Josephine Hyldgaard1,2, Pinar Bor1, Hans Jakob Ingerslev3, Niels Tørring4.
Abstract
BACKGROUND: Anti Mullerian hormone (AMH) has previously been measured using a manual method, but a fully automated assay from Roche Diagnostics was recently introduced. The aim of this study was to compare the results from the AMH gen II ELISA and Elecsys Cobas AMH methods in a clinical setting to evaluate whether the assays achieve the goals of analytical performance. A prospective observational study with 23 women seeking laparoscopic sterilization was conducted. Blood samples were collected preoperatively as well as 1 week and 1, 3 and 6 months postoperatively; they were evaluated with the AMH gen II ELISA and Elecsys Cobas AMH methods. The assays were validated according to the optimal performance of biochemical assays: CV Analytical < 0.25* CV Within Biological Variation.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26394617 PMCID: PMC4580367 DOI: 10.1186/s12958-015-0101-5
Source DB: PubMed Journal: Reprod Biol Endocrinol ISSN: 1477-7827 Impact factor: 5.211
Performance of Beckmann Coulter AMH Gen II ELISA and Elecsys Roche AMH assay
| ELISA ( | Cobas ( | |||
|---|---|---|---|---|
| Level | CV% | Level | CV% | |
| Control 1 | 40 pmol/l | 5.2 | 40 pmol/l | 3.3 |
| Control 2 | 20 pmol/l | 6.2 | - | - |
| Control 3 | 7 pmol/l | 10.3 | 7 pmol/l | 2.8 |
| Total CV% (<10 pmol/l) ( | 31.9 (11.9–47.3) | 28.0 (9.5–43.0 | ||
| Total CV% (>10 pmol/l) ( | 22.0 (11.9–35.2) | 21.3 (11.6–33.6) | ||
| Percentage of samples without quantitation | 15 % | 2 % | ||
Controls were analyzed on consecutive days of analysis over a period of 3 months. Data are given as CV% (Standard deviation x 100/ mean). Results of variance of clinical samples are given in two levels: < and >10 pmol/l (because of bias between the two methods, the level differ between the two groups.). Data are given as CV% (range). Percentage of samples without quantiation is determined as the percentage of results < 3 pmol/l (ELISA), and < 0.5 pmol/l (Elecsys Cobas)
Fig. 1Comparison between the Beckmann Coulter AMH Gen II ELISA and Elecsys Roche AMH (Anti Müllerian Hormone) assays. a: Passing-Bablok plot displaying the regression (solid line) and 95 % confidence interval (dotted lines) and b: Bias plot displaying the difference between methods (as a percentage). Results are given in pmol/l