Literature DB >> 18490526

Necessity for surgical revision of defibrillator leads implanted long-term: causes and management.

Jens Eckstein1, Michael T Koller, Markus Zabel, Dietrich Kalusche, Beat A Schaer, Stefan Osswald, Christian Sticherling.   

Abstract

BACKGROUND: Defibrillator lead malfunction is a potential long-term complication in patients with an implantable cardioverter-defibrillator (ICD). The aim of this study was to determine the incidence and causes of lead malfunction necessitating surgical revision and to evaluate 2 approaches to treat lead malfunction. METHODS AND
RESULTS: We included 1317 consecutive patients with an ICD implanted at 3 European centers between 1993 and 2004. The types and causes of lead malfunction were recorded. If the integrity of the high-voltage part of the lead could be ascertained, an additional pace/sense lead was implanted. Otherwise, the patients received a new ICD lead. Of the 1317 patients, 38 experienced lead malfunction requiring surgical revision and 315 died during a median follow-up of 6.4 years. At 5 years, the cumulative incidence was 2.5% (95% confidence interval, 1.5 to 3.6). Lead malfunction resulted in inappropriate ICD therapies in 76% of the cases. Implantation of a pace/sense lead was feasible in 63%. Both lead revision strategies were similar with regard to lead malfunction recurrence (P=0.8). However, the cumulative incidence of recurrence was high (20% at 5 years; 95% confidence interval, 1.7 to 37.7).
CONCLUSIONS: ICD lead malfunction necessitating surgical revision becomes a clinically relevant problem in 2.5% of ICD recipients within 5 years. In selected cases, simple implantation of an additional pace/sense lead is feasible. Regardless of the chosen approach, the incidence of recurrent ICD lead-related problems after lead revision is 8-fold higher in this population.

Entities:  

Mesh:

Year:  2008        PMID: 18490526     DOI: 10.1161/CIRCULATIONAHA.107.740670

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  24 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

2.  The Sprint Fidelis lead fracture story: time to come to our senses?

Authors:  P F H M van Dessel
Journal:  Neth Heart J       Date:  2010-01       Impact factor: 2.380

3.  The Entirely Subcutaneous Defibrillator - A New Generation and Future Expectations.

Authors:  Hussam Ali; Pierpaolo Lupo; Riccardo Cappato
Journal:  Arrhythm Electrophysiol Rev       Date:  2015-08

4.  [Lead survival and complications (except infections). Are we doing better nowadays?].

Authors:  Martin Seifert; Michael Neuss; Maren Schöpp; Cornel Koban; Christian Butter
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-08-06

5.  [Totally subcutaneous cardioverter-defibrillator (S-ICD®) : recent experience and future perspectives].

Authors:  J Köbe; S Zumhagen; F Reinke; E Schulze-Bahr; L Eckardt
Journal:  Herz       Date:  2011-10       Impact factor: 1.443

6.  The mismatch between patient life expectancy and the service life of implantable devices in current cardioverter-defibrillator therapy: a call for larger device batteries.

Authors:  Jörg Neuzner
Journal:  Clin Res Cardiol       Date:  2015-02-19       Impact factor: 5.460

Review 7.  Prophylactic implantable cardioverter defibrillator in heart failure: the growing evidence for all or Primum non nocere for some?

Authors:  Khang-Li Looi; Nigel Lever; Anthony Tang; Sharad Agarwal
Journal:  Heart Fail Rev       Date:  2017-05       Impact factor: 4.214

Review 8.  [The subcutaneous cardioverter-defibrillator: When less is more].

Authors:  J Kuschyk; B Rudic; I Akin; M Borggrefe; S Röger
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-06-10

Review 9.  Subcutaneous implantable defibrillator: State-of-the art 2013.

Authors:  Finn Akerström; Miguel A Arias; Marta Pachón; Alberto Puchol; Jesús Jiménez-López
Journal:  World J Cardiol       Date:  2013-09-26

Review 10.  Heart Failure With Reduced Ejection Fraction And A Narrow QRS Complex: Combination Of A Subcutaneous Defibrillator With Cardiac Contractility Modulation.

Authors:  Susanne Röger; Martin Borggrefe; Jürgen Kuschyk
Journal:  J Atr Fibrillation       Date:  2015-10-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.