Reliability of point-of-care capillary blood glucose measurements in the critical value range.

Ron B Schifman1, Tan T Nguyen, Susan T Page.   

Abstract

CONTEXT: Point-of-care glucose (POCG) testing on capillary blood specimens is central to maintaining glycemic control in patients with diabetes. Although there are known performance issues with POCG methods, especially for maintaining tight glucose control, there is little information about the accuracy of results in the critical ranges that may involve life-threatening conditions.
OBJECTIVES: To evaluate the reliability of POCG measurements in critical, high (>600 mg/dL) and low (<40 mg/dL) ranges.
DESIGN: One-year retrospective analysis of POCG (ACCU-CHEK glucose meter, Roche Diagnostics Corporation, Indianapolis, Indiana) results for routine patient care were obtained. The frequency and accuracy of repeat testing after critical POCG results was analyzed. A convenience sample of noncritical capillary POCG measurements retested on venous blood specimens by another point-of-care device (RAPIDPoint 405 analyzer, Siemens Medical Solutions USA, Malvern, Pennsylvania) was also evaluated.
RESULTS: Critical values were observed in 2.4 per 1000 POCG tests (256 of 105,928; 0.24%), with the highest rate (76 of 2289; 3.32%) from the emergency department. Twice as many critical high values as critical low values were seen. Nearly 80% of critical POCG tests (204 of 256) were repeated within 10 minutes. Of these 204 repeat measurements, 112 (54.9%) met accuracy criteria (±15 mg/dL of low and ±20% of high initial values). Accuracy was significantly higher when retesting was performed on the same meter or by the same operator (P ≤ .05). Comparison of capillary and venous POCG testing of noncritical results showed no significant difference (P = .95), with 89.8% (125 of 139) meeting accuracy criteria.
CONCLUSIONS: POCG measurements in the critical range are frequently erroneous, which is likely caused by preanalytic factors associated with sampling capillary blood. POCG testing practices should include retesting to confirm critical results.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24978924     DOI: 10.5858/arpa.2013-0455-OA

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  8 in total

1.  Evaluation of the Abbot FreeStyle Optium Neo H blood glucose meter in the hyperbaric oxygen environment.

Authors:  Carol R Baines; P David Cooper; Geraldine A O'Rourke; Charne Miller
Journal:  Diving Hyperb Med       Date:  2020-06-30       Impact factor: 0.887

2.  Utility of point-of-care vs reference laboratory testing for the evaluation of glucose levels.

Authors:  O M Andriankaja; F J Muñoz-Torres; J L Vergara; C M Pérez; K Joshipura
Journal:  Diabet Med       Date:  2019-03-01       Impact factor: 4.359

3.  Effective interventions to improve the quality of critically high point-of-care glucose meter results.

Authors:  Julie Lv Shaw; Christopher R McCudden; David A Colantonio; Ronald A Booth; Danny C Lin; Ivan M Blasutig; Thomas Moran; Dana Trofimczuk; Chantal Carriere; Anas Gharra; Cheryl Portelance; Chantal Tremblay; Dianne Dupaul; Nathalie Breton; Merina Angelkovski; Chetan Jariwala; Maureen Embleton; Christine Campbell; Kristina Groulx; Karen Larmour
Journal:  Pract Lab Med       Date:  2020-10-19

4.  Clinical staff knowledge and awareness of point-of-care-testing best practices at Tygerberg Hospital, South Africa.

Authors:  Thumeka P Jalavu; Megan Rensburg; Rajiv Erasmus
Journal:  Afr J Lab Med       Date:  2020-07-16

5.  Evaluation of CareSens POCT Devices for Glucose Testing in the Routine Hospital Setting.

Authors:  Kagan Huysal; Yasemin U Budak; Hakan Demirci; Mehmet Ali Önelge
Journal:  J Clin Diagn Res       Date:  2015-10-01

6.  Capillary versus Venous Blood Glucose in Patients with Coma.

Authors:  Viroj Wiwanitkit
Journal:  Adv Biomed Res       Date:  2017-07-14

7.  Comparison of glucose concentrations in canine whole blood, plasma, and serum measured with a veterinary point-of-care glucometer.

Authors:  Natalie D Suchowersky; Elizabeth A Carlson; Hollie P Lee; Ellen N Behrend
Journal:  J Vet Diagn Invest       Date:  2021-06-02       Impact factor: 1.569

8.  Using Plan-Do-Check-Act Circulation to Improve the Management of Panic Value in the Hospital.

Authors:  Suo-Wei Wu; Tong Chen; Yong Xuan; Xi-Wu Xu; Qi Pan; Liang-Yu Wei; Chao Li; Qin Wang
Journal:  Chin Med J (Engl)       Date:  2015-09-20       Impact factor: 2.628

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.