MRI scanning parameters and estimation of SNRs
Figure 1Gel-measured dose at 0.35 T for a TG 119 prostate plan. (a) TG 119 prostate plan with a 6-Gy target dose; (b) amplitude map of the simulated noise at 0.35 T with a voxel size of 5 mm × 5 mm × 5 mm and scanning time of 35.2 min; (c) gel-measured dose at 0.35T; (d) gamma comparison (3%/3 mm); (e) and (f) profiles comparisons: Planned dose (solid line) and gel measured dose (circle)
Figure 2Planned dose, gel-measured dose, and gamma map at 0.35T for the other 5 TG 119 plans. (a) AP; (b) bands; (c) C-shape; (d) H and N; (e) multi-target. Target dose: 6Gy. The mean gamma passing rate and standard deviation were 98.1% ± 0.6% using 3%/3 mm criteria and 88.1% ± 1.8% using 2%/2 mm criteria