Literature DB >> 26633935

A Meta-Analysis of Bilateral Essure® Procedural Placement Success Rates on First Attempt.

Gabriel Frietze1, Ophra Leyser-Whalen2, Mahbubur Rahman3, Mahta Rouhani3, Abbey B Berenson3.   

Abstract

Background: The Essure® (Bayer HealthCare Pharmaceuticals, Leverkusen, Germany) female sterilization procedure entails using a hysteroscope to guide a microinsert into the Fallopian tube openings. Failed placement can lead to patient dissatisfaction, repeat procedures, unintended or ectopic pregnancy, perforation of internal organs, or need for subsequent medical interventions. Additional interventions increase women's health risks, and costs for patients and the health care industry. Demonstrated successful placement rates are 63%-100%. To date, there have not been any systematic analyses of variables associated with placement rates.
Objectives: The aims of this review were: (1) to estimate the average rate of successful bilateral Essure microinsert placement on first attempt; and (2) to identify variables associated with successful placement. Materials and
Methods: A meta-analysis was conducted on 64 published studies and 19 variables. Following Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines, all published studies between November 2001 and February 2015 were reviewed. The studies were taken from from PubMed and Google Scholar, and by using the the "snowball" method that reported variables associated with successful bilateral Essure placement rates.
Results: The weighted average rate of successful bilateral microinsert placement on first attempt was 92% (0.92 [95% confidence interval: 0.904-0.931]). Variables associated with successful placements were: (1) newer device models; (2) higher body mass index; and (3) a higher percent of patients who received local anesthesia. Conclusions: The data gathered for this review indicate that the highest bilateral success rates may be obtained by utilizing the newest Essure device model with local anesthesia in heavier patients. More standardized data reporting in published Essure studies is recommended. (J GYNECOL SURG 31:308).

Entities:  

Year:  2015        PMID: 26633935      PMCID: PMC4657511          DOI: 10.1089/gyn.2015.0054

Source DB:  PubMed          Journal:  J Gynecol Surg        ISSN: 1042-4067


  60 in total

1.  How meta-analysis increases statistical power.

Authors:  Lawrence D Cohn; Betsy J Becker
Journal:  Psychol Methods       Date:  2003-09

2.  Bipolar hysteroscopic procedures and placement of Essure microinserts for tubal sterilization: a case control study.

Authors:  Pierre Panel; Isabelle Grosdemouge; Marie Houllier; Frédérique Renouvel; Ludovic Friederich; Arnaud Le Tohic
Journal:  Fertil Steril       Date:  2011-04-16       Impact factor: 7.329

3.  Safety and Effectiveness of NovaSure® Endometrial Ablation After Placement of Essure® Micro-Inserts.

Authors:  Cindy M Basinski; Pamela Price; Jamie Burkhart; Jamie Johnson
Journal:  J Gynecol Surg       Date:  2012-04

4.  Tubal risk markers for failure to place transcervical sterilization coils.

Authors:  Ophra Leyser-Whalen; Mahta Rouhani; Mahbubur Rahman; Abbey B Berenson
Journal:  Contraception       Date:  2011-10-27       Impact factor: 3.375

5.  A comparison of novice and experienced physicians performing hysteroscopic sterilization: an analysis of an FDA-mandated trial.

Authors:  Mark Levie; Scott G Chudnoff
Journal:  Fertil Steril       Date:  2011-07-23       Impact factor: 7.329

6.  Initial Asian experience in hysteroscopic sterilisation using the Essure permanent birth control device.

Authors:  Bernard Chern; Anthony Siow
Journal:  BJOG       Date:  2005-09       Impact factor: 6.531

7.  Efficacy of Essure hysteroscopic sterilization--5 years follow up of 1200 women.

Authors:  José E Ríos-Castillo; Esther Velasco; José E Arjona-Berral; José A Monserrat Jordán; Balbino Povedano-Cañizares; Camil Castelo-Branco
Journal:  Gynecol Endocrinol       Date:  2013-04-05       Impact factor: 2.260

8.  Analysis of pain and satisfaction with office-based hysteroscopic sterilization.

Authors:  Mark Levie; Gil Weiss; Bente Kaiser; Jennifer Daif; Scott G Chudnoff
Journal:  Fertil Steril       Date:  2009-08-14       Impact factor: 7.329

9.  Radiofrequency global endometrial ablation followed by hysteroscopic sterilization.

Authors:  Matthew R Hopkins; Douglas J Creedon; Sherif A El-Nashar; Douglas L Brown; Andrew E Good; Abimbola O Famuyide
Journal:  J Minim Invasive Gynecol       Date:  2007 Jul-Aug       Impact factor: 4.137

10.  Predictive factors of Essure implant placement failure: prospective, multicenter study of 495 patients.

Authors:  Pierre Panel; Isabelle Grosdemouge
Journal:  Fertil Steril       Date:  2008-11-19       Impact factor: 7.329

View more
  1 in total

1.  Reversible inhibition of sperm under guidance as an intratubular and reversible contraception in female rats: An experimental study.

Authors:  Abdul Salam Ansari; Kiran Sevliya; Ayesha Badar; Nirmal Kumar Lohiya
Journal:  Int J Reprod Biomed       Date:  2021-01-25
  1 in total