Literature DB >> 26894340

The impact of spectral filtration on image quality in micro-CT system.

Liqiang Ren1, Muhammad U Ghani, Di Wu, Bin Zheng, Yong Chen, Kai Yang, Xizeng Wu, Hong Liu.   

Abstract

This paper aims to evaluate the impact of spectral filtration on image quality in a microcomputed tomography (micro-CT) system. A mouse phantom comprising 11rods for modeling lung, muscle, adipose, and bones was scanned with 17 s and 2min, respectively. The current (μA) for each scan was adjusted to achieve identical entrance exposure to the phantom, providing a baseline for image quality evaluation. For each region of interest (ROI) within specific composition, CT number variations, noise levels, and contrast-to-noise ratios (CNRs) were evaluated from the reconstructed images. CT number variations and CNRs for bone with high density, muscle, and adipose were compared with theoretical predictions. The results show that the impact of spectral filtration on image quality indicators, such as CNR in a micro-CT system, is significantly associated with tissue characteristics. The findings may provide useful references for optimizing the scanning parameters of general micro-CT systems in future imaging applications.

Entities:  

Mesh:

Year:  2016        PMID: 26894340      PMCID: PMC4762071          DOI: 10.1120/jacmp.v17i1.5714

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

As a scaled‐down computed tomography (CT) imaging modality, micro‐CT was initially developed for industrial imaging applications and then for small animal imaging, but it recently has proven to be a useful tool in various clinical medical applications such as the assessments of the three‐dimensional (3D) bone microstructure and bone mineral density (BMD) performed on human legs or on bone biopsy specimens, the study of microvasculature anatomy, and the imaging of tumor perfusion and angiogenesis. Micro‐CT is capable of providing 3D imaging at high resolution on the order up to 10 μm, enhancing the way of noninvasively and precisely visualizing the microarchitecture inside the object. , , , , , Characterizing and optimizing the imaging performance of micro‐CT have been widely investigated in the past decades with simulation and experimental methods. Primary studies include the determination of the radiation dose levels using Monte Carlo procedure or ion‐chamber dosimeter, the assessment of system resolution in slanted edge and plane spread function methods, the investigation of noise propagation and reduction, and other evaluations for imaging quality assurance such as geometric accuracy, linearity, CT number variation, as well as uniformity. , , , Inherent and additional filtrations are generally applied in micro‐CT to eliminate low‐energy photons that make less contribution to image quality, but tend to increase the absorbed radiation dose. , , However, their effectiveness and influence on CT number variation, noise level, and image quality have not been fully characterized to date. Though the influence of spectral filtration on patient dose, as well as image quality, has been studied on a conventional CT scanner, the imaging parameters and scenario are different as in a micro‐CT system. Therefore, it is necessary and meaningful to perform a comprehensive investigation of the impact of spectral filtrations on image quality in a micro‐CT system. In this study, with the use of two imaging phantoms, we investigated the impact of spectral filtrations, including aluminum (Al), copper (Cu) and their combinations, on CT number variation, noise level, and image quality indicators such as contrast‐to‐noise ratio (CNR). Based on the evaluations of CNRs for soft tissue, muscle, and bone with low density, we further analyzed the influence of spectral filtrations on low‐contrast resolution, an important estimator in micro‐CT imaging. , Our overall objective is to assess the feasibility of developing and applying a new evaluation tool that can provide useful evaluation results or references for selecting the optimal scanning parameters in micro‐CT systems in future imaging applications.

MATERIALS AND METHODS

Micro‐CT scanner

A rotating gantry based micro‐CT system (Quantum FX, Perkin Elmer, Waltham, MA) was used in this study. The major components of the scanner consist of a polychromatic X‐ray source, of which the anode is constructed from tungsten (W), and a flat panel X‐ray imaging detector positioned opposite to each other on the gantry with a fixed source‐to‐detector distance of 265 mm. The X‐ray source is capable of providing target voltages from 30 to 90 kVp, with a current range from 20 to . Inherent filtrations with the purpose of reducing the fluence of low‐energy photons include a thick output window of beryllium (Be), Al, and Cu. Two modes are available for this scanner, live mode and CT scan mode. The live mode is used to view the subjects in real time with/without rotating the gantry, and the exposure lasts for up to 150 s, providing the possibility of spectral measurement. In contrast, various scan settings of standard scan and fine scan are provided in the CT scan mode, depending on the selected field of view (FOV).

Spectral measurement and evaluation

X‐ray spectra were measured using a cadmium telluride (CdTe) spectrometer (XR‐100TCdTe, Amptek, Bedford, MA). The spectrometer performing peak efficiency at includes an X‐ray and gamma ray detector, a preamplifier, and a cooler system using CdTe diode detector mounted on a two‐stage thermoelectric cooler. During the spectral measurement, the spectrometer was positioned with its axis perpendicular to the imaging detector. Considering the extremely limited space inside the gantry (i.e., extremely short distance between X‐ray source and spectrometer collimator), three pinholes with small diameters and 2 mm in thickness were applied with a brass spacer (36 mm in length and 3000 μm in diameter) separating them to cover the CdTe detector, in an effort to limit the photon rate to an acceptable level and prevent spectral distortions due to pileup. As shown in Fig. 1, two pinholes on top of the spacer had and diameters, and the one below had diameter. Since the received photon rate is very sensitive to the position of the spectrometer due to the collimator restriction, a good alignment of the spectrometer–pinhole assembly to the central ray from the X‐ray focal spot, therefore, becomes essential in order to achieve an accurate spectral measurement. The best position was determined through repeatedly moving the spectrometer until finding a “sweet spot” with the highest photon rate and proper spectral shape.
Figure 1

Configuration of spectrometer collimator used for covering the CdTe detector while limiting the X‐ray photon rate to an acceptable level.

The spectra were measured with inherent filtration and various additional spectral filtrations in the live mode under the condition of 90 kVp, and 150 s. The additional spectral filtrations included Al, Cu, and their combinations: 0.5 mmAl, 1.0 mmAl, 1.5 mmAl, 2.0 mmAl, 2.5 mmAl, 3.0 mmAl, 3.5 mmAl, 4.0 mmAl, 0.2 mmCu, and . For each measuring condition, three spectra were acquired and averaged to reduce the stochastic noise. Configuration of spectrometer collimator used for covering the CdTe detector while limiting the X‐ray photon rate to an acceptable level. To evaluate the spectral measurement accuracy, energy resolution (R) of the characteristic peak was calculated using Eq. (1), where is the full width in energy unit at half maximum, and is the central energy (59.32 keV) of : The beam quality generally defined as the overall ability of an X‐ray beam to penetrate an object was evaluated by calculating the mean energy and half value layer (HVL), which provide the criteria for beam quality improvements with increasingly additional filtrations. The mean energy was directly calculated from the measured spectra using Eq. (2), where ME is the mean energy, E is the energy for each channel, and N(E) is the photon count in the channel with energy E: In contrast, the HVL was measured using sheets of Al (type‐1100) and a recently calibrated ion chamber. During the HVL measurement, the ion chamber was flatly placed inside the gantry facing upward to the X‐ray focal spot and the sensitive as well as the geometric center of the ion chamber coincided with the isocenter of the rotating gantry.

Imaging phantoms and scan protocol

Two imaging phantoms were used in this study, a uniform water phantom with a diameter of 28 mm for measuring the CT number uniformity, and a water‐filled mouse phantom (Model 091, CIRS, Norfolk, VA) for evaluating the impact of the additional filtrations on the image quality. Note that these two phantoms were filled with deionized (DI) water to avoid any unnecessary imaging uncertainties. The mouse phantom consists of 11 rods of varying mineral loading and dimension in a water‐tight, polycarbonate housing which is very durable and resistant to many chemicals. Seven rods representing different compositions (i.e., lung, muscle, adipose, and bone with various densities of 0, 50, 250, and 750 mg/cc) within mammals were selected. The varied bone densities are realized through blending hydroxyapatite (HA), the principal constituent of teeth and bones within mammals, in a soft‐tissue equivalent, polymer background. The cross section containing the rods of interest is shown in Fig. 2 with more detailed dimensions.
Figure 2

Schematic of the water‐filled mouse phantom; the length is 30 mm for the rod representing lung, while 32 mm for other rods.

Under each spectral filtration condition, the water phantom and the mouse phantom were scanned with 40 mm FOV and 90 kVp, but different currents adjusted from to to provide the same entrance exposure to the object as a baseline for CT number uniformity and image quality comparison. The entrance exposure was measured using the same ion chamber and in the same way as in the HVL measurements. Through adjusting the current while keeping the same exposure time (e.g., 12 s), similar exposure levels were achieved, and the corresponding currents were determined for different scan conditions. Note that the process of acquiring equivalent entrance exposure was performed under the micro‐CT live mode, but the determined currents were applicable to the scan mode due to the constant source‐to‐object distance. Then, two CT scans, a standard scan of 17 s and a fine scan of 2 min, were performed on both imaging phantoms under each scan protocol. Schematic of the water‐filled mouse phantom; the length is 30 mm for the rod representing lung, while 32 mm for other rods.

CT number uniformity measurement and image quality assessment

With various spectral filtrations, the reconstructed images acquired from the water phantom were analyzed to compare the CT number uniformity, while those from the mouse phantom were to compare the CT numbers, noise levels, and CNRs of regions of interest (ROIs) for different materials within the mouse phantom. Since the source spectrum was significantly modified with the interaction of spectral filtrations, CT number calibration was performed with two specifically selected ROIs, one for air and the other for water, before the image acquisition under each scanning setting. The averaged CT number for the air ROI was calibrated as −1000, while 0 for the water ROI. Then, eleven circular ROIs with different diameters were defined, four for water as well as bones, and one for lung, muscle, adipose each, as in Fig. 3. The mean value within each ROI in the current slice was regarded as the CT number, ExperCTk for that corresponding material k of interest.
Figure 3

Eleven ROIs defined for water, lung, muscle, adipose, and bones.

Eleven ROIs defined for water, lung, muscle, adipose, and bones. In addition, the CT number for each material could be theoretically predicted using Eq. (3) with the a priori knowledge of mean energy and linear attenuation coefficients (LACs): where k indicates the material of interest (e.g., lung, muscle, adipose, bones, and water) and is the theoretical CT number for material k, and and represent the LACs under the mean energy for material and background of water. In order to minimize the background information such as ring artifacts, beam hardening artifacts, streaking artifacts, and cone‐beam artifacts, two sequential acquisitions of the image were acquired, and the noise was calculated from the subtraction of one acquisition from the other. Specifically, the noise level was determined through dividing the standard deviation (SD) in the four water ROIs on the subtraction image by a factor of , as shown in Eq. (4): Then, the experimental and theoretical CNRs of each material of interest were calculated using (5), (6). In Eq. (6), the relationship between theoretical CNR and LAC under mean energy for each material was established through substituting Eq. (3) and considering . The standard derivations ( and ) of experimental and theoretical CNRs were calculated using (7), (8), where , , , and were the standard derivations of ExperCTk, ExperCTbg, Nbg, μk,ME and μbg, ME either acquired experimentally (multiple measurements) or theoretically (e.g., curve fitting). In summary, 50 consecutive slices in the reconstructed images were selected for experimental assessments of CT number, noise level, and CNRs in each test.

RESULTS

Spectra evaluation and current determinations

As shown in the second column of Table 1, all spectra were acquired with acceptable photon rates less than 2000 photons per second (p/s) corresponding to a photon flux of about so as to avoid the pileup phenomenon. For each spectrum, the channel‐to‐energy calibration were performed using two characteristic peaks from K line of W, , and . Then, energy resolution and mean energy were derived using (1), (2) from the measured spectra and listed in the third and fourth column of Table 1. Due to the restriction of the received photon rate, high energy resolutions ranging from 1.46% to 2.05% were achieved. According to the calculation of mean energy and the measurement of HVL (fifth column), the beam quality was improved with increasingly additional filtrations. In the sixth column, the current (μA) for each filtration was determined to provide identical entrance exposure, as a baseline for image quality comparison with various spectral filtrations. Each entrance exposure lasted for 12 s; the values in the last column being the average of three repeated measurements, with a maximum relative deviation below 2.1%. All measured spectra are demonstrated in Fig. 4, indicating improved beam qualities with increasingly additional filtrations.
Table 1

Spectral evaluations with respect to photon rate, energy resolution, and mean energy, HVL measurement in terms of the thickness of Al, and current determination providing identical entrance exposure; all these data were calculated or measured at 90 kVp.

Additional Filtration (mm) Photon Rate (p/s) Energy Resolution (R: %) Mean Energy (keV) HVL (mmAl) Current (μA) Entrance Exposure (12s: mR)
None 1930±55 1.65±0.15 46.68±0.11 3.4863 474±5.0
0.5Al 1723±25 1.65±0.15 47.53±0.17 3.6972 470±7.3
1.0Al 1552±29 1.60±0.18 47.99±0.07 3.8380 480±6.8
1.5Al 1397±14 1.46±0.18 48.70±0.04 4.0487 474±5.4
2.0Al 1272±32 1.89±0.00 49.01±0.17 4.3498 471±9.9
2.5Al 1198±10 1.84±0.35 49.79±0.04 4.72107 480±1.8
3.0Al 1102±12 1.65±0.15 50.15±0.08 4.94118 472±2.6
3.5Al 1051±13 1.94±0.31 50.73±0.06 5.17129 474±8.8
4.0Al 971±7 1.65±0.15 51.09±0.09 5.41140 477±4.5
0.2Cu 938±11 1.97±0.11 51.16±0.10 5.45144 472±8.3
0.2Cu+2.5Al 724±5 2.05±0.11 52.94±0.04 6.81200 474±4.0
Figure 4

Measured spectra with (a) inherent filtration only and (b)–(f) various additional filtrations.

Spectral evaluations with respect to photon rate, energy resolution, and mean energy, HVL measurement in terms of the thickness of Al, and current determination providing identical entrance exposure; all these data were calculated or measured at 90 kVp. Measured spectra with (a) inherent filtration only and (b)–(f) various additional filtrations.

CT number uniformity and image quality analysis

The CT number uniformities were compared by plotting the CT number profile of the central line in each acquired image from the water phantom. The comparison with an internal filtration only and with an additional 4 mm Al filtration is presented in Fig. 5, showing that both uniformities are acceptable. This insignificant improvement may be because that the object used in this micro‐CT imaging system is relatively small, and the internal filtration of Al and Cu has already removed most of the X‐ray photons with low energies which are the major contributor to the beam hardening effect.
Figure 5

CT number profiles using a uniform water phantom, one with internal filtration only and the other with additional 4 mm Al filtration.

Five compositions (i.e., water, air, muscle, adipose, and bone with high density) were selected for the theoretical calculations and predications of CT number and CNR. The LACs of these compositions were obtained from the National Institute of Standards and Technology (NIST, Gaithersburg, MD), but only with discrete energies (e.g., 10 keV, 20 keV). In order to acquire the LAC data under the mean energies (46.68–52.94 keV) when different spectral filtrations were utilized, a second‐order polynomial fitting method was applied on the available data at the energies of 40 keV, 50 keV, and 60 keV for each composition, and the fitting results are listed in Table 2. Corresponding CT number calculated theoretically using Eq. (3) is presented, as well. It is not surprising that the LAC for each composition is decreased with the improved beam quality due to the increased mean energy, as well as the stronger penetrability. Also, the configuration is consistent for all filtrations because the CT number in air and water are very similar or equal, while for adipose, muscle, and bone, the absolute CT numbers are monotonically decreased with increasingly additional filtrations.
Table 2

Determination of LACs and CT numbers (Hounsfield unit (HU)) using a second‐order polynomial fitting and Eq. (3).

Additional Filtration (mm) Mean Energy (keV) Water (1.00g/cm3) Air (0.001205g/cm3) Adipose (0.95g/cm3) Muscle (1.05g/cm3) Bone (1.92g/cm3)
LAC CT LAC CT LAC CT LAC CT LAC CT
None46.680.2380 2.64E1 998.892 0.209 123.304 0.25047.9890.9402944.062
0.5Al47.530.2350 2.60E1 998.893 0.207 120.025 0.24647.6600.9052848.796
1.0A147.990.2340 2.59E1 998.893 0.206 118.293 0.24547.4860.8872798.434
1.5Al48.700.2310 2.56E1 998.894 0.204 115.685 0.24247.2230.8602722.482
2.0Al49.010.2300 2.54E1 998.894 0.204 114.571 0.24147.1100.8492690.035
2.5Al49.490.2280 2.51E1 998.895 0.202 111.844 0.23846.8340.8222610.437
3.0Al50.150.2260 2.50E1 998.896 0.201 110.622 0.23746.7090.8092574.734
3.5Al50.730.2250 2.48E1 998.896 0.200 108.706 0.23546.5130.7912518.663
4.0Al51.090.2240 2.47E1 998.896 0.200 107.550 0.23446.3940.7792484.795
0.2Cu51.160.2230 2.47E1 998.897 0.199 107.328 0.23446.3720.7772478.295
0.2Cu+2.5Al 52.940.2190 2.41E1 998.898 0.196 102.045 0.22945.8240.7262322.955
CT number profiles using a uniform water phantom, one with internal filtration only and the other with additional 4 mm Al filtration. The CT numbers acquired by experiments for water, lung, adipose, muscle, and bones are shown in 3, 4 under the scan condition of 17 s and 2 min, respectively, and a reconstructed image sample of mouse phantom is shown in Fig. 6. The results of CT number determination agree with the theoretical calculations with regard to water, adipose, and bone with high density. The CT number variations of muscle and bone with low density (e.g., 50 mg/cc) are not obvious, because their LACs are quite similar as that of water and, therefore, the results are prone to be affected by the imaging noise. Since most part of the lung is composed by air, the lung CT number shows a variation tendency towards to as that of air rather than to 0.
Table 3

CT numbers (HU) determined by experiments under a standard scan of 17 s, the values below being the average derived from 50 consecutive slices in the reconstructed image. (Note: the standard deviation value is not associated with the CT number in a given slice, but is derived as the variation of CT numbers from 50 consecutive slices in the reconstructed images for each material.)

Additional Filtration (mm) Water Lung Adipose Muscle Bone 0 mg/cc Bone 50 mg/cc Bone 250 mg/cc Bone 750 mg/cc
None 1.584±1.032 631.432±108.613 113.181±109.614 33.271±107.614 42.041±107.383 82.472±108.382 591.763±111.523 1543.741±111.532
0.5Al 4.495±1.042 642.102±106.194 113.952±105.102 32.771±104.603 43.883±105.431 76.683±105.712 577.132±108.151 1501.132±107.215
1.0Al 1.768±0.982 645.553±104.812 119.803±101.974 37.184±102.762 33.924±101.572 83.321±102.541 568.721±106.052 1483.112±105.045
1.5Al 2.172±0.990 647.632±104.632 105.421±102.402 37.973±102.302 28.841±101.762 84.084±103.123 561.462±105.189 1453.204±103.741
2.0Al 0.864±1.023 647.652±102.602 104.472±102.072 37.372±101.612 24.824±100.863 87.152±97.681 552.923±103.794 1418.161±107.314
2.5Al 1.037±1.024 649.341±101.182 102.871±97.364 39.324±98.362 19.721±96.614 85.905±98.734 544.462±100.893 1401.784±101.052
3.0Al 0.910±0.988 650.512±101.193 93.814±100.253 39.903±98.712 21.342±97.264 87.922±100.041 534.581±101.052 1371.182±105.142
3.5Al 0.505±0.986 648.403±102.042 99.403±99.962 35.432±99.703 18.352±99.662 83.631±99.431 524.612±101.812 1335.183±103.653
4.0Al 0.841±1.001 649.451±101.321 98.743±99.293 37.871±99.173 17.114±98.691 87.451±98.671 517.254±100.863 1326.264±97.426
0.2Cu 2.489±2.032 656.023±103.562 90.763±100.593 40.332±99.881 1.813±98.762 92.312±99.441 512.413±100.774 1284.083±114.835
0.2Cu+2.5Al 1.000±1.965 661.301±100.762 91.132±96.582 39.713±96.582 2.652±95.743 91.123±96.812 484.612±97.962 1182.583±118.071
Table 4

CT numbers (HU) determined by experiments under a fine scan of 2 min, the values below being the average derived from 50 consecutive slices in the reconstructed image. (Note: the standard deviation value is not associated with the CT number in a given slice, but is derived as the variation of CT numbers from fifty consecutive slices in the reconstructed images for each material.)

Additional Filtration (mm) Water Lung Adipose Muscle Bone 0 mg/cc Bone 50 mg/cc Bone 250 mg/cc Bone 750 mg/cc
None 8.588±1.001 642.721±48.174 124.842±41.213 31.338±41.214 48.291±40.112 81.352±40.762 588.091±43.662 1628.621±41.832
0.5Al 8.565±0.985 651.594±47.123 121.341±38.844 29.261±38.194 45.861±37.632 81.659±39.253 581.069±42.291 1602.741±40.925
1.0Al 6.964±1.012 651.162±47.043 117.082±39.001 31.232±39.112 39.737±38.924 84.992±39.371 572.038±42.382 1572.101±40.412
1.5Al 8.016±1.032 652.902±47.021 115.743±38.384 29.663±39.532 36.921±39.333 84.403±39.443 560.463±41.942 1536.771±40.325
2.0Al 6.721±0.974 651.812±46.072 113.253±39.473 32.112±38.971 32.718±38.363 85.214±39.162 551.593±41.682 1501.982±45.565
2.5Al 4.793±0.995 651.572±46.012 109.274±38.593 32.663±38.523 29.129±37.701 86.804±38.001 544.274±40.914 1481.413±37.864
3.0Al 4.582±0.985 653.402±45.232 109.650±38.614 32.001±38.001 26.314±37.662 89.342±37.863 535.663±40.871 1447.834±43.505
3.5Al 4.780±1.015 654.602±45.241 107.251±38.623 33.602±37.612 20.749±37.802 90.403±37.871 526.824±40.302 1420.723±43.065
4.0Al 4.201±1.014 655.162±45.583 106.552±38.261 34.592±37.301 18.833±37.332 89.381±37.624 519.687±40.241 1396.775±46.562
0.2Cu 0.392±2.021 659.331±49.421 102.193±38.831 38.102±38.832 8.041±37.615 96.923±38.221 514.692±40.704 1371.889±54.912
0.2Cu+2.5Al 2.929±2.012 662.041±49.672 98.001±38.614 36.003±38.492 6.962±37.703 91.932±37.473 481.197±39.562 1268.388±57.803
Figure 6

Reconstructed image sample of mouse phantom; the gray level values are scaled for muscle and 0 mg/cc.

The results determined by Eq. (4) indicate a reduction of noise with additional spectral filtrations under both standard and fine scan, as shown in Table 5. The trends seen in the table may be caused by improved beam quality and narrower spectra with filtration. Also, the reduction of noise provides a potential to achieve improved CNR for certain materials based on the tissue characteristic.
Table 5

Noise evaluation with various spectral filtrations under the condition of standard scan of 17 s and fine scan of 2 min.

Additional filtration (mm) 17 s 2 min
None 64.987±5.718 24.111±0.829
0.5Al 64.263±5.959 23.710±0.852
1.0Al 63.534±5.585 23.424±0.712
1.5Al 63.056±5.764 23.294±0.896
2.0Al 62.571±5.583 23.120±0.746
2.5Al 61.768±5.105 22.844±0.723
3.0Al 60.900±5.591 22.737±0.808
3.5Al 60.227±5.302 22.643±0.713
4.0Al 59.907±5.478 22.453±0.779
0.2Cu 59.542±5.443 22.434±0.681
0.2Cu+2.5Al 57.924±5.091 22.322±0.725
Table 6 summarizes the theoretical CNRs and the corresponding standard derivations derived from (6), (8) for muscle, adipose, and bone , respectively, while the experimental CNRs and their standard derivations directly determined using (5), (7) for lung, muscle, adipose, and bones are summarized in 7, 8. The theoretical and experimental results have high association. Both show that the CNR for bone with higher density (e.g., , 750 mg/cc, and 250 mg/cc) decreases with additional spectral filtrations. Similar phenomenon is also observed in soft tissue, such as adipose and bone with a density of 0 mg/cc. However, opposite situations occur in lung, muscle, and bone with lower density (50 mg/cc). Thus, the imaging quality indicator in terms of CNR significantly depends on the tissue characteristic, which indicates that experimental data are needed to provide useful references or guidelines for the utilization of additional spectral filtrations in future micro‐CT imaging applications.
Table 6

Theoretical CNRs for bone, adipose, and muscle with additional filtrations.

Additional Filtration (mm) Bone (1.92g/cm3) Adipose Muscle
17 s 2 min 17 s 2 min 17 s 2 min
None 45.303±3.993 122.102±4.205 1.897±0.165 5.114±0.174 0.738±0.068 1.990±0.072
0.5Al 44.330±4.114 120.152±4.322 1.868±0.172 5.062±0.181 0.742±0.068 2.010±0.071
1.0Al 44.046±3.861 119.470±3.622 1.862±0.166 5.050±0.155 0.747±0.065 2.027±0.061
1.5Al 43.175±3.947 116.873±4.498 1.835±0.169 4.966±0.193 0.749±0.069 2.027±0.079
2.0Al 42.991±3.837 116.349±3.757 1.831±0.161 4.955±0.158 0.753±0.068 2.038±0.067
2.5Al 42.262±3.486 114.271±3.610 1.811±0.153 4.896±0.158 0.758±0.059 2.050±0.061
3.0Al 42.278±3.889 113.238±4.030 1.816±0.167 4.865±0.173 0.767±0.073 2.054±0.076
3.5Al 41.819±3.677 111.231±3.501 1.805±0.162 4.801±0.155 0.772±0.065 2.054±0.062
4.0Al 41.477±3.782 110.666±3.826 1.795±0.164 4.790±0.165 0.774±0.068 2.066±0.069
0.2Cu 41.623±3.814 110.469±3.359 1.803±0.165 4.784±0.146 0.779±0.076 2.067±0.067
0.2Cu+2.Al 40.104±3.513 104.066±3.369 1.762±0.159 4.571±0.153 0.791±0.069 2.053±0.066
Table 7

Experimental CNRs with standard scan of 17 s.

Additional Filtration (mm) Bone (750 mg/cc) Bone (250 mg/cc) Bone (50 mg/cc) Bone (0 mg/cc) Adipose Muscle Lung
None 23.779±2.706 9.130±1.894 1.293±1.672 0.623±1.653 1.717±1.693 0.536±1.657 9.692±1.876
0.5Al 23.429±2.739 9.051±1.881 1.263±1.649 0.613±1.642 1.703±1.643 0.580±1.629 9.922±1.891
1.0Al 23.371±2.637 8.979±1.846 1.339±1.618 0.506±1.599 1.700±1.612 0.613±1.618 10.132±1.875
1.5Al 23.081±2.675 8.939±1.858 1.368±1.640 0.423±1.614 1.637±1.631 0.637±1.623 10.236±1.905
2.0Al 22.678±2.652 8.850±1.837 1.407±1.566 0.383±1.612 1.656±1.638 0.611±1.609 10.337±1.881
2.5Al 22.678±2.488 8.798±1.788 1.374±1.602 0.336±1.564 1.682±1.582 0.620±1.593 10.529±1.855
3.0Al 22.500±2.692 8.763±1.844 1.429±1.648 0.365±1.597 1.555±1.652 0.640±1.622 10.697±1.930
3.5Al 22.177±2.603 8.719±1.857 1.397±1.655 0.296±1.655 1.642±1.666 0.597±1.656 10.758±1.941
4.0Al 22.153±2.598 8.648±1.860 1.474±1.653 0.272±1.653 1.634±1.664 0.646±1.656 10.827±1.960
0.2Cu 21.524±2.755 8.564±1.865 1.509±1.676 0.072±1.659 1.566±1.695 0.636±1.678 11.060±2.012
0.2Cu+2.5Al 20.399±2.715 8.349±1.844 1.556±1.677 0.063±1.653 1.591±1.700 0.668±1.668 11.434±2.009
Table 8

Experimental CNRs with fine scan of 2 min.

Additional Filtration (mm) Bone (750 mg/cc) Bone (250 mg/cc) Bone (50 mg/cc) Bone (0 mg/cc) Adipose Muscle Lung
None 67.902±2.908 24.747±2.001 3.730±1.695 1.647±1.664 4.821±1.717 1.656±1.666 26.300±2.193
0.5Al 67.959±2.991 24.869±1.995 3.805±1.661 1.573±1.588 4.756±1.647 1.595±1.612 27.120±2.214
1.0Al 67.413±2.679 24.719±1.959 3.926±1.685 1.399±1.662 4.701±1.671 1.631±1.670 27.502±2.175
1.5Al 66.316±3.083 24.404±2.031 3.967±1.700 1.241±1.689 4.624±1.657 1.617±1.698 27.684±2.282
2.0Al 65.254±2.884 24.148±1.964 3.976±1.699 1.124±1.660 4.608±1.714 1.680±1.686 27.901±2.187
2.5Al 65.058±2.643 24.035±1.946 4.009±1.668 1.065±1.651 4.573±1.695 1.639±1.687 28.312±2.204
3.0Al 63.878±2.968 23.760±1.986 4.131±1.672 0.956±1.657 4.621±1.706 1.609±1.685 28.535±2.233
3.5Al 62.954±2.748 23.477±1.927 4.203±1.678 0.705±1.670 4.525±1.712 1.695±1.662 28.698±2.193
4.0Al 62.396±2.997 23.333±1.966 4.168±1.682 0.652±1.663 4.558±1.711 1.728±1.662 28.992±2.265
0.2Cu 61.169±3.071 22.960±1.943 4.338±1.709 0.341±1.676 4.538±1.736 1.716±1.748 29.372±2.376
0.2Cu+2.5Al 56.954±3.183 21.688±1.907 4.205±1.684 0.181±1.689 4.259±1.735 1.744±1.725 29.527±2.423
Determination of LACs and CT numbers (Hounsfield unit (HU)) using a second‐order polynomial fitting and Eq. (3). CT numbers (HU) determined by experiments under a standard scan of 17 s, the values below being the average derived from 50 consecutive slices in the reconstructed image. (Note: the standard deviation value is not associated with the CT number in a given slice, but is derived as the variation of CT numbers from 50 consecutive slices in the reconstructed images for each material.) CT numbers (HU) determined by experiments under a fine scan of 2 min, the values below being the average derived from 50 consecutive slices in the reconstructed image. (Note: the standard deviation value is not associated with the CT number in a given slice, but is derived as the variation of CT numbers from fifty consecutive slices in the reconstructed images for each material.) Reconstructed image sample of mouse phantom; the gray level values are scaled for muscle and 0 mg/cc. Noise evaluation with various spectral filtrations under the condition of standard scan of 17 s and fine scan of 2 min. Theoretical CNRs for bone, adipose, and muscle with additional filtrations. Experimental CNRs with standard scan of 17 s. Experimental CNRs with fine scan of 2 min.

DISCUSSION

Since certain bone disorders may be directly or indirectly associated with decreased BMD, better visualization of bone with low density is of significant importance in this situation. As reported in 7, 8, the CNRs of imaging low density bone (50 mg/cc), when additional filtration applied, were increased by up to 20.29% and 16.29% for 17s and 2 min scan, respectively, compared to the condition with no additional filtrations, since the imaging noise was greatly suppressed. Similarly, the CNRs of imaging muscle were increased by up to 24.61% and 5.31% for 17 s and 2 min scan, indicating that applying additional filtrations to the micro‐CT imaging system utilized in this study may also be beneficial for imaging muscle related disorders. Another potential clinical application of micro‐CT imaging technique is to detect and characterize breast cancer through performing a micro‐CT imaging on breast biopsy specimens. , For this clinical utilization, the application of additional filtrations may provide better low‐contrast resolution when imaging microcalcification with low density or imaging tumors with similar density (CT number) as that of muscle. Therefore, additional filtrations would be recommended under the aforementioned micro‐CT imaging conditions. Conversely, no additional filtrations would be recommended if the soft tissue, such as adipose, is preferentially to be visualized because the CNRs of imaging adipose are continuously decreased by up to 6.86% and 11.66% with maximum filtration applied. Though the impact of applying additional filtrations on imaging bones with high density and lung seems extremely significant compared to imaging others, it may not be the decisive factor, since their CNRs with respect to water or soft‐tissue background are always sufficient. In this study, we demonstrated a unique experimental approach to validate the accuracy and/or reliability of applying the theoretical models developed from conventional CT systems to a dedicated micro‐CT system. Comparing the experimental results with the results predicted by the theoretical models, we found that, as in the derivations of CT number for water, air, muscle, adipose, and bone, the corresponding LACs under the mean energy for each material were acquired by a second‐order polynomial fitting of the existing data. In this way, the CT number variation of certain material with known characteristic could be determined theoretically. However, establishing an accurate theoretical model of predicting and calculating the CNR is difficult because the noise level which is central for CNR determination often needs to be detected and measured by experiments rather than using mathematical calculation. Thus, it would be useful and meaningful to develop a noise evaluation model for this micro‐CT scanner using an experimental data‐driven learning approach, making it possible to predict CNR variation for various scanned materials with modified beam quality. An important issue that may affect the theoretical evaluations of CT numbers and CNRs is the utilization of the mean energy of the acquired spectrum instead of the effective energy, which is defined as the monoenergetic beam of photons that has the same HVL as the investigating spectrum of photons. This is because acquiring the effective energy and the HVL, which are related to not only the spectrum property but also the material characteristic, for each specific material seems difficult or even impossible. Nevertheless, it might account for the observation that the CNRs derived in theory and by experiment are slightly inconsistent, especially for muscle and adipose. Another reason is that the muscle and adipose used for theoretical calculations may not have identical features, such as density and homogeneity, as for experimental investigations. Spectral measurement in a micro‐CT system poses great challenges due to the limited space inside the gantry. Though the photon rates in this study have been successfully reduced to an acceptable level of less than 2000 p/s through using smaller pinhole collimators (i.e., the combination of and ), the process of spectrometer adjustment is highly reliant on the operation experience and may not be effective and reproducible each time. A possible alternative is to first measure the Compton scattered photons from a given sample (e.g., polymethylmetacrylate (PMMA), polyethylene, and carbon electrode) with known property. Then, the actual spectrum incident upon the scattering sample can be reconstructed from the scattered spectrum using an energy correction and the Klein‐Nishina function. This indirect method effectively avoids the pileup phenomenon, saves the space, and more importantly, is easily repeatable with larger pinholes such as or . Also, previously reported spectral modeling techniques can be utilized to generate the X‐ray spectrum and to predict the impact of spectral modifications on imaging quality indicators, such as CNR discussed in this study. However, the specificity of the modeling for certain micro‐CT system needs to be verified by experiments before they can be applied for theoretical calculation and analysis.

CONCLUSIONS

This study investigated the impact of additional spectral filtrations on image quality in a micro‐CT system. The evaluations of CT number, noise level, and CNR were performed using a water‐filled mouse phantom. The experimental results agreed well with the theoretical calculations that, with a baseline of identical entrance exposure to the imaged mouse phantom, the CNRs were degraded with improved beam quality for bone with high density and soft tissue, while those were enhanced for bone with low density, lung, and muscle. The findings in this study may provide useful references for optimizing the scanning parameters of general micro‐CT systems in future imaging applications.

ACKNOWLEDGMENTS

This research is supported in part by National Institutes of Health (NIH), RO1 CA193378, and supported in part by a grant from the University of Oklahoma Charles and Peggy Stephenson Cancer Center funded by the Oklahoma Tobacco Settlement Endowment Trust. The authors would like to acknowledge the support of Charles and Jean Smith Chair endowment fund, as well.
  17 in total

Review 1.  Evaluation of microvascular anatomy by micro-CT.

Authors:  A Lerman; E L Ritman
Journal:  Herz       Date:  1999-11       Impact factor: 1.443

2.  A comparison of two methods for measuring the signal to noise ratio on MR images.

Authors:  M J Firbank; A Coulthard; R M Harrison; E D Williams
Journal:  Phys Med Biol       Date:  1999-12       Impact factor: 3.609

Review 3.  High resolution X-ray computed tomography: an emerging tool for small animal cancer research.

Authors:  M J Paulus; S S Gleason; S J Kennel; P R Hunsicker; D K Johnson
Journal:  Neoplasia       Date:  2000 Jan-Apr       Impact factor: 5.715

4.  Small-animal X-ray dose from micro-CT.

Authors:  John M Boone; Orlando Velazquez; Simon R Cherry
Journal:  Mol Imaging       Date:  2004-07       Impact factor: 4.488

5.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

6.  Compton-scattering measurement of diagnostic x-ray spectrum using high-resolution Schottky CdTe detector.

Authors:  Koji Maeda; Masao Matsumoto; Akira Taniguchi
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

7.  Quantitative analysis of micro-CT imaging and histopathological signatures of experimental arthritis in rats.

Authors:  Matthew D Silva; Anneli Savinainen; Rasesh Kapadia; Jason Ruan; Elizabeth Siebert; Nicole Avitahl; Rebecca Mosher; Karen Anderson; Bruce Jaffee; Lisa Schopf; Sudeep Chandra
Journal:  Mol Imaging       Date:  2004-10       Impact factor: 4.488

8.  High-resolution X-ray microtomography for the detection of lung tumors in living mice.

Authors:  Nora M De Clerck; Kris Meurrens; Horst Weiler; Dirk Van Dyck; Greet Van Houtte; Piter Terpstra; Andrei A Postnov
Journal:  Neoplasia       Date:  2004 Jul-Aug       Impact factor: 5.715

9.  A flat-panel detector based micro-CT system: performance evaluation for small-animal imaging.

Authors:  Sang Chul Lee; Ho Kyung Kim; In Kon Chun; Myung Hye Cho; Soo Yeol Lee; Min Hyoung Cho
Journal:  Phys Med Biol       Date:  2003-12-21       Impact factor: 3.609

10.  Volumetric computed tomography (VCT): a new technology for noninvasive, high-resolution monitoring of tumor angiogenesis.

Authors:  Fabian Kiessling; Susanne Greschus; Matthias P Lichy; Michael Bock; Christian Fink; Silvia Vosseler; Jens Moll; Margareta M Mueller; Norbert E Fusenig; Horst Traupe; Wolfhard Semmler
Journal:  Nat Med       Date:  2004-09-07       Impact factor: 53.440

View more
  2 in total

1.  Practical alignment method for X-ray spectral measurement in micro-CT system based on 3D printing technology.

Authors:  Liqiang Ren; Di Wu; Yuhua Li; Bin Zheng; Yong Chen; Kai Yang; Hong Liu
Journal:  Biomed Phys Eng Express       Date:  2016-07-15

Review 2.  Guidelines for Micro-Computed Tomography Analysis of Rodent Dentoalveolar Tissues.

Authors:  Michael B Chavez; Emily Y Chu; Vardit Kram; Luis F de Castro; Martha J Somerman; Brian L Foster
Journal:  JBMR Plus       Date:  2021-03-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.