Harilaos Bogossian1, Dejan Mijic, Gerrit Frommeyer, Joachim Winter.
Abstract
Conductor externalization is a frequent complication with the St. Jude Medical Riata lead. Single case reports also reported externalization of conductors for dual-coil Biotronik leads. Up to now, conductor externalization has not yet been reported for any single coil leads. We report for the first time an externalization of conductors in a Biotronik Kentrox single-coil implantable cardioverter defibrillator (ICD) lead.Entities:
Keywords: Kentrox lead; externalized conduction; implantation/extraction; insulation failure
Mesh:
Year: 2014 PMID: 25302563 DOI: 10.1111/jce.12562