Accuracy of 3D fluoroscopy-navigated anterior transpedicular screw insertion in the cervical spine: an experimental study.

Jan Bredow1, Carolin Meyer2, Max Joseph Scheyerer2, Florian Siedek3, Lars Peter Müller2, Peer Eysel2, Gregor Stein2.   

Abstract

PURPOSE: The technique of pedicle screw stabilization is finding increasing popularity for use in the cervical spine. Implementing anterior transpedicular screws (ATPS) in cervical spine offers theoretical advantages compared to posterior stabilization. The goal of the current study was the development of a new setting for navigated insertion of ATPS, combining the advantage of reduced invasiveness of an anterior approach with the technical advantages of navigation.
METHODS: 20 screws were implanted in levels C3 to C6 of four cervical spine models (SAWBONES(®) Cervical Vertebrae with Anterior Ligament) with the use of 3D fluoroscopy navigation system [Arcadis Orbic 3D, Siemens and VectorVision fluoro 3D trauma software (BrainLAB)]. The accuracy of inserted screws was analyzed according to postoperative CT scans and following the modified Gertzbein and Robbins classification.
RESULTS: 20 anterior pedicle screws were placed in four human cervical spine models. Of these, eight screws were placed in C3, two screws in C4, six screws in C5, and four screws in C6. 16 of 20 screws (80 %) reached a grade 1 level of accuracy according to the modified Gertzbein and Robbins Classification. Three screws (15 %) were grade 2, and one screw (5 %) was grade 3. Grade 4 and 5 positions were not evident. Summing grades 1 and 2 together as "good" positions, 95 % of the screws achieved this level. Only a single screw did not fulfill these criteria.
CONCLUSION: The setting introduced in this study for navigated insertion of ATPS into cervical spine bone models is well implemented and shows excellent results, with an accuracy of 95 % (Gertzbein and Robbins grade 2 or better). Thus, this preliminary study represents a prelude to larger studies with larger case numbers on human specimens.

Entities:  

Keywords:  Accuracy of screw insertion; Anterior transpedicular screws; Navigated screw insertion; Spine surgery

Mesh:

Year:  2016        PMID: 26810977     DOI: 10.1007/s00586-016-4403-x

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  31 in total

1.  Computer-assisted posterior instrumentation of the cervical and cervico-thoracic spine.

Authors:  Marcus Richter; Thomas Mattes; Balkan Cakir
Journal:  Eur Spine J       Date:  2003-11-22       Impact factor: 3.134

2.  Load sharing properties of cervical pedicle screw-rod constructs versus lateral mass screw-rod constructs.

Authors:  Bradley J Dunlap; Eldin E Karaikovic; Hyung-Soon Park; Mark J Sokolowski; Li-Qun Zhang
Journal:  Eur Spine J       Date:  2010-02-02       Impact factor: 3.134

3.  Cervical anterior transpedicular screw fixation (ATPS)--Part II. Accuracy of manual insertion and pull-out strength of ATPS.

Authors:  Heiko Koller; Frank Acosta; Mark Tauber; Michael Fox; Hudelmaier Martin; Rosmarie Forstner; Peter Augat; Rainer Penzkofer; Christian Pirich; H Kässmann; Herbert Resch; Wolfgang Hitzl
Journal:  Eur Spine J       Date:  2008-01-26       Impact factor: 3.134

4.  Adverse events in surgically treated cervical spondylopathic myelopathy: a prospective validated observational study.

Authors:  Dennis Hartig; Juliet Batke; Nicolas Dea; Adrienne Kelly; Charles Fisher; John Street
Journal:  Spine (Phila Pa 1976)       Date:  2015-03-01       Impact factor: 3.468

5.  Pedicle screw fixation in the cervical spine.

Authors:  Rongming Xu; Nabil A Ebraheim; Martin Skie
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2008-08

6.  Accuracy of fluoroscopy versus computer-assisted navigation for the placement of anterior cervical pedicle screws.

Authors:  Andrew G Patton; Randal P Morris; Yong-Fang Kuo; Ronald W Lindsey
Journal:  Spine (Phila Pa 1976)       Date:  2015-04-01       Impact factor: 3.468

7.  [Trans-articular screw fixation of C1/C2 in atlanto-axial instability. Comparison between percutaneous and open procedures].

Authors:  M Blauth; M Richter; U Lange
Journal:  Orthopade       Date:  1999-08       Impact factor: 1.087

8.  Complications of cervical pedicle screw fixation for nontraumatic lesions: a multicenter study of 84 patients.

Authors:  Hiroaki Nakashima; Yasutsugu Yukawa; Shiro Imagama; Tokumi Kanemura; Mitsuhiro Kamiya; Makoto Yanase; Keigo Ito; Masaaki Machino; Go Yoshida; Yoshimoto Ishikawa; Yukihiro Matsuyama; Naoki Ishiguro; Fumihiko Kato
Journal:  J Neurosurg Spine       Date:  2011-12-16

9.  The three column spine and its significance in the classification of acute thoracolumbar spinal injuries.

Authors:  F Denis
Journal:  Spine (Phila Pa 1976)       Date:  1983 Nov-Dec       Impact factor: 3.468

10.  Anterior cervical pedicle screw and plate fixation using fluoroscope-assisted pedicle axis view imaging: a preliminary report of a new cervical reconstruction technique.

Authors:  Yasutsugu Yukawa; Fumihiko Kato; Keigo Ito; Hiroaki Nakashima; Masaaki Machino
Journal:  Eur Spine J       Date:  2009-04-03       Impact factor: 3.134

View more
  2 in total

1.  Accuracy of 3D fluoro-navigated anterior transpedicular screws in the subaxial cervical spine: an experimental study on human specimens.

Authors:  Jan Bredow; C Meyer; F Siedek; W F Neiss; L Löhrer; L P Müller; P Eysel; G Stein
Journal:  Eur Spine J       Date:  2017-07-27       Impact factor: 3.134

Review 2.  Progress of the Anterior Transpedicular Screw in Lower Cervical Spine: A Review.

Authors:  Yuan-Wei Zhang; Ting Zeng; Wen-Cheng Gao; Xin Xiao; Yan Xiao; Xi Chen; Su-Li Zhang; Liang Deng
Journal:  Med Sci Monit       Date:  2019-08-21
  2 in total