Literature DB >> 19454540

Comparison of blood glucose meters in a New Zealand diabetes centre.

Christopher Florkowski1, Christopher Budgen, Deborah Kendall, Helen Lunt, M Peter Moore.   

Abstract

BACKGROUND: To assess the accuracy and precision of the Roche Performa and Medisense Optium Xceed (5 and 10 s reading) blood glucose meters.
METHODS: Capillary blood samples were taken from 100 patients attending a diabetes centre and blood glucose measured on Roche Performa (n = 4) and Medisense Optium Xceed 5 s (n = 2) and 10 s reading (n = 2) meters. Venous plasma glucose from samples taken simultaneously was measured by the laboratory hexokinase method as reference standard. Imprecision was determined on the meters by replicate analysis (n = 20) of control solutions provided by the manufacturers and also patient venous whole-blood samples. Accuracy was assessed relative to the reference method by Bland-Altman plots, Passing and Bablok regression analysis, and both Clarke and consensus error grid analysis. Coefficients of variation (CVs) were calculated to determine imprecision.
RESULTS: Bland-Altman and Passing-Bablok analysis confirmed significant systematic bias for all meters, with relative under-reading of higher glucose concentrations. Error grid analysis showed that <5% readings exceeded +/-20% (or +/-0.83 mmol/L for readings <4 mmol/L) deviation from the reference method (1%, 2% and 4% for the Roche, Optium 5 and 10 s meters, respectively). CVs were all <4% for the control solutions and <6% for patient samples.
CONCLUSIONS: Both Roche Performa and Medisense Optium glucose meters (5 and 10 s readings) perform satisfactorily and are acceptable for operational use.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19454540     DOI: 10.1258/acb.2009.008193

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  4 in total

Review 1.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  Evaluation of CareSens POCT Devices for Glucose Testing in the Routine Hospital Setting.

Authors:  Kagan Huysal; Yasemin U Budak; Hakan Demirci; Mehmet Ali Önelge
Journal:  J Clin Diagn Res       Date:  2015-10-01

3.  Suboptimal performance of blood glucose meters in an antenatal diabetes clinic.

Authors:  Nimalie Jacintha Perera; Lynda Molyneaux; Maria Ines Constantino; Marg McGill; Dennis Koon-See Yue; Stephen Morris Twigg; Glynis Pauline Ross
Journal:  Diabetes Care       Date:  2011-01-07       Impact factor: 19.112

4.  The association of hyperglycaemia with prevalent tuberculosis: a population-based cross-sectional study.

Authors:  Sarah Lou Bailey; Helen Ayles; Nulda Beyers; Peter Godfrey-Faussett; Monde Muyoyeta; Elizabeth du Toit; John S Yudkin; Sian Floyd
Journal:  BMC Infect Dis       Date:  2016-12-05       Impact factor: 3.090

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.