with 4 s delay between each localization]. There is no visible contamination using interleaved acquisition of both VOIs. The signal loss was in the left (b) and in the right VOI (d). Contamination from the surrounding would result in a resonance peak at 0 ppm.
Signals Measured from Two 35 mm3 VOIs Placed Side By Side Inside a Homogeneous Phantom Close to the Coil
| Shared excitation slice | Area of signal peak (a.u) | SNR | Signal loss |
|---|---|---|---|
| Left voxel reference | 6563 ± 49 | 27.2 ± 1.1 | – |
| Right voxel reference | 7693 ± 48 | 22.8 ± 1.1 | – |
| Left voxel MVS shared excitation | 3527 ± 12 | 12.5 ± 0.4 | 46 ± 0.3% |
| Right voxel MVS shared excitation | 4991 ± 55 | 10.6 ± 0.4 | 35 ± 1% |
| Shared refocusing slices | |||
| Left voxel reference | 3994 ± 22 | 17.7 ± 0.6 | – |
| Right voxel reference | 6178 ± 55 | 17.7 ± 0.8 | – |
| Left voxel MVS (1 + 7) | 3607 ± 19 | 16.1 ± 0.6 | 10 ± 1% |
| Right voxel MVS (1 + 7) | 4860 ± 50 | 14.7 ± 0.6 | 21 ± 1% |
| Left voxel MVS (4 + 4) | 3737 ± 29 | 16.7 ± 0.8 | 6 ± 1% |
| Right voxel MVS (4 + 4) | 5563 ± 56 | 16.4 ± 0.7 | 10 ± 1% |
| Left voxel MVS (4 + 4) + 12.5% | 3880 ± 18 | 16.7 ± 0.7 | 3 ± 1% |
| Right voxel MVS (4 + 4) + 12.5% | 5681 ± 44 | 16.5 ± 0.4 | 8 ± 1% |
Both VOIs were acquired separately with T R = 8 s as reference signals. For multivoxel spectroscopy (MVS), the delay between the localizations was 1 s, followed by a delay of 7 s with a shared excitation slice only and vice versa for each voxel position. The influence of sharing both refocusing pulses was tested with asymmetric (1 + 7 s) and symmetric (4 + 4 s) timing. To reduce saturation effects caused by refocusing pulses, the measurement was repeated with the refocusing pulse voltage increased by 12.5% (from 160 to 180 V).
Figure 3Signals of a static VOI close to the coil acquired 1 s after the two adiabatic full passage refocusing pulses were applied to localize a second voxel. The second voxel was located on the left upper corner of the measured VOI and was shifted in 5 mm steps to the right after each measurement. Blue line: influence of one refocusing slice at different overlap positions. Red line: the second refocusing slice of the moving voxel constantly fully overlaps with the VOI. Both curves show the same features, the additional refocusing pulse induces an offset by −10%. Outside the dashed lines, the nominal voxels do not overlap.
Figure 4Time series of 31P‐MR spectra acquired in gastrocnemius medialis and soleus with a multivoxel protocol during one exercise‐recovery experiment comprising rest (2 min), exercise (5 min), and recovery (7 min). Unaveraged spectra are shown, time resolution 6 s (a). Spectra from gastrocnemius medialis and soleus at rest, representing 10 averaged spectra (1 min acquisition time), which were measured either separately using a singlevoxel protocol (b) or using the interleaved multivoxel protocol (c). Resting spectra have identical scale and are zoomed to metabolites, cutting off phosphocreatine, while the inserts show the full scale. All spectra were acquired with 2048 points, bandwidth of 5000 Hz and were filtered with 7 Hz apodisation and zero filled by a factor of 4 for display.