Kylie Pearson1, Matthew Long2, Josephine Prasad2, Ye Ying Wu3, Michael Bonifacio2.
Abstract
BACKGROUND: The manual Generation II (Gen II) ELISA method used to measure Anti-Müllerian Hormone (AMH) from Beckman Coulter has recently been superseded by a fully automated AMH immunoassay. The aim of this study was to evaluate the performance of the Access AMH assay and directly compare it to the modified Gen II ELISA method. A secondary aim was to verify that the fertile age-related AMH range previously established using the Gen II ELISA could be used to interpret results from the new automated Access assay.Entities:
Mesh:
Substances:
Year: 2016 PMID: 26879773 PMCID: PMC4754992 DOI: 10.1186/s12958-016-0143-3
Source DB: PubMed Journal: Reprod Biol Endocrinol ISSN: 1477-7827 Impact factor: 5.211
AMH intra- and inter-assay precision on Access2 and DxI800 analysers
| Intra-assay precision | Inter-assay precision | ||||||
|---|---|---|---|---|---|---|---|
| Mean, pmol/L | SD, pmol/L | CV, % | Mean, pmol/L | SD, pmol/L | CV, % | ||
| Access2 | Control 1 | 6.27 | 0.09 | 1.51 | 6.23 | 0.24 | 3.85 |
| Control 2 | 31.78 | 0.45 | 1.41 | 31.69 | 0.96 | 3.04 | |
| Control 3 | 97.85 | 1.64 | 1.68 | 97.19 | 3.35 | 3.45 | |
| DxI800 | Control 1 | 6.40 | 0.21 | 3.30 | 6.31 | 0.28 | 4.36 |
| Control 2 | 33.36 | 0.68 | 2.04 | 32.49 | 1.87 | 5.76 | |
| Control 3 | 103.18 | 2.98 | 2.89 | 98.54 | 4.52 | 4.59 | |
AMH stability under different storage conditions
| Stored at 4 °C | Freeze/thaw | ||||||
|---|---|---|---|---|---|---|---|
| Mean, pmol/L | SD, pmol/L | Deviation, % | Mean, pmol/L | SD, pmol/L | Deviation, % | ||
| Day 1 | 5.96 | 0.16 | 0.00 | Fresh | 5.96 | 0.16 | 0.00 |
| Day 2 | 5.83 | 0.09 | −2.29 | x1 | 5.96 | 0.08 | −0.11 |
| Day 3 | 5.77 | 0.03 | −3.24 | x2 | 5.87 | 0.04 | −1.57 |
| Day 4 | 5.92 | 0.04 | −0.67 | x3 | 5.91 | 0.14 | −0.89 |
| Day 5 | 5.75 | 0.18 | −3.58 | x4 | 5.74 | 0.09 | −3.80 |
| Day 6 | 6.09 | 0.05 | 2.07 | x5 | 5.81 | 0.18 | −2.63 |
| Day 7 | 5.75 | 0.01 | −3.52 | x6 | 5.74 | 0.04 | −3.75 |
| Day 8 | 5.83 | 0.22 | −2.24 | x7 | 5.78 | 0.07 | −3.13 |
Fig. 1Linearity of the Access AMH assay on the Access2 and DxI800 analysers using a. calibrator material or b. patient sera. c. Linearity at the low end of the concentration range using patient sera. Regression line (blue), 95 % confidence intervals (red). All results are shown in pmol/L
Comparison of different dilution methods using the Access AMH Assay
| Dilution in patient sera | Dilution in sample diluent A | ||||||
|---|---|---|---|---|---|---|---|
| Dilution | Mean, pmol/L | CV, % | Deviation, % | Mean, pmol/L | CV, % | Deviation, % | |
| Sample 1 | Neat | 76.90 | 0.50 | 0.00 | 76.90 | 0.50 | 0.00 |
| 1:5 | 74.82 | 1.09 | −2.71 | 68.90 | 1.73 | −10.41 | |
| 1:10 | 74.63 | 1.59 | −2.96 | 66.28 | 2.74 | −13.81 | |
| 1:16 | 77.81 | 1.11 | 1.18 | 68.04 | 1.36 | −11.52 | |
| Sample 2 | Neat | 170.64 | 1.83 | 0.00 | 170.64 | 1.83 | 0.00 |
| 1:5 | 177.69 | 1.10 | 4.13 | 164.44 | 1.24 | −3.63 | |
| 1:10 | 176.71 | 0.64 | 3.56 | 154.80 | 0.46 | −9.28 | |
| 1:16 | 179.53 | 3.55 | 5.21 | 153.65 | 3.07 | −9.96 | |
Fig. 2a. Correlation between Access AMH and Gen II ELISAs on the Access2 and DxI800 instruments (n = 142). b. Correlation between results from Access2 analysers in different locations (n = 46) and Access2 and DxI800 analysers (n = 142). Upper panels represent the Passing-Bablok diagram with the regression line (blue) and the 95 % confidence interval (red). Lower panels represent the Bland-Altman plots with mean (blue) and 95 % confidence intervals (red). All results are shown in pmol/L
Fig. 3Correlation between automated AMH and Gen II ELISA normal female reference ranges. Individual results (blue points) and 10th, 50th and 90th percentile median values for 5-year age groups for Access AMH and Gen II ELISA methods (n = 489)