Radiofrequency catheter ablation using cooled electrodes: impact of irrigation flow rate and catheter contact pressure on lesion dimensions.

Christian Weiss1, Matthias Antz, Olaf Eick, Kablai Eshagzaiy, Thomas Meinertz, Stephan Willems.   

Abstract

Irrigation of radiofrequency current (RF) ablation reduces the risk of thrombus formation. The aim of this study was to investigate the impact of different irrigation catheter flow rates and contact pressures from the catheter on the development of lesion dimension and thrombus formation. A thigh muscle preparation was achieved in six sheep to create a cradle that was filled and perfused with heparinized blood (250 mL/min, 37 C degrees). RF ablation (30 s, 30 W) was initially performed with three different irrigation flow rates (5 mL/min, 10 mL/min, and 20 mL/min) and a perpendicular position (0.1 N contact pressure) of the irrigated ablation catheter ("Sprinklr," Medtronic, Inc., Minneapolis, MN, USA). The next lesions were induced with constant contact pressure of 0.05 Newton (N); 0.1 N; 0.3 and 0.5 N and a parallel or perpendicular orientation of the catheter, respectively. A constant irrigation flow of 10 mL/min was maintained during these RF applications. Cross sections of the lesions were investigated with regard to maximal depth and maximal diameter at and below the surface. During high flow irrigation (20 mL/min) the surface diameter was significantly smaller (0.63 +/- 0.1 cm) compared to irrigation flowrates of 5 mL/min (0.88 +/- 0.2 cm) and 10 mL/min (1 +/- 0.1 cm). Thrombus formation was not observed during any RF application. Only in perpendicular catheter orientations with a contact pressure of 0.5 N were significantly deeper lesions (0.85 +/- 0.12 cm) induced compared to 0.05 N (0.55 +/- 0.02 cm), 0.1 N (0.7 +/- 0.01 cm) and 0.3 N (0.67 +/- 0.01 cm) contact pressure. There was no significant difference in lesion depth with different flow rates. Irrigated RF ablation even with low flow rates and high catheter contact pressure prevented thrombus formation at the electrode. Smaller lesion diameters have been created with high irrigation flow rates. The deeper lesion created with high catheter contact pressure might be caused by a greater power transmission to the tissue.

Entities:  

Mesh:

Year:  2002        PMID: 11991372     DOI: 10.1046/j.1460-9592.2002.00463.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  18 in total

1.  Measuring luminal esophageal temperature during pulmonary vein isolation of atrial fibrillation.

Authors:  Daisuke Sato; Kunihiro Teramoto; Hiroki Kitajima; Naoto Nishina; Yoshitomi Kida; Hiroki Mani; Masahiro Esato; Yeong-Hwa Chun; Toshiji Iwasaka
Journal:  World J Cardiol       Date:  2012-05-26

2.  Catheter ablation of an incessant ventricular tachycardia originating from the left aortic sinus cusp in an adolescent with subacute myocarditis.

Authors:  Klaus Kettering; Christoph Kampmann; Hanke Mollnau; Karl-Friedrich Kreitner; Thomas Münzel; Christian Weiss
Journal:  Clin Res Cardiol       Date:  2008-10-13       Impact factor: 5.460

3.  2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.

Authors:  Edmond M Cronin; Frank M Bogun; Philippe Maury; Petr Peichl; Minglong Chen; Narayanan Namboodiri; Luis Aguinaga; Luiz Roberto Leite; Sana M Al-Khatib; Elad Anter; Antonio Berruezo; David J Callans; Mina K Chung; Phillip Cuculich; Andre d'Avila; Barbara J Deal; Paolo Della Bella; Thomas Deneke; Timm-Michael Dickfeld; Claudio Hadid; Haris M Haqqani; G Neal Kay; Rakesh Latchamsetty; Francis Marchlinski; John M Miller; Akihiko Nogami; Akash R Patel; Rajeev Kumar Pathak; Luis C Saenz Morales; Pasquale Santangeli; John L Sapp; Andrea Sarkozy; Kyoko Soejima; William G Stevenson; Usha B Tedrow; Wendy S Tzou; Niraj Varma; Katja Zeppenfeld
Journal:  J Interv Card Electrophysiol       Date:  2020-10       Impact factor: 1.900

4.  Comparison of the change in the dimension of the pulmonary vein ostia immediately after pulmonary vein isolation for atrial fibrillation-open irrigated-tip catheters versus non-irrigated conventional 4 mm-tip catheters.

Authors:  Takashi Yamamoto; Takumi Yamada; Yukihiko Yoshida; Yasuya Inden; Naoya Tsuboi; Hirohiko Suzuki; Monami Ando; Masayuki Shimano; Rei Shibata; Haruo Hirayama; Toyoaki Murohara
Journal:  J Interv Card Electrophysiol       Date:  2014-07-16       Impact factor: 1.900

5.  Simultaneous assessment of contact pressure and local electrical coupling index using robotic navigation.

Authors:  Antonio Dello Russo; Gaetano Fassini; Michela Casella; Fabrizio Bologna; Osama Al-Nono; Daniele Colombo; Viviana Biagioli; Pasquale Santangeli; Luigi Di Biase; Martina Zucchetti; Benedetta Majocchi; Vittoria Marino; Joseph J Gallinghouse; Andrea Natale; Claudio Tondo
Journal:  J Interv Card Electrophysiol       Date:  2014-03-16       Impact factor: 1.900

6.  Use of impedance-based catheter tip-to-tissue contact assessment (electroanatomic coupling index, ECI) in typical right atrial flutter ablation.

Authors:  Massimiliano Maines; Francesco Peruzza; Alessandro Zorzi; Domenico Catanzariti; Carlo Angheben; Maurizio Del Greco
Journal:  J Interv Card Electrophysiol       Date:  2018-04-21       Impact factor: 1.900

7.  2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.

Authors:  Edmond M Cronin; Frank M Bogun; Philippe Maury; Petr Peichl; Minglong Chen; Narayanan Namboodiri; Luis Aguinaga; Luiz Roberto Leite; Sana M Al-Khatib; Elad Anter; Antonio Berruezo; David J Callans; Mina K Chung; Phillip Cuculich; Andre d'Avila; Barbara J Deal; Paolo Della Bella; Thomas Deneke; Timm-Michael Dickfeld; Claudio Hadid; Haris M Haqqani; G Neal Kay; Rakesh Latchamsetty; Francis Marchlinski; John M Miller; Akihiko Nogami; Akash R Patel; Rajeev Kumar Pathak; Luis C Sáenz Morales; Pasquale Santangeli; John L Sapp; Andrea Sarkozy; Kyoko Soejima; William G Stevenson; Usha B Tedrow; Wendy S Tzou; Niraj Varma; Katja Zeppenfeld
Journal:  Europace       Date:  2019-08-01       Impact factor: 5.214

8.  Higher contact force during radiofrequency ablation leads to a much larger increase in edema as compared to chronic lesion size.

Authors:  Samuel Thomas; Josh Silvernagel; Nathan Angel; Eugene Kholmovski; Elyar Ghafoori; Nan Hu; John Ashton; Derek J Dosdall; Rob MacLeod; Ravi Ranjan
Journal:  J Cardiovasc Electrophysiol       Date:  2018-06-06

Review 9.  Better Lesion Creation And Assessment During Catheter Ablation.

Authors:  Saurabh Kumar; Chirag R Barbhaiya; Samuel Balindger; Roy M John; Laurence M Epstein; Bruce A Koplan; Usha B Tedrow; William G Stevenson; Gregory F Michaud
Journal:  J Atr Fibrillation       Date:  2015-10-31

10.  Impact of Variableorientation and Flow Rates on Radiofrequency Ablation Lesions Created by Externally Irrigated Catheters: An Ex-Vivo Study.

Authors:  Nishaki Mehta; Katerina Morgaenko; William Sauer; William Stevenson; David Haines
Journal:  J Atr Fibrillation       Date:  2020-10-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.