Thierry Scheerlinck1,2, Mathias Polfliet3,4, Rudi Deklerck3,4, Gert Van Gompel2,5, Nico Buls2,5, Jef Vandemeulebroucke3,4.
Abstract
BACKGROUND ANDEntities:
Mesh:
Year: 2015 PMID: 26634843 PMCID: PMC4812075 DOI: 10.3109/17453674.2015.1123569
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Figure 1.Schematic representation of the proposed method. A. Segmentation of the bone and prosthesis images based on the source images. B. Bone registration. C. Prosthesis registration. D. Quantification of the migration and its visualization.
Figure 2.Translations and rotations are expressed based on a coordinate system with the tip of the prosthesis as origin. Positive translations along the X-axis correspond to medial migration, those along the Y-axis to proximal migration, and those along the Z-axis to anterior migration. Positive rotations about the X-axis correspond to anterior tilt, those about the Y-axis to retroversion, and those about the Z-axis to valgus.
Figure 3.To determine accuracy, a CPT hip implant (Stem) was cemented in a proximal femur (Bone) and removed from the cement mantle. The bone was fixed to a wooden board and the stem was fixed to 3 stages, allowing movement of the stem in controlled steps along the Y- and Z-axes (translation) and about the Y-axis (rotation). To assess translations along the X-axis, the bone and stem were rotated 90° about the Y-axis and the “Z-stage” was used to translate the stem in a mediolateral direction.
Overview of the 39 comparisons between the baseline CT scanner and different degrees of translations and rotations used to assess the accuracy of the CTSA tool. Values are number
| Pure translations | X-axis | Y-axis | Z-axis | |
|---|---|---|---|---|
| 0.1 mm to 0.5 mm (every 0.1 mm) | 5 | 5 | 5 | |
| 1.0 mm and 1.5 mm | 2 | 2 | 2 | |
| 2.0 mm, 3.0 mm, and 4.0 mm | 0 | 3 | 0 | |
| Pure rotations | ||||
| 0.5° to 3.0° (every 0.5°) | 0 | 6 | 0 | |
| Combined translations and rotations | ||||
| 0.5 mm/0.5° to 1.5 mm/1.5° (every 0.5 mm and 0.5°) | 0 | 3 | 0 | |
| Combined translations | X- and | Y- and | ||
| Y-axes | Z-axes | |||
| 0.5 mm/0.5 mm to 1.5 mm/1.5 mm (every 0.5 mm and 0.5 mm) | 3 | 3 | ||
The accuracy of the CTSA tool for imposed migrations in the mechanical model, assessed for 39 comparisons
| MSE | SDSE | 95% TI | p-SW | MAE | SDAE | MaxAE | |
|---|---|---|---|---|---|---|---|
| Mediolateral translation (X-axis) in mm | –0.004 | 0.053 | −0.147; 0.139 | 0.106 | 0.053 | 0.043 | 0.119 |
| Superoinferior translation (Y-axis) in mm | –0.042 | 0.068 | −0.224; 0.141 | 0.179 | 0.030 | 0.020 | 0.148 |
| Anteroposterior translation (Z-axis) in mm | –0.054 | 0.081 | −0.271; 0.163 | 0.000 | 0.040 | 0.045 | 0.190 |
| Anteroposterior tilt (around X-axis) in ° | 0.038 | 0.057 | −0.117; 0.192 | 0.596 | 0.042 | 0.032 | 0.148 |
| Retro- and anteversion tilt (around Y-axis) in ° | –0.006 | 0.036 | −0.103; 0.091 | 0.085 | 0.064 | 0.047 | 0.068 |
| Varus-valgus tilt (around Z-axis) in ° | 0.038 | 0.047 | −0.087; 0.163 | 0.230 | 0.081 | 0.052 | 0.188 |
p < 0.05, the signed error data are not likely to be normally distributed.
Figure 4.Bland-Altman plot of translations (left panel) and rotations (right panel) of the CTSA tool in the mechanical model. Horizontal lines refer to average values of the signed error for X-, Y-, and Z-translations and rotations (solid lines) and their 95% tolerance intervals (dotted lines).
The reliability coefficients and the parameters of the linear regression of the accuracy of the CTSA tool in the mechanical model, assessed for 39 comparisons
| rxx | r2 | Slope (m) | Intercept (b) | |
|---|---|---|---|---|
| Mediolateral translation (X-axis) | 0.968 | 0.984 | 0.960 | 0.004 |
| Superoinferior translation (Y-axis) | 0.998 | 0.995 | 0.996 | −0.040 |
| Anteroposterior translation (Z-axis) | 1.007 | 0.959 | 0.986 | −0.051 |