Literature DB >> 23298082

Dosimetric accuracy of a deterministic radiation transport based (192)Ir brachytherapy treatment planning system. Part III. Comparison to Monte Carlo simulation in voxelized anatomical computational models.

K Zourari1, E Pantelis, A Moutsatsos, L Sakelliou, E Georgiou, P Karaiskos, P Papagiannis.   

Abstract

PURPOSE: To compare TG43-based and Acuros deterministic radiation transport-based calculations of the BrachyVision treatment planning system (TPS) with corresponding Monte Carlo (MC) simulation results in heterogeneous patient geometries, in order to validate Acuros and quantify the accuracy improvement it marks relative to TG43.
METHODS: Dosimetric comparisons in the form of isodose lines, percentage dose difference maps, and dose volume histogram results were performed for two voxelized mathematical models resembling an esophageal and a breast brachytherapy patient, as well as an actual breast brachytherapy patient model. The mathematical models were converted to digital imaging and communications in medicine (DICOM) image series for input to the TPS. The MCNP5 v.1.40 general-purpose simulation code input files for each model were prepared using information derived from the corresponding DICOM RT exports from the TPS.
RESULTS: Comparisons of MC and TG43 results in all models showed significant differences, as reported previously in the literature and expected from the inability of the TG43 based algorithm to account for heterogeneities and model specific scatter conditions. A close agreement was observed between MC and Acuros results in all models except for a limited number of points that lay in the penumbra of perfectly shaped structures in the esophageal model, or at distances very close to the catheters in all models.
CONCLUSIONS: Acuros marks a significant dosimetry improvement relative to TG43. The assessment of the clinical significance of this accuracy improvement requires further work. Mathematical patient equivalent models and models prepared from actual patient CT series are useful complementary tools in the methodology outlined in this series of works for the benchmarking of any advanced dose calculation algorithm beyond TG43.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23298082     DOI: 10.1118/1.4770275

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Current state of the art brachytherapy treatment planning dosimetry algorithms.

Authors:  P Papagiannis; E Pantelis; P Karaiskos
Journal:  Br J Radiol       Date:  2014-07-16       Impact factor: 3.039

Review 2.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17

3.  Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter - Part I: Core algorithms and validation.

Authors:  Alexander Maslowski; Adam Wang; Mingshan Sun; Todd Wareing; Ian Davis; Josh Star-Lack
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

4.  Dosimetric comparison of Acuros™ BV with AAPM TG43 dose calculation formalism in breast interstitial high-dose-rate brachytherapy with the use of metal catheters.

Authors:  Mourougan Sinnatamby; Vivekanandan Nagarajan; Sathyanarayana Reddy K; Gunaseelan Karunanidhi; Vivekanandam Singhavajala
Journal:  J Contemp Brachytherapy       Date:  2015-09-14

5.  Film based verification of calculation algorithms used for brachytherapy planning-getting ready for upcoming challenges of MBDCA.

Authors:  Grzegorz Zwierzchowski; Grzegorz Bielęda; Janusz Skowronek; Magdalena Mazur
Journal:  J Contemp Brachytherapy       Date:  2016-08-16

Review 6.  A brief look at model-based dose calculation principles, practicalities, and promise.

Authors:  Ron S Sloboda; Hali Morrison; Brie Cawston-Grant; Geetha V Menon
Journal:  J Contemp Brachytherapy       Date:  2017-02-08

7.  BrachyGuide: a brachytherapy-dedicated DICOM RT viewer and interface to Monte Carlo simulation software.

Authors:  Evaggelos Pantelis; Vassiliki Peppa; Vasileios Lahanas; Eleftherios Pappas; Panagiotis Papagiannis
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

8.  Impact of GBBS algorithm on post-mastectomy scar boost irradiation of breast using catheter flap.

Authors:  Mourougan Sinnatamby; Vijayaprabhu Neelakandan; Gunaseelan Karunanidhi; Saravanan Kandasamy; Seenisamy Ramapandian; Muniyappan Kannan; Elakiya Sampath
Journal:  J Contemp Brachytherapy       Date:  2021-05-24

9.  Head and neck (192)Ir HDR-brachytherapy dosimetry using a grid-based Boltzmann solver.

Authors:  Frank-André Siebert; Sabine Wolf; George Kóvacs
Journal:  J Contemp Brachytherapy       Date:  2013-12-18

10.  A comparative assessment of inhomogeneity and finite patient dimension effects in 60Co and 192Ir high-dose-rate brachytherapy.

Authors:  Irina Fotina; Kyveli Zourari; Vasileios Lahanas; Evaggelos Pantelis; Panagiotis Papagiannis
Journal:  J Contemp Brachytherapy       Date:  2018-02-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.