Evaluation of filtering methods for acquiring radial intra-artery blood pressure waveforms.

Lawrence T Hersh1, Bruce Friedman2, William Luczyk2, Jean Sesing2.   

Abstract

The methods for evaluating noninvasive blood pressure (NIBP) monitors using an intra-arterial reference are detailed in the ANSI/AAMI/ISO 81060-2:2009 standard. In a recent study, GE Healthcare obtained invasive radial arterial blood pressure waveforms. The work presented here describes the development of filtering strategies for obtaining high fidelity intra-arterial pressure waveforms for NIBP accuracy testing using the 81060-2 standard. The natural frequency and damping factor of each subject-catheter-transducer system was computed from fast-flush transients. These parameters were used to construct filters for removing or reducing resonance artifacts. Additionally, new optimal damping factors were evaluated for designing compensation filters. Theoretical measurement systems using actual damping factors (< 0.4) and natural frequencies were found capable of generating significant systolic resonance artifacts (≥ 8 mmHg). Typical filters that may be standardly available in monitoring equipment were observed to be potentially inadequate in removing resonance artifact. Filters with particular optimal damping factors (0.6-0.7) were effective in removing resonance artifact. Clinicians need to understand that resonance artifacts potentially exist in intra-arterial waveforms and that the adjustments of monitoring systems may not be adequate. Optimal filters for obtaining intra-arterial waveforms should take into account the damping factor and natural frequency of the measuring system. In research and device evaluation studies it is necessary that optimal filtering be done to minimize the effects of under-damping.

Keywords:  Damping factor; Filter design; Intra-arterial blood pressure waveform filtering; Natural frequency; Noninvasive blood pressure

Mesh:

Year:  2014        PMID: 25516162     DOI: 10.1007/s10877-014-9649-4

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  6 in total

Review 1.  Understanding natural frequency and damping and how they relate to the measurement of blood pressure.

Authors:  B Kleinman
Journal:  J Clin Monit       Date:  1989-04

2.  Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems.

Authors:  Stefano Romagnoli; Salvatore Mario Romano; Sergio Bevilacqua; Chiara Lazzeri; Gian Franco Gensini; Carlo Pratesi; Diego Quattrone; Daniele Dini; Angelo Raffaele De Gaudio
Journal:  J Crit Care       Date:  2010-10-30       Impact factor: 3.425

3.  Electronic compensator for pressure waveform distortion by fluid-filled catheters.

Authors:  A Damenstein; R L Stout; H U Wessel; M H Paul
Journal:  Med Biol Eng       Date:  1976-03

4.  Semiautomatic algorithm to remove resonance artifacts from the direct radial artery pressure.

Authors:  H A Schwid
Journal:  Biomed Instrum Technol       Date:  1989 Jan-Feb

5.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

6.  Validation of a conical cuff on the forearm for estimating radial artery blood pressure.

Authors:  Lawrence T Hersh; Jean C Sesing; William J Luczyk; Bruce A Friedman; Shanggen Zhou; Paul B Batchelder
Journal:  Blood Press Monit       Date:  2014-02       Impact factor: 1.444

  6 in total
  2 in total

1.  Avoiding Heparinization of Arterial Line and Maintaining Acceptable Arterial Waveform After Cardiac Surgery: A Randomized Clinical Trial.

Authors:  Azin Alizadehasl; Mohsen Ziyaeifard; Mohammadmehdi Peighambari; Rasoul Azarfarin; Ghodrat Golbargian; Hooman Bakhshandeh
Journal:  Res Cardiovasc Med       Date:  2015-08-01

2.  Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery.

Authors:  Irwin Gratz; Edward Deal; Francis Spitz; Martin Baruch; I Elaine Allen; Julia E Seaman; Erin Pukenas; Smith Jean
Journal:  BMC Anesthesiol       Date:  2017-03-21       Impact factor: 2.217

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.