Kyeong-Bong Lee1, Min-Kyu Kim1, Ju-Ri Jeong1, Wan-Hee Lee1.
Abstract
BACKGROUND: The purpose of this study was to examine the inter- and intra-rater reliability of an electronic inspiratory loading device for the assessment of pulmonary functions: maximum inspiratory pressure, peak inspiratory flow, and vital capacity. MATERIAL/Entities:
Mesh:
Year: 2016 PMID: 26782369 PMCID: PMC4725617 DOI: 10.12659/msm.895573
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
General characteristics of subjects.
| Characteristic | Mean ±SD |
|---|---|
| Gender, | 26/24 |
| Age, years | 55.96±12.81 |
| Height, cm | 164.77±6.97 |
| Weight, kg | 63.11±10.53 |
| BMI, kg/m2 | 23.12±2.68 |
| Affected side, | 29/21 |
| Duration, days | 228.66±150.82 |
SD – standard deviation; BMI – body mass index.
Intra-rater reliability between measures for the variables.
| Variables | Mean ±SD | ICC | 95% CI | SEM | MDC | |
|---|---|---|---|---|---|---|
| 1st test | 2nd test | |||||
| Maximal inspiratory pressure (cmH2O) | 52.02±20.49 | 53.52±21.06 | .986 | .976–.992 | 4.380 | 7.370 |
| Peak inspiratory flow ( | 2.92±1.20 | 2.97±1.20 | .992 | .986–.995 | 0.003 | 0.011 |
| Vital capacity ( | 1.99±0.69 | 1.99±0.58 | .959 | .928–.977 | 0.122 | 0.057 |
SD – standard deviation; ICC – intraclass correlation coefficient; 95% CI – 95% confidence interval; SEM – standard error of the mean; MDC – minimal detectable change.
Inter-rater reliability between examiners for the variables.
| Variables | Mean ±SD | ICC | 95% CI | SEM | MDC | |
|---|---|---|---|---|---|---|
| E1 | E2 | |||||
| Maximal inspiratory pressure (cmH2O) | 52.02±20.49 | 52.09±21.40 | .984 | .971–.991 | 3.575 | 4.912 |
| Peak inspiratory flow ( | 2.92±1.20 | 3.01±1.24 | .985 | .974–.992 | 0.006 | 0.000 |
| Vital capacity ( | 1.99±0.69 | 1.99±0.58 | .933 | .881–.962 | 0.179 | 0.123 |
E1 – examiner 1; E2 – examiner 2; SD – standard deviation; ICC – intraclass correlation coefficient; 95% CI – 95% confidence interval; SEM – standard error of the mean; MDC – minimal detectable change.
Figure 1Bland-Altman plots illustrating the measurement difference of maximal inspiratory pressure and peak inspiratory flow volume (A): Maximal inspiratory at intra-rater measurement (B): Maximal inspiratory at inter-rater measurement (C): Peak inspiratory flow volume at intra-rater measurement (D): Peak inspiratory flow volume at inter-rater measurement (E): Vital capacity at intra-rater measurement (F): Vital capacity at inter-rater measurement