Evaluation of neutron dose equivalent levels at the maze entrance of medical accelerator treatment rooms.

P H McGinley1, E K Butker.   

Abstract

This paper evaluates the accuracy of Kersey's method of calculation of the dose level due to photoneutrons at the maze entrance of medical accelerators. In this study measurements and calculations of the neutron dose equivalent were made for 13 medical accelerator facilities. The group of accelerators was composed of four different models from two manufacturers. The maze length for the treatment rooms varied from 3.0 to 8.54 m and 10 of the mazes had a single 90-deg turn with the remaining having two 90-deg turns. Moderated activation detectors and a portable neutron remmeter were used for measurements in the treatment room and maze, respectively. It was found that the maximum disagreement between the measured and calculated values was a factor of 2.3 with the calculated value exceeding the measured value. The majority of the measured values were within 25% of the calculated levels. It was concluded that Kersey's method is suitable for use in designing medical accelerator mazes.

Mesh:

Year:  1991        PMID: 2046615     DOI: 10.1118/1.596713

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

Review 1.  A review on photoneutrons characteristics in radiation therapy with high-energy photon beams.

Authors:  Alireza Naseri; Asghar Mesbahi
Journal:  Rep Pract Oncol Radiother       Date:  2010-09-22

2.  A new analytical formula for neutron capture gamma dose calculations in double-bend mazes in radiation therapy.

Authors:  Hosein Ghiasi; Asghar Mesbahi
Journal:  Rep Pract Oncol Radiother       Date:  2012-05-24

3.  Monte Carlo characterizations mapping of the (γ,n) and (n,γ) photonuclear reactions in the high energy X-ray radiation therapy.

Authors:  Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2013-07-31

4.  Safe bunker designing for the 18 MV Varian 2100 Clinac: a comparison between Monte Carlo simulation based upon data and new protocol recommendations.

Authors:  Manije Beigi; Fatemeh Afarande; Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2015-11-17

5.  Optimization of the photoneutron target geometry for e-accelerator based BNCT.

Authors:  Nahid Chegeni; Saleh Boveiry Pur; Sasan Razmjoo; Seydeh Khadijed Hoseini
Journal:  Electron Physician       Date:  2017-06-25
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.