Safe bunker designing for the 18 MV Varian 2100 Clinac: a comparison between Monte Carlo simulation based upon data and new protocol recommendations.

Manije Beigi1, Fatemeh Afarande2, Hosein Ghiasi3.   

Abstract

AIM: The aim of this study was to compare two bunkers designed by only protocols recommendations and Monte Carlo (MC) based upon data derived for an 18 MV Varian 2100Clinac accelerator.
BACKGROUND: High energy radiation therapy is associated with fast and thermal photoneutrons. Adequate shielding against the contaminant neutron has been recommended by IAEA and NCRP new protocols.
MATERIALS AND METHODS: The latest protocols released by the IAEA (safety report No. 47) and NCRP report No. 151 were used for the bunker designing calculations. MC method based upon data was also derived. Two bunkers using protocols and MC upon data were designed and discussed.
RESULTS: From designed door's thickness, the door designed by the MC simulation and Wu-McGinley analytical method was closer in both BPE and lead thickness. In the case of the primary and secondary barriers, MC simulation resulted in 440.11 mm for the ordinary concrete, total concrete thickness of 1709 mm was required. Calculating the same parameters value with the recommended analytical methods resulted in 1762 mm for the required thickness using 445 mm as recommended by TVL for the concrete. Additionally, for the secondary barrier the thickness of 752.05 mm was obtained.
CONCLUSION: Our results showed MC simulation and the followed protocols recommendations in dose calculation are in good agreement in the radiation contamination dose calculation. Difference between the two analytical and MC simulation methods revealed that the application of only one method for the bunker design may lead to underestimation or overestimation in dose and shielding calculations.

Keywords:  Monte Carlo; Protocols; Radiation protection; Shielding

Year:  2015        PMID: 26900357      PMCID: PMC4716400          DOI: 10.1016/j.rpor.2015.10.003

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  10 in total

1.  Analysis of photoneutron spectra produced in medical accelerators.

Authors:  C Ongaro; A Zanini; U Nastasi; J Ródenas; G Ottaviano; C Manfredotti; K W Burn
Journal:  Phys Med Biol       Date:  2000-12       Impact factor: 3.609

2.  Experimental and computational determination of neutron dose equivalent around radiotherapy accelerators.

Authors:  E J Waller; T J Jamieson; D Cole; T Cousins; R B Jammal
Journal:  Radiat Prot Dosimetry       Date:  2003       Impact factor: 0.972

3.  Evaluation of neutron dose equivalent levels at the maze entrance of medical accelerator treatment rooms.

Authors:  P H McGinley; E K Butker
Journal:  Med Phys       Date:  1991 Mar-Apr       Impact factor: 4.071

4.  Evaluation of the photoneutron field produced in a medical linear accelerator.

Authors:  H S Kim; Y H Park; B C Koo; J W Kwon; J-S Lee; H S Choi
Journal:  Radiat Prot Dosimetry       Date:  2006-10-30       Impact factor: 0.972

Review 5.  A review on photoneutrons characteristics in radiation therapy with high-energy photon beams.

Authors:  Alireza Naseri; Asghar Mesbahi
Journal:  Rep Pract Oncol Radiother       Date:  2010-09-22

6.  A new analytical formula for neutron capture gamma dose calculations in double-bend mazes in radiation therapy.

Authors:  Hosein Ghiasi; Asghar Mesbahi
Journal:  Rep Pract Oncol Radiother       Date:  2012-05-24

7.  Photoneutron and capture gamma dose equivalent for different room and maze layouts in radiation therapy.

Authors:  Asghar Mesbahi; Hosein Ghiasi; Seyed Rabee Mahdavi
Journal:  Radiat Prot Dosimetry       Date:  2010-01-18       Impact factor: 0.972

8.  Monte Carlo characterizations mapping of the (γ,n) and (n,γ) photonuclear reactions in the high energy X-ray radiation therapy.

Authors:  Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2013-07-31

9.  Robust demarcation of basal cell carcinoma by dependent component analysis-based segmentation of multi-spectral fluorescence images.

Authors:  Ivica Kopriva; Antun Persin; Neira Puizina-Ivić; Lina Mirić
Journal:  J Photochem Photobiol B       Date:  2010-04-03       Impact factor: 6.252

10.  Neutron and capture gamma along the mazes of linear accelerator vaults.

Authors:  Raymond K Wu; Patton H McGinley
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  10 in total
  3 in total

1.  MC safe bunker designing for an 18 MV linac with nanoparticles included primary barriers and effect of the nanoparticles on the shielding aspects.

Authors:  Amir Ghasemi-Jangjoo; Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-06

2.  Monte Carlo study on the secondary cancer risk estimations for patients undergoing prostate radiotherapy: A humanoid phantom study.

Authors:  Amir Ghasemi-Jangjoo; Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2020-01-10

3.  Application of the phase-space distribution approach of Monte Carlo for radiation contamination dose estimation from the (n,γ), (γ,n) nuclear reactions and linac leakage photons in the megavoltage radiotherapy facility.

Authors:  Amir Ghasemi-Jangjoo; Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2020-01-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.