| Literature DB >> 26576110 |
Hyungjin Kim1, Jeong Min Lee2, Jeong Hee Yoon3, Jin-Young Jang4, Sun-Whe Kim4, Ji Kon Ryu5, Stephan Kannengiesser6, Joon Koo Han2, Byung Ihn Choi2.
Abstract
OBJECTIVE: To investigate the image quality (IQ) and apparent diffusion coefficient (ADC) of reduced field-of-view (FOV) di-ffusion-weighted imaging (DWI) of pancreas in comparison with full FOV DWI.Entities:
Keywords: Apparent diffusion coefficient; Diffusion-weighted imaging; Image quality; Magnetic resonance imaging; Pancreas; Reduced field-of-view
Mesh:
Year: 2015 PMID: 26576110 PMCID: PMC4644742 DOI: 10.3348/kjr.2015.16.6.1216
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
Final Diagnoses of 102 Pancreatic Lesions
| Diagnosis | No. of Patients | Lesion Location | Size (mm)* | |||
|---|---|---|---|---|---|---|
| Uncinate Process | Head | Body | Tail | |||
| Adenocarcinoma | 44 (43.1) | 5 (11.4) | 19 (43.2) | 10 (22.7) | 10 (22.7) | 27.6 ± 11.5 |
| Noninvasive IPMN | 16 (15.7) | 2 (12.5) | 6 (37.5) | 5 (31.3) | 3 (18.8) | 25.1 ± 7.3 |
| IPMC | 5 (4.9) | 2 (40) | 1 (20) | 1 (20) | 1 (20) | 31.2 ± 22.2 |
| Neuroendocrine tumor | 18 (17.6) | 1 (5.6) | 7 (38.9) | 2 (11.1) | 8 (44.4) | 21.7 ± 13.7 |
| Serous cystic tumor | 8 (7.8) | 0 (0) | 2 (25) | 3 (37.5) | 3 (37.5) | 39.4 ± 26.6 |
| Mucinous cystic tumor | 3 (2.9) | 0 (0) | 0 (0) | 0 (0) | 3 (100) | 42.2 ± 21.4 |
| Others† | 8 (7.8) | 0 (0) | 3 (37.5) | 1 (12.5) | 4 (50) | 24.7 ± 10.5 |
Data are number of patients unless otherwise specified. Data in parentheses are percentages. *Data are mean ± standard deviation, †Benign epithelial cysts (n = 3), metastases (n = 2), schwannoma (n = 1), solid pseudopapillary tumor (n = 1), and desmoid tumor (n = 1) were included. IPMC = intraductal papillary mucinous carcinoma, IPMN = intraductal papillary mucinous neoplasm
Fig. 158-year-old woman with 1.5 cm sized neuroendocrine tumor (arrow) in pancreas uncinate process.
A. Mass shows rim enhancement on enhanced T1-weighted image. B. Full field-of-view (FOV) diffusion-weighted imaging (DWI) sequence at b = 0 s/mm2. Lesion shows ill-defined hyperintensity. C. Full FOV DWI at b = 500 s/mm2. Note that margin of lesion is severely blurred and barely delineable from background. D. Corresponding apparent diffusion coefficient (ADC) map of full FOV DWI. E. Reduced FOV DWI at b = 0 s/mm2. F. Reduced FOV DWI at b = 400 s/mm2. Lesion is more clearly visualized on reduced FOV image. G. Corresponding ADC map of reduced FOV DWI.
MR Imaging Parameters
| Reduced FOV | Full FOV | |
|---|---|---|
| TR/TE (ms) | 3200/70 | 5000/70 |
| FOV (cm) | 28 × 8.5 | 38 × 38 |
| Matrix size | 192 × 35 | 136 × 136 |
| In-plane resolution (mm) | 1.5 × 2.4 | 2.8 × 2.8 |
| Section thickness (mm) | 5 | 8 |
| Slice number | 15 | 22 |
| Intersection gap (%) | 10 | 0 |
| Flip angle (degree) | 90 | 90 |
| NEX | 6 | 3 |
| B-value (s/mm2) | 0, 400, 800 | 0, 25, 50, 75, 100, 150, 200, 500, 800, 1000 or 0, 25, 50, 75, 100, 150, 200, 500, 1500 |
| Acquisition time | 2 min 18 sec | 6 min 30 sec or 7 min 15 sec |
FOV = field-of-view, NEX = number of excitations, TE = echo time, TR = repetition time
Fig. 238-year-old man with neuroendocrine tumor (not shown) in pancreas tail.
Images are cropped to focus on pancreas. A. Full field-of-view (FOV) diffusion-weighted imaging (DWI) sequence at b = 0 s/mm2. Margin of pancreas is blurred and duct is not visible. Normal lobulated appearance of pancreas is hardly recognizable. B. Full FOV DWI at b = 500 s/mm2. C. Corresponding apparent diffusion coefficient (ADC) map of full FOV DWI. D. Reduced FOV DWI at b = 0 s/mm2. Note that lobulated appearance of pancreas border and pancreatic duct are more clearly visualized on reduced FOV image. E. Reduced FOV DWI at b = 400 s/mm2. F. Corresponding ADC map of reduced FOV DWI.
Qualitative Image Analysis Scores Compared between Reduced FOV and Full FOV Diffusion-Weighted Imaging Sequences
| Anatomic Structures | Lesion Conspicuity | Artifacts | Total Image Quality Score | Subjective Clinical Utility† | |
|---|---|---|---|---|---|
| Reduced FOV (b = 0 s/mm2) | |||||
| Reader 1 | 2.95 ± 0.83 | 3.19 ± 0.99 | 2.79 ± 0.78 | 8.93 ± 2.20 | 3.40 ± 0.68 |
| Reader 2 | 2.57 ± 0.91 | 2.99 ± 1.14 | 2.50 ± 0.71 | 8.06 ± 2.16 | 3.44 ± 0.75 |
| Average | 2.76 ± 0.79 | 3.11 ± 0.99 | 2.65 ± 0.68 | 8.51 ± 2.05 | 3.41 ± 0.64 |
| Full FOV (b = 0 s/mm2) | |||||
| Reader 1 | 2.20 ± 0.63 | 2.55 ± 1.10 | 2.41 ± 0.74 | 7.16 ± 2.03 | 2.87 ± 0.86 |
| Reader 2 | 2.16 ± 0.69 | 2.55 ± 1.21 | 2.40 ± 0.69 | 7.11 ± 1.87 | 3.40 ± 0.77 |
| Average | 2.18 ± 0.59 | 2.55 ± 1.07 | 2.41 ± 0.63 | 7.13 ± 1.83 | 3.14 ± 0.70 |
| < 0.001 | < 0.001 | < 0.001 | < 0.001 | < 0.001 | |
| Reduced FOV (b = 400 s/mm2) | |||||
| Reader 1 | 2.89 ± 0.69 | 3.09 ± 0.95 | 2.95 ± 0.65 | 8.93 ± 1.74 | N/A |
| Reader 2 | 2.74 ± 0.69 | 3.21 ± 0.89 | 2.71 ± 0.67 | 8.65 ± 1.78 | |
| Average | 2.81 ± 0.64 | 3.15 ± 0.79 | 2.83 ± 0.57 | 8.79 ± 1.60 | |
| Full FOV (b = 500 s/mm2) | |||||
| Reader 1 | 2.47 ± 0.50 | 2.70 ± 0.90 | 2.62 ± 0.56 | 7.78 ± 1.45 | |
| Reader 2 | 2.65 ± 0.62 | 3.09 ± 1.00 | 2.82 ± 0.53 | 8.56 ± 1.47 | |
| Average | 2.56 ± 0.47 | 2.89 ± 0.86 | 2.72 ± 0.45 | 8.17 ± 1.31 | |
| < 0.001 | 0.002 | 0.061 | < 0.001 |
Data are mean ± standard deviation. *Wilcoxon signed-rank test was performed between reduced FOV and full FOV sequences using averaged image quality scores of two readers, †Scoring of subjective clinical utility was performed once for each DWI sequence. DWI = diffusion-weighted imaging, FOV = field-of-view, N/A = not available
Weighted κ Statistics for Inter-Reader Agreement of Qualitative Image Quality Scores
| Reduced FOV | Full FOV | |||
|---|---|---|---|---|
| b = 0 s/mm2 | b = 400 s/mm2 | b = 0 s/mm2 | b = 500 s/mm2 | |
| Anatomic structures | 0.473 (0.352, 0.594) | 0.613 (0.493, 0.732) | 0.561 (0.418, 0.703) | 0.365 (0.219, 0.512) |
| Lesion conspicuity | 0.660 (0.554, 0.767) | 0.355 (0.227, 0.483) | 0.618 (0.511, 0.726) | 0.461 (0.346, 0.576) |
| Artifacts | 0.496 (0.376, 0.616) | 0.388 (0.241, 0.534) | 0.431 (0.284, 0.578) | 0.251 (0.096, 0.405) |
| Total image quality score | 0.543 (0.443, 0.644) | 0.490 (0.386, 0.593) | 0.564 (0.474, 0.654) | 0.382 (0.278, 0.485) |
| Subjective clinical utility | 0.470 (0.312, 0.627) | 0.305 (0.165, 0.445) | ||
Data in parentheses are 95% confidence intervals. FOV = field-of-view
Apparent Diffusion Coefficient Measurements and Comparisons between Reduced Field-of-View (FOV) and Full FOV Diffusion-Weighted Imaging Sequences
| Reduced FOV (× 10-3 mm2/s) | Full FOV (× 10-3 mm2/s) | ||
|---|---|---|---|
| Adenocarcinoma (n = 26) | 1.061 ± 0.133 | 1.079 ± 0.135 | 0.438 |
| Noninvasive IPMN (n = 7) | 2.620 ± 0.252 | 2.679 ± 0.337 | 0.176 |
| IPMC (n = 3) | 1.160 ± 0.046 | 1.208 ± 0.088 | 0.285 |
| Neuroendocrine tumor (n = 7) | 0.983 ± 0.152 | 1.004 ± 0.153 | 0.499 |
| Serous cystic tumor (n = 6) | 2.607 ± 0.277 | 2.582 ± 0.289 | 0.463 |
| Mucinous cystic tumor (n = 3) | 2.709 ± 0.455 | 2.736 ± 0.346 | 0.593 |
| Pancreas parenchyma (n = 27) | 1.191 ± 0.125 | 1.218 ± 0.103 | 0.070 |
Data are mean ± standard deviation. *Wilcoxon signed-rank test was performed. IPMC = intraductal papillary mucinous carcinoma, IPMN = intraductal papillary mucinous neoplasm