Literature DB >> 26848762

[Multiple inappropriate defibrillator shocks due to insulation failure of a Biotronik Linox defibrillator lead with externalized conductor].

Hamdi Elfarra1, Rainer Moosdorf1, Leszek Rybinski1, Wolfram Grimm2.   

Abstract

In this article the case of a patient who received a total of 35 inappropriate defibrillator shocks due to insulation failure with externalized conductor of a Biotronik Linox® lead is described. The implanted defibrillator was immediately inactivated and the failed lead was extracted using a laser sheath system.

Entities:  

Keywords:  Complications; Defibrillator lead; Inappropriate shocks; Insulation failure; Sudden cardiac death

Mesh:

Year:  2016        PMID: 26848762     DOI: 10.1007/s00399-016-0418-6

Source DB:  PubMed          Journal:  Herzschrittmacherther Elektrophysiol        ISSN: 0938-7412


  10 in total

1.  Longitudinal follow-up of Riata leads reveals high annual incidence of new conductor externalization and electrical failure.

Authors:  Christian Steinberg; Jean-François Sarrazin; François Philippon; Jean Champagne; Marc-André Bouchard; Franck Molin; Isabelle Nault; Louis Blier; Gilles O'Hara
Journal:  J Interv Card Electrophysiol       Date:  2014-11-16       Impact factor: 1.900

2.  Repetitive inappropriate implantable cardioverter-defibrillator shocks due to insulation failure with externalized conductor cables of a Biotronik Linox SD ICD lead.

Authors:  Tobias Reichlin; Michael Kühne; Christian Sticherling
Journal:  Europace       Date:  2015-10-07       Impact factor: 5.214

3.  Transvenous extraction of implantable cardioverter-defibrillator leads under advisory--a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads.

Authors:  Michael P Brunner; Edmond M Cronin; Jessen Jacob; Valeria E Duarte; Khaldoun G Tarakji; David O Martin; Thomas Callahan; P Peter Borek; Daniel J Cantillon; Mark J Niebauer; Walid I Saliba; Mohamed Kanj; Oussama Wazni; Bryan Baranowski; Bruce L Wilkoff
Journal:  Heart Rhythm       Date:  2013-06-28       Impact factor: 6.343

Review 4.  [ICD leads].

Authors:  Carsten W Israel; Mohamed Karim Sheta
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06

5.  Failure of fluoroscopy to detect "inside-out" insulation failure and externalized conductors in a Riata ICD lead.

Authors:  Edmond M Cronin; Bryan J Baranowski; David O Martin
Journal:  Heart Rhythm       Date:  2012-11-30       Impact factor: 6.343

6.  Outcomes of Sprint Fidelis and Riata lead extraction: Data from 2 high-volume centers.

Authors:  Mikhael F El-Chami; Faisal M Merchant; Mathew Levy; Mian Bilal Alam; Rohit Rattan; Michael H Hoskins; Jonathan J Langberg; David Delurgio; Michael S Lloyd; Angel R Leon; Samir Saba
Journal:  Heart Rhythm       Date:  2015-05-18       Impact factor: 6.343

7.  Conductor externalization of the Biotronik Kentrox internal cardioverter-defibrillator lead: the tip of another iceberg?

Authors:  Bernard Abi-Saleh; Marwan M Refaat; Maurice Khoury; Bruce Wilkoff
Journal:  Heart Rhythm       Date:  2014-05-10       Impact factor: 6.343

8.  Insulation failure and externalized conductor of a single-coil Kentrox lead: an ongoing story?

Authors:  Harilaos Bogossian; Dejan Mijic; Gerrit Frommeyer; Joachim Winter
Journal:  J Cardiovasc Electrophysiol       Date:  2014-11-12

9.  Early failure of the Biotronik Linox implantable cardioverter defibrillator lead.

Authors:  Gareth J Padfield; Christian Steinberg; Shahzad S Karim; Stanley Tung; Matthew T Bennett; John P Le Maitre; Jamil Bashir
Journal:  J Cardiovasc Electrophysiol       Date:  2014-12-22

Review 10.  Insulation Failure of the Linox Defibrillator Lead: A Case Report and Retrospective Review of a Single Center Experience.

Authors:  Andrew J Howe; Nicholas A McKeag; Carol M Wilson; Kyle P Ashfield; Michael J Roberts
Journal:  J Cardiovasc Electrophysiol       Date:  2015-04-18
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.