Insulation Failure of the Linox Defibrillator Lead: A Case Report and Retrospective Review of a Single Center Experience.

Andrew J Howe1, Nicholas A McKeag1, Carol M Wilson1, Kyle P Ashfield1, Michael J Roberts1.   

Abstract

Implantable cardioverter defibrillator (ICD) lead insulation failure and conductor externalization have been increasingly reported. The 7.8F silicon-insulated Linox SD and Linox S ICD leads (Biotronik, Berlin, Germany) were released in 2006 and 2007, respectively, with an estimated 85,000 implantations worldwide. A 39-year-old female suffered an out-of-hospital ventricular fibrillation (VF) arrest with successful resuscitation. An ICD was implanted utilizing a single coil active fixation Linox(Smart) S lead (Biotronik, Berlin, Germany). A device-triggered alert approximately 3 years after implantation confirmed nonphysiological high rate sensing leading to VF detection. A chest X-ray showed an abnormality of the ICD lead and fluoroscopic screening confirmed conductor externalization proximal to the defibrillator coil. In view of the combined electrical and fluoroscopic abnormalities, urgent lead extraction and replacement were performed. A review of Linox (Biotronik) and Vigila (Sorin Group, Milan, Italy) lead implantations within our center (n = 98) identified 3 additional patients presenting with premature lead failure, 2 associated with nonphysiological sensed events and one associated with a significant decrease in lead impedance. All leads were subsequently removed and replaced. This case provides a striking example of insulation failure affecting the Linox ICD lead and, we believe, is the first to demonstrate conductor externalization manifesting both electrical and fluoroscopic abnormalities.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  Biotronik; Linox lead failure; implantable cardioverter defibrillator; insulation failure; lead failure

Mesh:

Year:  2015        PMID: 25711237     DOI: 10.1111/jce.12638

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

1.  What did we learn from Riata™?

Authors:  Avi Fischer; C Jenney
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06-09

2.  [Multiple inappropriate defibrillator shocks due to insulation failure of a Biotronik Linox defibrillator lead with externalized conductor].

Authors:  Hamdi Elfarra; Rainer Moosdorf; Leszek Rybinski; Wolfram Grimm
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-02-05

3.  First time description of early lead failure of the Linox Smart lead compared to other contemporary high-voltage leads.

Authors:  Vanessa Weberndörfer; Tobias Nyffenegger; Ian Russi; Miriam Brinkert; Benjamin Berte; Stefan Toggweiler; Richard Kobza
Journal:  J Interv Card Electrophysiol       Date:  2018-05-01       Impact factor: 1.900

Review 4.  Externalized conductors and insulation failure in Biotronik defibrillator leads: History repeating or a false alarm?

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Clin Cases       Date:  2017-02-16       Impact factor: 1.337

5.  Inside-out abrasion and contained conductor cable externalization in a defibrillation lead with asymmetric conductor cable lumen distribution.

Authors:  Ernest W Lau; Michael J D Roberts
Journal:  HeartRhythm Case Rep       Date:  2018-01-11
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.