Precise 3D/2D calibration between a RGB-D sensor and a C-arm fluoroscope.

Xiang Wang1,2, Severine Habert3, Meng Ma3, Chun-Hao Huang3, Pascal Fallavollita3, Nassir Navab3,4.   

Abstract

PURPOSE: Calibration and registration are the first steps for augmented reality and mixed reality applications. In the medical field, the calibration between an RGB-D camera and a C-arm fluoroscope is a new topic which introduces challenges.
METHOD: A convenient and efficient calibration phantom is designed by combining the traditional calibration object of X-ray images with a checkerboard plane. After the localization of the 2D marker points in the X-ray images and the corresponding 3D points from the RGB-D images, we calculate the projection matrix from the RGB-D sensor coordinates to the X-ray, instead of estimating the extrinsic and intrinsic parameters simultaneously. VALIDATION: In order to evaluate the effect of every step of our calibration process, we performed five experiments by combining different steps leading to the calibration. We also compared our calibration method to Tsai's method to evaluate the advancement of our solution. At last, we simulated the process of estimating the rotation movement of the RGB-D camera using MATLAB and demonstrate that calculating the projection matrix can reduce the angle error of the rotation.
RESULTS: A RMS reprojection error of 0.5 mm is achieved using our calibration method which is promising for surgical applications. Our calibration method is more accurate when compared to Tsai's method. Lastly, the simulation result shows that using a projection matrix has a lower error than using intrinsic and extrinsic parameters in the rotation estimation.
CONCLUSIONS: We designed and evaluated a 3D/2D calibration method for the combination of a RGB-D camera and a C-arm fluoroscope.

Entities:  

Keywords:  3D/2D calibration; C-arm fluoroscope; Calibration phantom; RGB-D sensor

Mesh:

Year:  2016        PMID: 26811080     DOI: 10.1007/s11548-015-1347-2

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  14 in total

1.  Surface transparency makes stereo overlays unpredictable: the implications for augmented reality.

Authors:  Laura G Johnson; Philip Edwards; David Hawkes
Journal:  Stud Health Technol Inform       Date:  2003

2.  Time-of-Flight Sensor Calibration for a Color and Depth Camera Pair.

Authors:  Jiyoung Jung; Joon-Young Lee; Yekeun Jeong; In So Kweon
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  2015-07       Impact factor: 6.226

3.  DVV: a taxonomy for mixed reality visualization in image guided surgery.

Authors:  Marta Kersten-Oertel; Pierre Jannin; D Louis Collins
Journal:  IEEE Trans Vis Comput Graph       Date:  2012-02       Impact factor: 4.579

4.  Intra-operative augmented reality in distal locking.

Authors:  Roberto Londei; Marco Esposito; Benoit Diotte; Simon Weidert; Ekkehard Euler; Peter Thaller; Nassir Navab; Pascal Fallavollita
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-03-27       Impact factor: 2.924

5.  Intra-op measurement of the mechanical axis deviation: an evaluation study on 19 human cadaver legs.

Authors:  Lejing Wang; Pascal Fallavollita; Alexander Brand; Okan Erat; Simon Weidert; Peter-Helmut Thaller; Ekkehard Euler; Nassir Navab
Journal:  Med Image Comput Comput Assist Interv       Date:  2012

6.  New method for geometric calibration and distortion correction of conventional C-arm.

Authors:  Cai Meng; Jun Zhang; Fugen Zhou; Tianmiao Wang
Journal:  Comput Biol Med       Date:  2014-06-25       Impact factor: 4.589

7.  Multi-modal intra-operative navigation during distal locking of intramedullary nails.

Authors:  Benoit Diotte; Pascal Fallavollita; Lejing Wang; Simon Weidert; Ekkehard Euler; Peter Thaller; Nassir Navab
Journal:  IEEE Trans Med Imaging       Date:  2014-10-02       Impact factor: 10.048

8.  Seven years of clinical experience with teleconsultation in craniomaxillofacial surgery.

Authors:  Rolf Ewers; Kurt Schicho; Arne Wagner; Gerhard Undt; Rudolf Seemann; Michael Figl; Michael Truppe
Journal:  J Oral Maxillofac Surg       Date:  2005-10       Impact factor: 1.895

9.  Prospective motion correction of X-ray images for coronary interventions.

Authors:  Guy Shechter; Barak Shechter; Jon R Resar; Rafael Beyar
Journal:  IEEE Trans Med Imaging       Date:  2005-04       Impact factor: 10.048

10.  Image intensifier distortion correction for fluoroscopic RSA: the need for independent accuracy assessment.

Authors:  Angela E Kedgley; Anne-Marie V Fox; Thomas R Jenkyn
Journal:  J Appl Clin Med Phys       Date:  2012-01-05       Impact factor: 2.102

View more
  1 in total

1.  Toward an End-to-End Calibration for Mobile C-Arm in Combination with a Depth Sensor for Surgical Augmented Reality Applications.

Authors:  Sahar Hosseinian; Hossein Arefi; Nassir Navab
Journal:  Sensors (Basel)       Date:  2019-12-19       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.