| Literature DB >> 26712454 |
Simon Daniel Robinson1,2, Barbara Dymerska1,2, Wolfgang Bogner1,2, Markus Barth3, Olgica Zaric1,2, Sigrun Goluch1,4, Günther Grabner1,2, Xeni Deligianni5,6, Oliver Bieri5,6, Siegfried Trattnig1,2.
Abstract
PURPOSE: To develop a simple method for combining phase images from multichannel coils that does not require a reference coil and does not entail phase unwrapping, fitting or iterative procedures. THEORY AND METHODS: At very short echo time, the phase measured with each coil of an array approximates to the phase offset to which the image from that coil is subject. Subtracting this information from the phase of the scan of interest matches the phases from the coils, allowing them to be combined. The effectiveness of this approach is quantified in the brain, calf and breast with coils of diverse designs.Entities:
Keywords: parallel transmit; phase combination; phase imaging; phased array coils; ultra-high field
Mesh:
Year: 2015 PMID: 26712454 PMCID: PMC5217082 DOI: 10.1002/mrm.26093
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668
Figure 1Steps in the COMPOSER method, illustrated for eight channels of a 32‐channel head coil. Single‐channel phase images from the scan to be reconstructed (a) the target—a high‐resolution GRE acquisition—show disparities resulting from surface coil sensitivities as well as common variation due to local deviation from the static magnetic field (eg, in frontal regions). Phase images from a short‐echo‐time reference scan (b) reflect phase offsets; the superposition of a circularly symmetric variation from B1 + (the same in all images) and a channel‐specific B1 − are apparent, but no visible susceptibility‐related contribution. SER images are coregistered to the high‐resolution GRE scan, and the SER phase subtracted from the phase of the target scan, after which these appear identical (c). Combined phase (d) and magnitude images (e) are generated according to Eqs. (4) and (5).
Figure 2A comparison of the quality of phase images generated with COMPOSER and the other approaches tested (one subject). The left panel shows the same slice (slice 26). Magnitude images (MSS) are scaled identically and the quality parameter Q is defined in Eq. (7) (low Q values reveal poor phase matching). No Correction: The magnitude and Q values were generally low and many slices had regions of very low SNR (see slice 46 in right panel). Hammond: The magnitude and Q values were high at the center of the image but low in distal slices, as illustrated in the slice in the right panel. Adaptive Combined: No magnitude or Q maps were available, but phase images showed open‐ended fringe lines, indicating complete signal loss. After processing, these persisted as isolated point artifacts (see arrow 1). MCPC‐3D‐II: Phase matching was generally very good, with most regions having Q values close to 100%. The origin of isolated artifacts (at arrows) is described in the main text and Supporting Figure S3. With both the Roemer method and COMPOSER, the Q values were close to 100% throughout the brain and in regions of disconnected tissue such as the eyes and scalp. No artifacts were apparent in either approach, although there was a larger background phase variation in the Roemer reconstruction.
Figure 3Quantitative comparison of the quality of phase matching, assessed using the quality metric Q, defined in Eq. (7). Plots are mean values over the brains of six subjects, except the Roemer method, which is for one subject. The ordinate has been scaled logarithmically to allow comparison of the relative number of voxels with poor matching.
Figure 4A comparison of the quality of a COMPOSER phase image and a phase difference reconstruction, illustrated in a detail of an axial midbrain slice at the level of the superior colliculus (CP, cerebral peduncles; SN, substantia nigra; RN, red nucleus; MB, mammillary body). In the multi‐echo data (NE=3; top and center), the COMPOSER reconstruction of the second echo (middle) is less noisy than the phase difference between the first and second echo (top). Additional SNR gain is made with a single‐echo acquisition with a lower bandwidth (bottom, from the protocol acquired for the group of six subjects), which was only possible with the COMPOSER approach. The images have been scaled to yield similar contrast.
Figure 5Demonstration of the quality of phase matching and reconstruction with COMPOSER with a calf coil with no volume reference. The phase images from the two proton channels (“GE” column) show little similarity before phase matching, but appear identical after phase matching. The Q values are close to 100% throughout the image. The combined phase image could be spatially unwrapped.