Azin Alizadehasl1, Mohsen Ziyaeifard1, Mohammadmehdi Peighambari1, Rasoul Azarfarin1, Ghodrat Golbargian1, Hooman Bakhshandeh1.
Abstract
BACKGROUND: Invasive and continuous blood pressure (BP) monitoring is crucial after cardiac surgery. Accuracy of BP measurement mostly depends on patency of arterial catheter and acceptable waveform. Heparinized saline flush usually used for this purpose may be accompanied by potential heparin adverse effects.Entities:
Keywords: Cardiac Surgery; Heparin; Saline Solution
Year: 2015 PMID: 26436074 PMCID: PMC4588703 DOI: 10.5812/cardiovascmed.28086v2
Source DB: PubMed Journal: Res Cardiovasc Med ISSN: 2251-9572
Figure 1.The Study Flow Diagram
Demographic and Clinical Characteristics of Patients in the Two Study Groups [a,b,c]
| Parameters | Heparinized Saline Group (n = 50) | Normal Saline Group (n = 50) | P Value |
|---|---|---|---|
|
| 49.7 ± 11.2 | 53.5 ± 12.2 | 0.194 |
|
| |||
| Male | 30 (60) | 39 (78) | 0.052 |
| Female | 20 (40) | 11 (22) | |
|
| 162 ± 14 | 160 ± 15 | 0.492 |
|
| 69.1 ± 11.1 | 67.4 ± 10.8 | 0.440 |
|
| |||
| CABG | 33 (66) | 27 (54) | 0.710 |
| Valve | 22 (44) | 28 (56) | |
| aPTT, sec [ | 33.7 ± 4.9 | 32.1 ± 4.2 | 0.083 |
| Platelet count, × 103
[ | 226 ± 51 | 240 ± 57 | 0.199 |
|
| |||
| 20 | 46 (92) | 48 (96) | 0.678 |
| 18 | 4 (8) | 2 (4) | |
|
| |||
| Radial | 42 (84) | 41 (82) | |
| Brachial | 5 (10) | 6 (12) | |
| Femoral | 2 (4) | 0 | 0.791 |
a Abbreviations: aPTT, Activated partial prothrombin time; and CABG, Coronary artery bypass grafting.
b Data are presented as No. (%) or Mean ± SD.
c In second postoperative day.
Frequency of Acceptable Arterial Waveform in First, Second and Third Postoperative Days [a]
| Variable | Heparinized Saline Group (n = 50) | Normal Saline Group (n = 50) | P Value |
|---|---|---|---|
|
| 39 (78) | 36 (72) | 0.774 |
|
| 40 (80) | 39 (78) | 0.793 |
|
| 33 (66) | 32 (65.3) | 0.887 |
a Data are presented as No. (%).