Juan Cuadros1, Ramón Pérez-Tanoira2,3,4,5, Laura Prieto-Pérez2,3,4, Ines Martin-Martin2,6, Pedro Berzosa7, Vicenta González7, Gebre Tisiano2, Seble Balcha2, José Manuel Ramos2,8, Miguel Górgolas2,3,4.
Abstract
BACKGROUND: In up to one third of the hospitals in some rural areas of Africa, laboratory services in malaria diagnosis are limited to microscopy by thin film, as no capability to perform thick film exists (gold standard in terms of sensitivity for malaria diagnosis). A new rapid molecular malaria diagnostic test called Loop-mediated isothermal DNA amplification (LAMP) has been recently validated in clinical trials showing exceptional sensitivity and specificity features. It could be a reliable diagnostic tool to be implemented without special equipment or training.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26555068 PMCID: PMC4640523 DOI: 10.1371/journal.pone.0142842
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Comparison between LAMP, RDTs and on-site thin film microscopy in LAMP positive patients.
RDT: Rapid Diagnostic Test; Mic: Microscopy; Hb: Haemoglobin (g/dl); Pla: Platelets (/μl), m: male; f: female, N negative, Pf: Plasmodium falciparum, Pv Plasmodium vivax.
| Age | Sex | RDT | LAMP | Mic | Hb | Pla |
|---|---|---|---|---|---|---|
| 6 | m | Pf | Pf | N | 8.5 | 140,000 |
| 30 | f | N | Pf | N | 12.7 | 25,000 |
| 22 | f | N | Pf | N | 12.8 | 90,000 |
| 22 | f | N | Pf | N | 5.1 | 123,000 |
| 11 | f | N | Pf | N | 14.8 | 153,000 |
| 15 | m | N | Pv | N | 16.5 | 172,000 |
| 10 | m | N | Pv | N | 11.7 | 92,000 |
| 15 | m | Pf | Pf | N | 6.6 | 131,000 |
| 38 | f | N | Pf | N | 7.6 | 35,000 |
| 20 | f | N | Pf | N | 12.6 | 379,000 |
| 65 | m | Pf | Pf | N | 12.4 | 61,000 |
Sensitivity, specificity, PPV and NPV for RDTs were 27% (CI95%: 0.58%), 100% (95% CI: 99.6–100%), 100% (95% CI: 83.3%) and 94% (95% CI: 91.2–98.6%), respectively.
Quantitative diagnostic parameters and qualitative diagnostic parameters for LAMP positive patients.
|
|
|
|
|
|
| 23.091±16.670 | 24.720±14.971 | 0.593 |
|
| 5.182±3.573 | 6.618±3.880 | 0.186 |
|
| 8054.545±4887.610 | 8722.302±5031.078 | 0.397 |
|
| 2294.027±1789.285 | 1957.252±1311.938 | 0.840 |
|
| 5152.400±3267.980 | 5440.999±4529.712 | 0.979 |
|
| 9.900±4.832 | 13.021±3.226 | 0.015 |
|
| 110666.667±43247.351 | 256869.919±104907.606 | ≤0.001 |
|
|
|
|
|
|
| 6(54.54%) | 83(58.041%) | 0.821 |
|
| 2(18.18%) | 17(11.18%) | 0.667 |
|
| 9(81.82%) | 73(48.03%) | 0.091 |
|
| 6(54.54%) | 27(17.76%) | 0.012 |
|
| 3(27.27%) | 26(17.10%) | 0.395 |
|
| 8(72.73%) | 81(53.29%) | 0.472 |
*Other symptoms included general malaise, abdominal pain and cough.
A cost and time-to-completion analysis for all the methods used was done.
| Method | Cost of diagnosis/sample | Time for diagnosis | treatment cost | Total cost |
|---|---|---|---|---|
|
| 8.18$/sample | 1 hour/10 samples | 57,6$ | 1382.60$ |
|
| 0.60–1.00$/sample | 20 min-30min/10samples | 115.2$ | 212.4–277.2$ |
|
| 0.32–1.27$/sample | 30–60 min/sample | 158.4$ | 210.40–364.14$ |
1: DNA extraction takes 10min + LAMP reaction and reading 50min. Capacity of hot-block is 24 samples: each determination comprises two tubes, one for P. falciparum and one for Plasmodium spp.+ controls (positive and negative for P. falciparum and P. vivax).
2: To obtain the same sensitivity of examination as that for thick film at high power fields (with 100x oil immersion objective) for 10 min, a thin film should be examined for at least 30 min[13].
3: Only treatment with ACT for patients with negative diagnoses with each method were considered.
4: Treatment with ACT was considered exclusively.
5: Total cost of diagnosis + total cost of treatment.