Daniel Engelman1,2,3, Joseph H Kado4,5, Bo Reményi6,7, Samantha M Colquhoun8,9,10, Jonathan R Carapetis11,12, Nigel J Wilson13,14, Susan Donath15,16, Andrew C Steer17,18,19.
Abstract
BACKGROUND: Echocardiographic screening for rheumatic heart disease (RHD) has the potential to detect subclinical cases for secondary prevention, but is constrained by inadequate human resources in most settings. Training non-expert health workers to perform focused cardiac ultrasound (FoCUS) may enable screening at a population-level. We aimed to evaluate the quality and agreement of FoCUS for valvular regurgitation by briefly trained health workers.Entities:
Mesh:
Year: 2016 PMID: 26830341 PMCID: PMC4736281 DOI: 10.1186/s12872-016-0205-7
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Fig. 1Map of study sites in Fiji. Pins indicate school assessment sites
Twelve step focused cardiac ultrasound protocol for assessment of regurgitation
| View | Valve | Mode | Identify regurgitation | Measure longest jet |
|---|---|---|---|---|
| 1. PLAX | Mitral | 2D | ||
| Mitral | CD | Yes | Yes | |
| Aortic | 2D | |||
| Aortic | CD | Yes | Yes | |
| 2. PSAX | Aortic | 2D | ||
| Aortic | CD | Yes | No | |
| Mitral | 2D | |||
| Mitral | CD | Yes | No | |
| 3. Apical | Mitral | 2D | ||
| Mitral | CD | Yes | Yes | |
| Aortic | 2D | |||
| Aortic | CD | Yes | Yes |
PLAX parasternal long-axis view, PSAX, parasternal short-axis view, 2D two-dimensional imaging mode, CD, color Doppler mode
Quality of focused cardiac ultrasound by nurses
| View | Quality | Nurse 1 | Nurse 2 | Nurse 3 | Nurse 4 | Nurse 5 | Nurse 6 | Nurse 7 | Total | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| n = 250 | n = 264 | n = 340 | n = 340 | n = 274 | n = 311 | n = 239 | n = 2018 | ||||||||||
| n | % | n | % | n | % | n | % | n | % | n | % | n | % | n | % | ||
| Overall | 1 | 248 | 99.2 | 216 | 81.8 | 338 | 99.4 | 337 | 99.1 | 266 | 97.1 | 308 | 99.0 | 236 | 98.7 | 1949 | 96.6 |
| 2 | 2 | 0.8 | 42 | 15.9 | 0 | 0.0 | 2 | 0.6 | 7 | 2.6 | 3 | 1.0 | 2 | 0.8 | 58 | 2.9 | |
| 3 | 0 | 0.0 | 6 | 2.3 | 2 | 0.6 | 1 | 0.3 | 1 | 0.4 | 0 | 0.0 | 1 | 0.4 | 11 | 0.5 | |
| MV 2D | 1 | 247 | 98.8 | 204 | 77.3 | 337 | 99.1 | 337 | 99.1 | 265 | 96.7 | 308 | 99.0 | 235 | 98.3 | 1933 | 95.8 |
| 2 | 3 | 1.2 | 59 | 22.3 | 1 | 0.3 | 3 | 0.9 | 9 | 3.3 | 3 | 1.0 | 4 | 1.7 | 82 | 4.0 | |
| 3 | 0 | 0.0 | 1 | 0.4 | 2 | 0.6 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 | 3 | 0.2 | |
| MV CD | 1 | 249 | 99.6 | 210 | 79.5 | 337 | 99.1 | 334 | 98.2 | 256 | 93.4 | 308 | 97.1 | 228 | 95.4 | 1916 | 94.9 |
| 2 | 1 | 0.4 | 48 | 18.2 | 1 | 0.3 | 5 | 1.5 | 17 | 6.2 | 3 | 2.9 | 10 | 4.2 | 91 | 4.5 | |
| 3 | 0 | 0.0 | 6 | 2.3 | 2 | 0.6 | 1 | 0.3 | 1 | 0.4 | 0 | 0.0 | 1 | 0.4 | 11 | 0.5 | |
| AV 2D | 1 | 244 | 97.6 | 215 | 81.4 | 329 | 96.7 | 325 | 95.6 | 268 | 97.8 | 302 | 99.0 | 236 | 98.7 | 1925 | 95.4 |
| 2 | 6 | 2.4 | 49 | 18.6 | 9 | 2.7 | 15 | 4.4 | 5 | 1.8 | 9 | 1.0 | 3 | 1.3 | 90 | 4.5 | |
| 3 | 0 | 0.0 | 0 | 0.0 | 2 | 0.6 | 0 | 0.0 | 1 | 0.4 | 0 | 0.0 | 0 | 0.0 | 3 | 0.1 | |
| AV CD | 1 | 248 | 99.2 | 224 | 84.8 | 338 | 99.4 | 340 | 100.0 | 266 | 97.1 | 308 | 99.4 | 236 | 98.3 | 1960 | 97.1 |
| 2 | 2 | 0.8 | 38 | 14.4 | 0 | 0.0 | 0 | 0.0 | 7 | 2.6 | 3 | 0.6 | 3 | 1.7 | 53 | 2.6 | |
| 3 | 0 | 0.0 | 2 | 0.8 | 2 | 0.6 | 0 | 0.0 | 1 | 0.4 | 0 | 0.0 | 0 | 0.0 | 5 | 0.3 | |
Quality: 1, adequate for diagnosis; 2, poor but diagnosis made; 3, not interpretable. AV aortic valve, MV mitral valve, 2D two-dimensional mode, CD color Doppler mode
Agreement between nurse and cardiologist on the presence of regurgitation
| View | Jet length (cm) | Mitral kappa (CI) | Aortic kappa (CI) |
|---|---|---|---|
| One view | >0 | 0.60 (0.56, 0.64) | 0.44 (0.32, 0.57) |
| PLAX | >0 | 0.72 (0.70, 0.74) | 0.54 (0.42, 0.62) |
| PSAX | >0 | 0.55 (0.53, 0.57) | 0.53 (0.45, 0.61) |
| Apical | >0 | 0.61 (0.58, 0.64) | 0.44 (0.33, 0.58) |
| Two views | >0 | 0.75 (0.71, 0.80) | 0.61 (0.46, 0.77) |
| Two views | 0.5 | 0.76 (0.71, 0.80) | 0.64 (0.49, 0.79) |
| Two views | 1.0 | 0.78 (0.74, 0.83) | 0.65 (0.49, 0.81) |
| Two views | 1.5 | 0.81 (0.76, 0.86) | n/aa |
| Two views | 2.0 | 0.78 (0.71, 0.86) | n/aa |
CI 95 % confidence interval, PLAX parasternal long axis, PSAX parasternal short axis. Jet length refers to minimum jet length in at least one view. aAnalysis was not performed for jet lengths greater than the World Heart Federation diagnostic cut-off point
Fig. 2Difference versus mean (Bland-Altman) plots comparing cardiologist and nurse measurements of regurgitation length. a: Mitral regurgitation; b: Aortic regurgitation. ULA and LLA indicate upper and lower 95 % limits of agreement
Fig. 3Examples of cardiac ultrasound performed by nurses. a. Parasternal long axis view demonstrating mitral regurgitation. This image was categorised adequate quality. b. Apical 5 chamber view demonstrating aortic regurgitation. The regurgitation was not identified by the nurse. This image was categorised poor quality. c and d. Parasternal long axis view demonstrating closing volume. Color is seen across the mitral valve in single frame at the time of valve closure, but not in mid-systole