Pieter W J Lozekoot1, Monique M J de Jong1, Sandro Gelsomino2,3, Orlando Parise1, Francesco Matteucci1, Fabiana Lucà1, N Kumar1, Jan Nijs4, Jens Czapla4, Paul Kwant1, Daniele Bani5, Gian Franco Gensini6, Laurent Pison1, Harry J G M Crijns1, Jos G Maessen1, Mark La Meir1,4.
Abstract
PURPOSE: Data on epicardial contact force efficacy in dual epicardial-endocardial atrial fibrillation ablation procedures are lacking. We present an in vitro study on the importance of epicardial and endocardial contact forces during this procedure.Entities:
Keywords: Atrial arrhythmias; Atrial fibrillation; Catheter ablation; Computer model simulation; Contact force
Mesh:
Year: 2016 PMID: 26728030 PMCID: PMC4757611 DOI: 10.1007/s10840-015-0089-y
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.900
Fig. 1a–b The ABLA-BOX. 1 plexiglass box, 2 plexiglass septum, 3 endocardial compartment, 4 epicardial compartment, 5 12-mm trocar, 6 linear lesion entrance, 7 thermocouple entrance, 8 rail for thermocouples, 9 magnetic tissue holder, 10 stirring motor, 11 outlet, 12 heating mat, 13 inlet
Fig. 2The ABLA-BOX setup. 1 plexiglass box, 2 force transducer, 3 catheter holder, 4 catheter holder rail, 5 digital forces monitor, 6 heating mat temperature controller, 7 stirring motor controller, 8 catheter holder fine-tuning system
Fig. 3The in vitro circuit explained. A roller pump circulates blood through a circuit consisting of a heat exhanger and the ABLA-BOX
Lesion dimension by contact force
| Epicardial CF | |||||
|---|---|---|---|---|---|
| 100 g | 300 g | 500 g | |||
| 20 g | Max diam. area (mm) | 6.72 ± 2.15* | |||
| Min diam. area (mm) | NT | NT | 5.23 ± 1.82* | ||
| Total area (mm2) | 39.61 ± 23.25* | ||||
| Total volume (mm3) | 3.96 ± 2.32* | ||||
| Endocardial CF | 30 g | Max diam. area (mm) | 6.97 ± 1.51 | 6.72 ± 2.15* | 7.85 ± 1.83* |
| Min diam. area (mm) | 5.42 ± 1.28 | 5.23 ± 1.82* | 6.27 ± 1.60†‡ | ||
| Total area (mm2) | 40.30 ± 16.20 | 39.61 ± 23.25* | 53.44 ± 24.15†‡ | ||
| Total volume (mm3) | 4.03 ± 1.62 | 3.96 ± 2.32* | 5.34 ± 2.4†‡ | ||
| 40 g | Max diam. area (mm) | 7.29 ± 2.15 | 7.71 ± 1.96* | 8.46 ± 1.93†‡ | |
| Min diam. area (mm) | 5.70 ± 1.89 | 6.07 ± 1.70* | 6.73 ± 1.71†‡ | ||
| Total area (mm2) | 46.26 ± 29.58 | 50.79 ± 24.18* | 60.96 ± 27.18†‡ | ||
| Total volume (mm3) | 4.63 ± 2.96 | 5.08 ± 2.42* | 6.10 ± 2.72†‡ | ||
| 50 g | Max diam. area (mm) | 8.42 ± 2.29 | 8.91 ± 1.75* | 9.46 ± 1.87* | |
| Min diam. area (mm) | 6.70 ± 2.05 | 7.12 ± 1.56* | 7.37 ± 1.68* | ||
| Total area (mm2) | 61.79 ± 35.36 | 66.90 ± 27.18* | 71.68 ± 29.45* | ||
| Total volume (mm3) | 6.18 ± 3.54 | 6.69 ± 2.72* | 7.17 ± 2.95* | ||
CF contact force, NT not transmural
*No significant difference mean epicardial contact forces, †significant difference means epicardial contact force 500 g vs. epicardial contact force 100 g, ‡significant difference means epicardial contact force 500 g vs. epicardial contact force 300 g; p ≤ 0.05
Fig. 4a Samples of middle sections of the left atrial tissue after in vitro hybrid ablation, employing different contact forces, cryopreserved and cut in consecutive sections. One hundred-micrometer thick, parallel to the endocardial surface. See text. b Ablation lesion formation progression during different tissue depths in a non-transmural ablation sample (endocardial force 20 g, epicardial 100 g), cryopreserved and cut in consecutive sections, 100-μm thick, parallel to the endocardial surface. c Ablation lesion formation progression during different tissue depths in a transmural ablation sample (endocardial force 30 g, epicardial 100 g), cryopreserved and cut in consecutive sections, 100-μm thick, parallel to the endocardial surface
Mean tissue thickness of the different used tissue samples
| Epicardial CF | ||||
|---|---|---|---|---|
| 100 g (mm) | 300 g (mm) | 500 g (mm) | ||
| Epicardial CF | 20 g | 3.9 ± 0.7 | 4.8 ± 0.5§ | 4.5 ± 0.2* |
| 30 g | 4.3 ± 0.1 | 4.4 ± 0.1* | 4.1 ± 0.3* | |
| 40 g | 5.0 ± 0.5 | 4.5 ± 0.4* | 4.3 ± 0.3† | |
| 50 g | 4.3 ± 0.3 | 4.3 ± 0.2* | 4.6 ± 0.1* | |
CF contact force
*No significant difference mean epicardial contact forces, †significant difference means epicardial contact force 500 g vs. epicardial contact force 100 g, ‡significant difference means epicardial contact force 500 g vs. epicardial contact force 300 g, §significant difference means epicardial contact force 300 g vs. epicardial contact force 100 g; p ≤ 0.05