INTRODUCTION: Navigated procedures in spinal surgery have been established due to an increasing demand for precision. Especially, 3D C-arms connected to navigation systems are being used more often and can be utilised intraoperatively for the planning and controlling of screw positions. This prospective study analyses our experiences with 3D-based navigation in posterior stabilisations in the cervical and thoracic spine.
METHODS: A 3D C-Arm (Ziehm Vision Vario 3D(®)) was connected to a navigation system (VectorVision, Brainlab(®)) and used for the placement of, in total, 451 screws among 67 patients. Of those, 14 patients had to undergo operations in the cervical and 53 in the thoracic spine. Postoperatively, the positioning was observed with computed tomography (CT).
RESULTS: The application time is approximately 6 min. In total, 354/451 (78.5%) screws could be inserted assisted with navigation, and 272/451 (60.3%) were controlled intraoperatively. Regarding the cervical spine, in 87.1% (61/70) of the screws, the navigation procedure was uneventful. The positioning of 63.2% (43/68) of the screws was checked intraoperatively. In the upper thoracic spine, 77% (293/381) could be placed with navigation and 59.6% (227/381) were controlled intraoperatively. Occasionally, the scanning setup was problematic. Correct placement was seen in 92.7% of screws; for the remaining screws, no revision was needed.
CONCLUSIONS: Intraoperative 3D imaging navigation for posterior spinal stabilisations is technically feasible and reliable in clinical use. The image quality depends on the individual bone density. With undisturbed visibility of the vertebral body, the reliability of 3D-based navigation is comparable to that of CT-based procedures. Additionally, it has the advantage of skipping the preoperative acquisition of data as well as the matching process, with reduced radiation doses.

Entities:  

Keywords:  3D navigation; Cervical and thoracic spine; Image-guided surgery

Year:  2011        PMID: 26814949     DOI: 10.1007/s00068-011-0098-1

Source DB:  PubMed          Journal:  Eur J Trauma Emerg Surg        ISSN: 1863-9933            Impact factor:   3.693


  36 in total

1.  Intraoperative Iso-C C-arm navigation in craniospinal surgery: the first 60 cases.

Authors:  Jonathan S Hott; Vivek R Deshmukh; Jeffrey D Klopfenstein; Volker K H Sonntag; Curtis A Dickman; Robert F Spetzler; Stephen M Papadopoulos
Journal:  Neurosurgery       Date:  2004-05       Impact factor: 4.654

2.  [Intraoperative three-dimensional navigation for pedicle screw placement].

Authors:  P A Grützner; T Beutler; K Wendl; J von Recum; A Wentzensen; L-P Nolte
Journal:  Chirurg       Date:  2004-10       Impact factor: 0.955

3.  Transpedicular screw placement evaluated by axial computed tomography of the cervical pedicle.

Authors:  Takeshi Sakamoto; Masashi Neo; Takashi Nakamura
Journal:  Spine (Phila Pa 1976)       Date:  2004-11-15       Impact factor: 3.468

4.  Clinical accuracy of three-dimensional fluoroscopy-based computer-assisted cervical pedicle screw placement: a retrospective comparative study of conventional versus computer-assisted cervical pedicle screw placement.

Authors:  Yoshimoto Ishikawa; Tokumi Kanemura; Go Yoshida; Zenya Ito; Akio Muramoto; Shuichiro Ohno
Journal:  J Neurosurg Spine       Date:  2010-11

5.  Navigation with three-dimensional rotational radiographic data for transpedicular percutaneous needle introduction: feasibility and comparison with fluoroscopic guidance.

Authors:  Theo van Walsum; Everine B van de Kraats; Maurits H J Voormolen; Jorrit-Jan Verlaan; Willem P Th M Mali; Wiro J Niessen
Journal:  J Vasc Interv Radiol       Date:  2006-09       Impact factor: 3.464

6.  Iso-C3D fluoroscopy-based navigation in direct pedicle screw fixation of Hangman fracture: a case report.

Authors:  S Rajasekaran; S Vidyadhara; Ajoy P Shetty
Journal:  J Spinal Disord Tech       Date:  2007-12

7.  Timing of paired points and surface matching registration in three-dimensional (3D) image-guided spinal surgery.

Authors:  Eric W Nottmeier; Tracey L Crosby
Journal:  J Spinal Disord Tech       Date:  2007-06

8.  Morphologic characteristics of human cervical pedicles.

Authors:  E E Karaikovic; M D Daubs; R W Madsen; R W Gaines
Journal:  Spine (Phila Pa 1976)       Date:  1997-03-01       Impact factor: 3.468

9.  Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine.

Authors:  L P Amiot; K Lang; M Putzier; H Zippel; H Labelle
Journal:  Spine (Phila Pa 1976)       Date:  2000-03-01       Impact factor: 3.468

10.  An anatomic, radiographic, and biomechanical assessment of extrapedicular screw fixation in the thoracic spine.

Authors:  M Dvorak; S MacDonald; K R Gurr; S I Bailey; R G Haddad
Journal:  Spine (Phila Pa 1976)       Date:  1993-09-15       Impact factor: 3.468

View more
  6 in total

Review 1.  [Intraoperative 3D imaging in spinal surgery].

Authors:  O Gonschorek; S Hauck; V Bühren
Journal:  Unfallchirurg       Date:  2016-10       Impact factor: 1.000

2.  Anatomical considerations for insertion of pedicular screw in cervicothoracic junction.

Authors:  Morteza Faghih-Jouibari; Keisan Moazzeni; Amir Amini-Navai; Sara Hanaei; Sina Abdollahzadeh; Ramin Khanmohammadi
Journal:  Iran J Neurol       Date:  2016-10-07

3.  [Multiplanar reconstruction with mobile 3D image intensifier. Surgical treatment of proximal humerus fractures].

Authors:  P Hepp; J Theopold; J-S Jarvers; B Marquaß; N von Dercks; C Josten
Journal:  Unfallchirurg       Date:  2014-05       Impact factor: 1.000

4.  Navigation-assisted surgery for tumors of the spine.

Authors:  S Bandiera; R Ghermandi; A Gasbarrini; G Barbanti Bròdano; S Colangeli; S Boriani
Journal:  Eur Spine J       Date:  2013-09-24       Impact factor: 3.134

5.  Detection of articular perforations of the proximal humerus fracture using a mobile 3D image intensifier - a cadaver study.

Authors:  Jan Theopold; Kevin Weihs; Christine Feja; Bastian Marquaß; Christoph Josten; Pierre Hepp
Journal:  BMC Med Imaging       Date:  2017-08-01       Impact factor: 1.930

6.  State Recognition of Bone Drilling Based on Acoustic Emission in Pedicle Screw Operation.

Authors:  Fengqing Guan; Yu Sun; Xiaozhi Qi; Ying Hu; Gang Yu; Jianwei Zhang
Journal:  Sensors (Basel)       Date:  2018-05-09       Impact factor: 3.576

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.