Extramedullary versus intramedullary tibial alignment technique in total knee arthroplasty: A meta-analysis of randomized controlled trials.

Huan Bei Zeng1, Xiao Zhou Ying1, Guang Jun Chen1, Xia Qing Yang1, Duo Duo Lin1, Zhi Jie Li1, Hai Xiao Liu1.   

Abstract

The aim of this study was to establish whether the use of an extramedullary or intramedullary tibial cutting guide leads to superior mechanical leg axis and implant positioning. A meta-analysis of six randomized controlled trials including 350 knees was performed. For the mechanical axis, frontal tibial component angle and tibial slope, there were no significant differences in the mean values or the number of outliers (±3°) between the extramedullary and intramedullary groups. A reduced tourniquet time was associated with the intramedullary guide. No significant difference in the complication rate was noted between the two groups. Neither extramedullary nor intramedullary tibial alignment was more accurate in facilitating the tibial cut. Use of an intramedullary guide results in a shorter tourniquet time and exhibits a similar complication rate as the extramedullary guide.

Entities:  

Mesh:

Year:  2015        PMID: 26598086      PMCID: PMC4602384          DOI: 10.6061/clinics/2015(10)10

Source DB:  PubMed          Journal:  Clinics (Sao Paulo)        ISSN: 1807-5932            Impact factor:   2.365


INTRODUCTION

Total knee arthroplasty (TKA) is a successful procedure for the treatment of pain and for restoring physical function in patients with severe arthritis 1-4. Lower extremity alignment is one of the paramount factors determining the long-term success of TKA. The survival rate of TKA is increased if leg alignment is restored within 3° of valgus or varus on the mechanical axis 5,6. Malpositioning of the implant can lead to early wear and loosening as well as inferior functional performance 7,8, which potentially exposes patients to reduced implant longevity 9-11. Both intramedullary (IM) and extramedullary (EM) techniques are popular for guiding tibial component alignment in TKA. However, significant debate still exists regarding the optimal alignment guide for the placement of the tibial component. Most TKA systems offer both methods at the choice of the surgeon. Each tibial instrumentation method is reliable, although different authors have presented opposing results with respect to which type of tibial instrumentation results in better component alignment 12-14. Some randomized controlled trials (RCTs) have been published comparing the IM with the EM guide for tibial component alignment in TKA. However, a meta-analysis evaluating the radiographical outcomes between the two guiding techniques has not been performed. In this study, we conducted a meta-analysis of pooled data from relevant RCTs to evaluate whether an IM or EM tibial guide is more accurate in assuring correct tibial positioning. Moreover, the tourniquet time and complication rate were also compared between these two techniques.

MATERIALS AND METHODS

We conducted a meta-analysis of all English and non-English articles identified from electronic databases including Medline, Embase, Cochrane Library, China National Knowledge Infrastructure, Wan Fang Chinese Periodical and Google. In addition, we also manually searched for other relevant studies including those from the reference lists of all included studies. The last search was conducted on September 2, 2014. We used the following key words: arthroplasty, replacement, knee, total knee arthroplasty, randomized, randomised, intramedullary and extramedullary. These key words were used in combination with the Boolean operators AND or OR. The search strategy is presented in Figure 1.
Figure 1

Flow chart of the study selection and inclusion process.

Exclusion criteria and quality criteria

We included all published RCTs comparing EM guides with IM guides in patients undergoing primary TKA. Exclusion criteria comprised the following: trials with a retrospective design and trials that did not randomize patients into two relevant groups. Quality criteria included the randomization method, concealment of allocation, blinding and intention-to-treat analysis.

Data extraction

For each eligible study, two of the authors of this meta-analysis independently extracted all relevant data. Disagreement was resolved by discussion with a third investigator. The following data were extracted: 1) the participants' demographic data; 2) the mean value of the mechanical axis and the number of outliers (±3°); 3) the mean value of the frontal tibial component (FTC) angle and the number of outliers (±3°); 4) the mean value of the tibial slope; 5) the tourniquet time; 6) and the complication rate. When data were incomplete or unclear, attempts were made to contact the investigators for clarification.

Radiologic limb alignment

The mechanical axis was defined as a line bisecting the center of the femoral head, the center of the knee and the center of the ankle. The FTC angle was defined as the angle measured between the articular surface of the tibial component and the mechanical axis of the tibia with a goal of 90°. The tibial slope was measured as the angle between the tibial plateau and a line perpendicular to the anatomical axis of the tibia. Up to 3° of deviation from neutral alignment was considered acceptable, whereas values outside of this range were classified as outliers.

Data analysis

This meta-analysis was conducted using RevMan 5.0 (Cochrane Collaboration, Oxford, UK). We assessed the statistical heterogeneity using a standard chi-square test (statistical heterogeneity was considered to be present at p<0.1 and I2 values >50%). When comparing trials exhibiting heterogeneity, pooled data were meta-analyzed using a random effects model; otherwise, a fixed effects model was used. Mean differences and 95% confidence intervals (CIs) were calculated for continuous outcomes and risk ratios (RR) and 95% CIs were calculated for dichotomous outcomes.

RESULTS

A total of 363 potentially relevant papers were identified. By screening titles and reading the abstracts and the entire articles, six studies with 350 knees (173 in the EM group and 177 in the IM group) were included in the final meta-analysis. All of these RCTs were published in English. The sample sizes ranged from 50 to 100 knees. Five studies assessed radiological alignment 13-18 and one study assessed functional outcomes 19. The key characteristics of the included RCTs are summarized in Table 1.
Table 1

Characteristics of the included studies.

AuthorCountryPatients(EM)/(IM)Age(EM)/(IM)Gender(EM)/(IM)Total knee systemEvaluation of alignment
Blakeney 2011 (17)    2014 (19)Australia34/36    35/36<70 years:11/19;  >70 years: 23(24)/17Male:12/15    Female: 23/21Genesis II total knee system  (Smith&Nephew)6 weeks for X ray;  3 months for CT
Kroon 2012 (18)Netherlands24/26Not statedNot statedGenesis II MIS-TKA  (Smith&Nephew)4 to 6 monthsX ray
Lozano 2008 (16)Spain38/3170 years old/ 69 years oldMale:7/3    Female: 32/28Profix total knee system    (Smith&Nephew)Examination time: Not stated X ray
Chin 2005 (15)Singapore30/3065.6 years old/    66.9 years oldMale:7/6    Female: 23/24PFC Sigma total knee system     (Depuy)Examination time: Not stated X ray
Reed 2002 (14)United Kingdom46/5468 years old/  69 years oldMale: 24/22    Female: 28/26Not stated3 months X ray

EM: Extramedullary group; IM: Intramedullary group.

Methodological quality

The methodological quality of the six included studies was variable. The reported methods of generating allocation sequences were adequate in two studies and five trials reported allocation concealment. Blinding of the surgeon and the patients was reported in four studies and five of the studies blinded their assessors to the outcome. The methodological quality of the studies is presented in Figure 2. Judgment with respect to each risk of bias item is presented as a percentage for all of the included studies, as shown in Figure 3.
Figure 2

Methodological quality of the included studies. This risk of bias tool incorporates assessment of randomization (sequence generation and allocation concealment), blinding (participants, personnel and outcome assessors), completeness of outcome data, selection of outcomes reported and other sources of bias. The items were scored with “yes”, “no”, or “unsure”.

Figure 3

Risk of bias. Each risk of bias item is presented as a percentage across all included studies and indicates the proportional level for each risk of bias item.

Comparison of radiologic limb alignment

The pooled results indicated that there was no significant difference between the two groups in terms of the mean mechanical axis (p=0.31, Figure 4a). Outliers in the mechanical axis occurred in 37.3% of knees (31/83) in the EM group compared with 33.7% (30/89) in the IM group and there was no significant difference (p=0.63, Figure 4b). No significant difference in the mean FTC angle was noted between the two groups (p=0.60, Figure 4c). The outliers in the FTC angle occurred in 26.3% of knees (20/76) in the EM group compared with 25% (21/84) in the IM group and there was no significant difference (p=0.89, Figure 4d). There was no significant difference between the two groups in terms of the mean tibial slope (p=0.42, Figure 4e).
Figure 4

Comparison of the mean mechanical axis between the extramedullary and the intramedullary technique. b. Relative risk of producing a deviation of greater than 3o from neutral in the mechanical axis in extramedullary group vs. the intramedullary group. c. Comparison of the mean FTC angle between the extramedullary and intramedullary technique. d. Relative risk of producing deviation of greater than 3o from neutral in the FTC angle for the extramedullary vs. intramedullary techniques. e. Comparison of the mean tibial slope between the extramedullary and intramedullary groups.

Comparison of the tourniquet time and the complication rate

The tourniquet time was shorter in the IM group compared with the EM group (p=0.01, Figure 5a). TKA-related complications occurred in 4.2% of knees (3/72) in the EM group compared with 3.0% (2/67) in the IM group with no significant difference (p=0.70, Figure 5b). Blakeney et al. 17 stated that there was one case of pulmonary embolism, one deep infection and one case of knee stiffness in the EM group and one case of knee stiffness in the IM group. Chin et al. 15 reported that one patient in the IM group had a mild stroke. The functional scores were evaluated, although insufficient data were available for the meta-analysis. Only one study measured functional knee scores (Oxford knee score) with mean scores of 37.6 in the EM group and 36.8 in the IM group.
Figure 5

Comparison of the mean tourniquet time between the extramedullary group and the intramedullary group. b. Comparison of the complication rate between the extramedullary group and the intramedullary group.

DISCUSSION

Our meta-analysis compared the radiographic outcomes between the EM and the IM guiding techniques in patients undergoing TKA. No significant differences were found between the two groups in terms of the mean values of the mechanical axis, the FTC angle or the tibial slope. Moreover, neither the EM nor the IM guiding techniques offer an advantage over the other method in reducing outliers of greater than 3°. The IM guide is associated with a shorter tourniquet time and exhibits a similar complication rate as the EM guide. A comparison of functional outcomes between the two groups could not be performed in this study because RCT research on functional outcomes is scarce. A previous study of British orthopedic surgeons found that 75.6% prefer EM and 20.3% prefer IM jigs when determining tibial alignment with the remainder using both or neither 20. The published literature is divided as to which jig is superior. According to the results of our literature review, approximately 52.6% studies argue that IM and EM guides are equally accurate for tibial alignment 12,16-18,. Approximately 36.8% suggest that an IM guide is more accurate (13,14,27-30,34) and 10.5% suggest that an extramedullary guide is more accurate 31,32. However, very few studies included large samples or RCTs for a comparison of the two methods. Regarding the accuracy of the tibial cutting, this current meta-analysis study suggests that neither EM nor IM tibial alignment is more accurate than the other approach. Most surgeons prefer to use the extramedullary guide either because they are more experienced in its use or because of the possible complications of the IM guide. However, because the center of the talus is slightly medial to the midpoint between the malleoli, the surgeon must estimate the location of the center of the talus based on these bony landmarks, which may be obscured by soft tissue in obese patients or by bony abnormalities 33. For IM guides, the entry point position is a key factor and the ideal entry point position is located on the tibial articular surface corresponding to the proximal continuation of the tibial canal and should be preoperatively determined with the help of X-rays 34. Moreover, decompression of the medullary contents using suction before instrumenting the canal is also recommended to decrease the risk of embolizing the medullary contents 35. Therefore, it is important for the surgeon to appreciate the benefits and deficiencies of each guide and to use whichever method is most appropriate for each particular case, although both the EM and IM systems allow for satisfactory alignment. Most surgeons have accepted that a postoperative mechanical axis of 0°±3° will result in less pain, better knee function, faster rehabilitation and improved quality of life 36-39. Recently, several studies found that a postoperative mechanical axis of 0°±3° did not result in better long-term survival of TKA implants compared with a group of outliers 39-43. In one of the most influential studies, Parratte et al. 40 retrospectively reviewed the data of 398 primary TKAs and found that a mechanical axis of 0°±3° did not improve the rate of survival 15 years postoperatively. This result implies that the accuracy of the mechanical axis likely provides limited value with regards to long-term durability. In addition, although computer-assisted TKA improves the mechanical leg axis and component orientation compared with the conventional technique, there is currently no proven clinical benefit of this approach. Therefore, future research on tibial guiding techniques should not only assess radiological alignment but also consider functional outcomes. This present meta-analysis has several limitations. First, only six studies were included and the sample size of the included studies was small, which might have affected our results. Second, most of the trials focused on short-term radiographic outcomes and only one RCT study evaluated functional outcome. Therefore, we could not perform a valid statistical comparison of the functional outcomes between the two groups. Therefore, further high-quality RCTs with long-term follow-up should be designed to assess radiographic outcomes, knee function and implant survival rate.
  42 in total

1.  Coronal alignment after total knee replacement.

Authors:  R S Jeffery; R W Morris; R A Denham
Journal:  J Bone Joint Surg Br       Date:  1991-09

2.  A comparison of intramedullary and extramedullary alignment systems for tibial component placement in total knee arthroplasty.

Authors:  D A Brys; A V Lombardi; T H Mallory; B K Vaughn
Journal:  Clin Orthop Relat Res       Date:  1991-02       Impact factor: 4.176

3.  Extramedullary versus intramedullary alignment guides in total knee arthroplasty.

Authors:  Y Ishii; G Ohmori; J E Bechtold; R B Gustilo
Journal:  Clin Orthop Relat Res       Date:  1995-09       Impact factor: 4.176

4.  A comparative study of extramedullary and intramedullary alignment systems in total knee arthroplasty.

Authors:  E D Tillett; G A Engh; T Petersen
Journal:  Clin Orthop Relat Res       Date:  1988-05       Impact factor: 4.176

5.  Comparative experience with intramedullary and extramedullary alignment in total knee arthroplasty.

Authors:  G A Engh; T L Petersen
Journal:  J Arthroplasty       Date:  1990-03       Impact factor: 4.757

6.  Accuracy of intramedullary versus extramedullary tibial alignment cutting systems in total knee arthroplasty.

Authors:  K E Teter; D Bregman; C W Colwell
Journal:  Clin Orthop Relat Res       Date:  1995-12       Impact factor: 4.176

7.  Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty.

Authors:  D A Dennis; M Channer; M H Susman; E A Stringer
Journal:  J Arthroplasty       Date:  1993-02       Impact factor: 4.757

8.  Computer-assisted techniques versus conventional guides for component alignment in total knee arthroplasty: a randomized controlled trial.

Authors:  William G Blakeney; Riaz J K Khan; Simon J Wall
Journal:  J Bone Joint Surg Am       Date:  2011-08-03       Impact factor: 5.284

9.  Insall Award paper. Why are total knee arthroplasties failing today?

Authors:  Peter F Sharkey; William J Hozack; Richard H Rothman; Shani Shastri; Sidney M Jacoby
Journal:  Clin Orthop Relat Res       Date:  2002-11       Impact factor: 4.176

10.  Intramedullary instrumentation: safer and more accurate than extramedullary instrumentation.

Authors:  R S Laskin
Journal:  Orthopedics       Date:  2001-08       Impact factor: 1.390

View more
  4 in total

1.  Tibial alignment technique and its influence on clinical and functional outcomes following total knee arthroplasty.

Authors:  Mohit R Bansal; Diane Back; Peter Earnshaw; N A Sandiford
Journal:  J Clin Orthop Trauma       Date:  2020-04-28

2.  Accuracy of tibial positioning in the frontal plane: a prospective study comparing conventional and innovative techniques in total knee arthroplasty.

Authors:  R K Zahn; F Graef; J L Conrad; L Renner; C Perka; H Hommel
Journal:  Arch Orthop Trauma Surg       Date:  2020-03-02       Impact factor: 3.067

3.  The EOS 3D imaging system reliably measures posterior tibial slope.

Authors:  Andreas Hecker; Till D Lerch; Rainer J Egli; Emanuel F Liechti; Frank M Klenke
Journal:  J Orthop Surg Res       Date:  2021-06-16       Impact factor: 2.359

4.  A study on dynamic monitoring, components, and risk factors of embolism during total knee arthroplasty.

Authors:  Kang Lu; Mingtao Xu; Wei Li; Kai Wang; Dawei Wang
Journal:  Medicine (Baltimore)       Date:  2017-12       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.