Literature DB >> 26578490

Accurate ultrasound imaging based on range point migration method for the depiction of fetal surface.

Hirofumi Taki1, Shinya Tanimura2, Takuya Sakamoto3, Tsuyoshi Shiina3, Toru Sato3.   

Abstract

PURPOSE: The purpose of this study is to evaluate the performance of a modified range point migration (RPM) method using a semi-broad transmit beam for fetal surface imaging.
METHODS: The conventional RPM method depicts accurate images of target surfaces by estimating the reflection point on a target surface from the path length of plural transmit-and-receive element combinations. However, the conventional RPM method depicts false images when echoes from different targets are received simultaneously. For the elimination of false images in the employment of the RPM method, we propose a modified RPM method with a semi-broad transmit beam to decrease the number of targets in each measurement region.
RESULTS: The modified RPM method depicted two acrylic cylinders of 2 cm in diameter with a root-mean-square error (RMSE) of 0.062 mm, where the RMSE of the migration method was 0.145 mm. The modified RPM method also succeeded in depicting a 7-month fetal phantom with a RMSE of 0.058 mm relative to a 3D image acquired using optical measurement.
CONCLUSION: This study shows the potential of the modified RPM method in achieving accurate surface imaging of multiple targets using a semi-broad beam, indicating that the method is suitable for fetal surface imaging.

Keywords:  Fetal imaging; Fetal surface; Migration; Range point migration; Three-dimensional ultrasonography

Mesh:

Year:  2014        PMID: 26578490     DOI: 10.1007/s10396-014-0574-4

Source DB:  PubMed          Journal:  J Med Ultrason (2001)        ISSN: 1346-4523            Impact factor:   1.314


  18 in total

Review 1.  Three-dimensional ultrasound imaging.

Authors:  A Fenster; D B Downey; H N Cardinal
Journal:  Phys Med Biol       Date:  2001-05       Impact factor: 3.609

2.  High range resolution ultrasonographic vascular imaging using frequency domain interferometry with the Capon method.

Authors:  Hirofumi Taki; Kousuke Taki; Takuya Sakamoto; Makoto Yamakawa; Tsuyoshi Shiina; Motoi Kudo; Toru Sato
Journal:  IEEE Trans Med Imaging       Date:  2011-10-06       Impact factor: 10.048

3.  Three-dimensional synthetic aperture focusing using a rocking convex array transducer.

Authors:  Henrik Andresen; Svetoslav Ivanov Nikolov; Mads Møller Pedersen; Daniel Buckton; Jørgen Arendt Jensen
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2010-05       Impact factor: 2.725

4.  Comparative evaluation of despeckle filtering in ultrasound imaging of the carotid artery.

Authors:  Christos P Loizou; Constantinos S Pattichis; Christodoulos I Christodoulou; Robert S H Istepanian; Marios Pantziaris; Andrew Nicolaides
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2005-10       Impact factor: 2.725

5.  Volume rendering of 3D medical ultrasound data using direct feature mapping.

Authors:  E Steen; B Olstad
Journal:  IEEE Trans Med Imaging       Date:  1994       Impact factor: 10.048

6.  High frame-rate blood vector velocity imaging using plane waves: simulations and preliminary experiments.

Authors:  Jesper Udesen; Fredrik Gran; Kristoffer Lindskov Hansen; Jørgen Arendt Jensen; Carsten Thomsen; Michael Bachmann Nielsen
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2008-08       Impact factor: 2.725

7.  High-frame-rate echocardiography with reduced sidelobe level.

Authors:  Hideyuki Hasegawa; Hiroshi Kanai
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2012-11       Impact factor: 2.725

8.  High-frame-rate echocardiography using diverging transmit beams and parallel receive beamforming.

Authors:  Hideyuki Hasegawa; Hiroshi Kanai
Journal:  J Med Ultrason (2001)       Date:  2011-05-07       Impact factor: 1.314

9.  Small calcification depiction in ultrasound B-mode images using decorrelation of echoes caused by forward scattered waves.

Authors:  Hirofumi Taki; Takuya Sakamoto; Makoto Yamakawa; Tsuyoshi Shiina; Kenichi Nagae; Toru Sato
Journal:  J Med Ultrason (2001)       Date:  2011-02-25       Impact factor: 1.314

10.  Small calcification indicator in ultrasonography using correlation of echoes with a modified Wiener filter.

Authors:  Hirofumi Taki; Takuya Sakamoto; Makoto Yamakawa; Tsuyoshi Shiina; Toru Sato
Journal:  J Med Ultrason (2001)       Date:  2012-03-09       Impact factor: 1.314

View more
  2 in total

1.  Discrimination of thoracic spine from muscle based on their difference in ultrasound reflection and scattering characteristics.

Authors:  Tomohiro Yokoyama; Shohei Mori; Mototaka Arakawa; Eiko Onishi; Masanori Yamauchi; Hiroshi Kanai
Journal:  J Med Ultrason (2001)       Date:  2019-08-21       Impact factor: 1.314

2.  Synthetic aperture ultrasound imaging with a ring transducer array: preliminary ex vivo results.

Authors:  Xiaolei Qu; Takashi Azuma; Takeshi Yogi; Shiho Azuma; Hideki Takeuchi; Satoshi Tamano; Shu Takagi
Journal:  J Med Ultrason (2001)       Date:  2016-06-14       Impact factor: 1.314

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.