Literature DB >> 26868850

Image-Guided High-Dose Rate Brachytherapy in Cervix Carcinoma Using Balloon Catheter and Belt Immobilization System.

Qiyong Fan1,2, Anamaria R Yeung2, Robert Amdur2, Richard Helmig2, Justin Park2, Jonathan Li2, Darren Kahler2, Chihray Liu2, Bo Lu2.   

Abstract

PURPOSE: The efficacy of image-guided high-dose rate brachytherapy for cervical cancer is limited by the ineffective rectal sparing devices available commercially and the potential applicator movement. We developed a novel device using a balloon catheter and a belt immobilization system, serving for rectal dose reduction and applicator immobilization purposes, respectively.
METHODS: The balloon catheter is constructed by gluing a short inflatable tube to a long regular open-end catheter. Contrast agent (10) cm3 is injected into the inflatable end, which is affixed to the tandem and ring applicator, to displace the posterior vaginal wall. The belt immobilization system consists of a specially designed bracket that can hold and fix itself to the applicator, a diaper-like Velcro fastener package used for connecting the patient's pelvis to the bracket, and a buckle that holds the fasteners to stabilize the whole system. The treatment data for 21 patients with cervical cancer using both balloon catheter and belt immobilization system were retrospectively analyzed. Computed tomography and magnetic resonance images, acquired about 30 minutes apart, were registered to evaluate the effectiveness of the immobilization system.
RESULTS: In comparison with a virtual rectal blade, the balloon decreased the rectal point dose by 34% ± 4.2% (from 276 ± 57 to 182 ± 38 cGy), corresponding to an extra sparing distance of 7.9 ± 1.1 mm. The maximum sparing distance variation per patient is 1.4 ± 0.6 mm, indicating the high interfractional reproducibility for rectum sparing. With the immobilization system, the mean translational and rotational displacements of the applicator set are <3 mm and <1.5°, respectively, in all directions.
CONCLUSIONS: The rectal balloon provides significant dose reduction to the rectum and it may potentially minimize patient discomfort. The immobilization system permits almost no movement of the applicator during treatment. This work has the potential to be promoted as a standardized solution for high-dose rate treatment of cervical cancer.

Entities:  

Keywords:  applicator immobilization; cervical cancer; high-dose rate brachytherapy; rectal sparing; tandem and ring applicator

Mesh:

Substances:

Year:  2016        PMID: 26868850      PMCID: PMC5616039          DOI: 10.1177/1533034616631122

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  21 in total

1.  The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix.

Authors:  S Nag; B Erickson; B Thomadsen; C Orton; J D Demanes; D Petereit
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-08-01       Impact factor: 7.038

2.  Dose-volume histogram parameters and late side effects in magnetic resonance image-guided adaptive cervical cancer brachytherapy.

Authors:  Petra Georg; Stefan Lang; Johannes C A Dimopoulos; Wolfgang Dörr; Alina E Sturdza; Daniel Berger; Dietmar Georg; Christian Kirisits; Richard Pötter
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-10       Impact factor: 7.038

3.  Patient outcome study of concurrent chemoradiation, external beam radiotherapy, and high-dose rate brachytherapy in locally advanced carcinoma of the cervix.

Authors:  Jennifer Louise Forrest; Ida Ackerman; Lisa Barbera; Elizabeth A Barnes; Melanie Davidson; Alex Kiss; Gillian Thomas
Journal:  Int J Gynecol Cancer       Date:  2010-08       Impact factor: 3.437

4.  Proposed guidelines for image-based intracavitary brachytherapy for cervical carcinoma: report from Image-Guided Brachytherapy Working Group.

Authors:  Subir Nag; Higinia Cardenes; Silvia Chang; Indra J Das; Beth Erickson; Geoffrey S Ibbott; Jessica Lowenstein; Joye Roll; Bruce Thomadsen; Mahesh Varia
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-11-15       Impact factor: 7.038

5.  Rectal dose sparing with a balloon catheter and ultrasound localization in conformal radiation therapy for prostate cancer.

Authors:  Rakesh R Patel; Nigel Orton; Wolfgang A Tomé; Rick Chappell; Mark A Ritter
Journal:  Radiother Oncol       Date:  2003-06       Impact factor: 6.280

6.  Design and development of an inflatable latex balloon to reduce rectal and bladder doses for patients undergoing high dose rate brachytherapy.

Authors:  P Raghukumar; Raghu Ram K Nair; Abi S Aprem; Saju Bhasi; Suja Sisupal; V Padmanbhan
Journal:  J Med Phys       Date:  2009-01

7.  Inter- and intrafractional tumor and organ movement in patients with cervical cancer undergoing radiotherapy: a cinematic-MRI point-of-interest study.

Authors:  Philip Chan; Robert Dinniwell; Masoom A Haider; Young-Bin Cho; David Jaffray; Gina Lockwood; Wilfred Levin; Lee Manchul; Anthony Fyles; Michael Milosevic
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-12-31       Impact factor: 7.038

8.  Significant rectal and bladder dose reduction via utilization of Foley balloon catheters in high-dose-rate tandem and ovoid intracavitary brachytherapy of the uterine cervix.

Authors:  Tony Y Eng; Clifton D Fuller; Sean X Cavanaugh; Melissa M Blough; Amir Sadeghi; Terence Herman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-05-01       Impact factor: 7.038

9.  Clinical impact of computed tomography-based image-guided brachytherapy for cervix cancer using the tandem-ring applicator - the Addenbrooke's experience.

Authors:  L T Tan; C E Coles; C Hart; E Tait
Journal:  Clin Oncol (R Coll Radiol)       Date:  2008-12-19       Impact factor: 4.126

10.  Bladder-rectum spacer balloon in high-dose-rate brachytherapy in cervix carcinoma.

Authors:  Bhavana Rai; Firuza D Patel; Santam Chakraborty; Suresh C Sharma; Rakesh Kapoor; Abi Santhosh Aprem
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-04-01       Impact factor: 7.038

View more
  2 in total

1.  Rectum Protection by Rectal Gel Injection in Cervical Cancer Brachytherapy: A Dosimetric Study via Deformable Surface Dose Accumulation and Machine-Learning-Based Discriminative Modeling.

Authors:  Xuetao Wang; Bailin Zhang; Qiang He; Yilin Kong; Zhenhui Dai; Haoyu Meng; Fangjun Huang; Shengfeng Zhang; Yuanhu Zhu; Xiang Tan; Xin Zhen
Journal:  Front Oncol       Date:  2021-04-16       Impact factor: 6.244

2.  Comparison of the Dosimetric Influence of Applicator Displacement on 2D and 3D Brachytherapy for Cervical Cancer Treatment.

Authors:  Ailin Wu; Du Tang; Aidong Wu; Yunqin Liu; Liting Qian; Lei Zhu
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec
  2 in total